Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 77 - 77
4 Apr 2023
Sharrock M Fermor H Redmond A Brockett C
Full Access

This study aims to assess the changes in mechanical behaviour over time in ‘haemarthritic’ articular cartilage compared to ‘healthy’ articular cartilage.

Pin-on-plate and indentation tests were used to determine the coefficient of friction (COF) and deformation of ‘healthy’ and ‘haemarthritic articular cartilage. Osteochondral pins (8 mm) were extracted from porcine tali and immersed in exposure fluid for two hours prior to test. Pins were articulated against a larger bovine femoral plate for 3600 seconds under a load of 50 N. Osteochondral pins (8 mm) were loaded during indentation testing for 3600 seconds under a load of 0.25 N. To mimic the effect of a joint bleed in vitro; serum, whole blood and 50% v/v were used as exposure and lubricant fluids. COF and deformation were expressed as mean (n=3) and statistically analysed using a one-way ANOVA and post-hoc Tukey test (p>0.05).

The serum condition yielded a COF of 0.0428 ± 0.02 with 0.08mm ± 0.04 deformation. The 50% v/v condition produced a higher COF of 0.0485 ± 0.02 and 0.21mm ± 0.04 deformation. The lowest COF and deformation were produced by the whole blood condition (0.0292 ± 0.02 and 0.06mm ± 0.006 respectively). Statistical analysis indicated no significant difference across the friction test conditions but a significant difference across all indentation test conditions (ANOVA, p>0.05). Combination of creep deformation and wear was observed on the articular surface up to 24 hours post-test in 50% v/v and whole blood conditions.

The average haemophilia patient can experience multiple joint bleeds per year of which this study demonstrates the effect of just one joint bleed. This study has provided evidence of potential reversible and irreversible mechanical changes to articular cartilage surface during a joint bleed.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 62 - 62
1 Mar 2021
Talbott H Wilkins R Cooper R Redmond A Brockett C Mengoni M
Full Access

Abstract

OBJECTIVE

Flattening of the talar dome is observed clinically in haemarthropathy as structural and functional changes advance but has not been quantified yet. In order to confirm clinical observation, and assess the degree of change, morphological measurements were derived from MR images.

METHODS

Four measurements were taken, using ImageJ (1.52v), from sagittal MRI projections at three locations – medial, lateral and central: Trochlear Tali Arc Length (TaAL), Talar Height (TaH), Trochlear Tali Length (TaL), and Trochlear Tali Radius (TaR). These measurements were used to generate three ratios of interest: TaR:TaAL, TaAL:TaL, and TaL:TaH. With the hypothesis of a flattening of the talar dome with haemarthropathy, it was expected that TaR:TaAL and TaL:TaH should be greater for haemophilic ankles, and TaAL:TaL should be smaller. A total of 126 MR images (ethics: MEEC 18–022) were included to assess the difference in those ratios between non-diseased ankles (33 images from 11 volunteers) and haemophilic ankles (93 images from 8 patients’ ankles). Non-diseased control measurements were compared to literature to assess the capacity of doing measurements on MRI instead of radiographs or CT.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 38 - 38
1 Mar 2021
Vasiljeva K Lunn D Chapman G Redmond A Wang L Thompson J Williams S Wilcox R Jones A
Full Access

Abstract

Objectives

The importance of cup position on the performance of total hip replacements (THR) has been demonstrated in in vitro hip simulator tests and clinically. However, how cup position changes during gait has not been considered and may affect failure scenarios. The aim of this study was to assess dynamic cup version using gait data.

Methods

Pelvic movement data for walking for 39 unilateral THR patients was acquired (Leeds Biomedical Research Centre). Patient's elected walking speed was used to group patients into high- and low-functioning (mean speed, 1.36(SD 0.09)ms−1 and 0.85(SD 0.08)ms−1 respectively). A computational algorithm (Python3.7) was developed to calculate cup version during gait cycle. Inputs were pelvic angles and initial cup orientation (assumed to be 45° inclination and 7° version, anterior pelvic plane was parallel to radiological frontal plane). Outputs were cup version angles during a gait cycle (101 measurements/cycle). Minimum, maximum and average cup version during gait cycle were measured for each patient. Two-sample t-test (p=0.05) was used to compare groups.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 31 - 31
1 Mar 2021
Hopwood J Redmond A Chapman G Richards L Collins S Brockett C
Full Access

Abstract

Objectives

Implant loosening remains a common cause of total ankle replacement (TAR) revision, and has been associated with wear-mediated osteolysis. Limited pre-clinical studies for TARs have been reported and the variety of experiment settings make it difficult to compare wear rates. Factors such as simulator control mechanism; whether pneumatic or electromechanical, may influence the integrity of the simulator outputs with respect to input profiles. This study compares the wear of a TAR, tested in electromechanical and pneumatic experimental simulators under identical input conditions.

Methods

Twelve medium BOX® (MatOrtho Ltd) TARs (n=6 for each simulator) were tested in an electromechanical and pneumatic knee simulator (Simulation Solutions, UK) for 3 million cycles (Mc). Standard ‘Leeds’ displacement-controlled inputs were used. Kinematic performance was investigated by comparing the output profiles against the maximum demanded input values. The lubricant used was 25% new-born calf serum and wear was determined gravimetrically.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_2 | Pages 33 - 33
1 Jan 2019
Hopwood J Chapman G Redmond A Richards L Brockett C
Full Access

Total ankle replacement (TAR) is a substitute to ankle fusion, replacing the degenerated joint with a mechanical motion-conserving alternative. Compared with hip and knee replacements, TARs remain to be implanted in much smaller numbers, due to the surgical complexity and low mid-to-long term survival rates. TAR manufacturers have recently explored the use of varying implant sizes to improve TAR performance. This would allow surgeons a wider scope for implanting devices for varying patient demographics. Minimal pre-clinical testing has been demonstrated to date, while existing wear simulation standards lack definition. Clinical failure of TARs and limited research into wear testing defined a need for further investigation into the wear performance of TARs to understand the effects of the kinematics on varying implant sizes.

Six medium and six extra small BOX® (MatOrtho) TARs will be tested in a modified knee simulator for 5 million cycles (Mc). The combinations of simulator inputs that mimic natural gait conditions were extracted from ankle kinematic profiles defined in previous literature. The peak axial load will be 3.15 kN, which is equivalent to 4.5 times body weight of a 70kg individual. The flexion profile ranges from 15° plantarflexion to 15° dorsiflexion. Rotation about the tibial component will range from −2.3° of internal rotation to 8° external rotation, while the anterior/posterior displacement will be 7mm anterior to −2mm posterior throughout the gait cycle. The components will be rotated through the simulation stations every Mc to account for inter-station variability. Gravimetric measurements of polyethylene wear will be taken at every Mc stage. A contact profilometer will also be used to measure average surface roughness of each articulating surface pre-and-post simulation.

The development of such methods will be crucial in the ongoing improvement of TARs, and in enhancing clinical functionality, through understanding the envelope of TAR performance.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_16 | Pages 44 - 44
1 Oct 2016
Stratton-Powell A Tipper JL Williams SD Redmond A Brockett CL
Full Access

Total ankle replacement (TAR) has a mean survivorship of 77% at 10 years which is poor compared to other types of joint arthroplasty. Osteolysis and aseptic loosening are commonly cited TAR failure modes, the mechanisms of which are unknown. Retrieval analyses of TAR devices may reveal mechanisms of failure similar or dissimilar to other joint replacements. This study investigated whether TAR explants exhibit similar damage modes to those recognised in other total joint replacements.

22 Ankle Evolution System TARs (Transystème, Nimes, France) were implanted and retrieved by the same surgeon. Mean implantation time was 7.8 yrs (5.3 to 12.1 range). Pain and/or loosening were the indications for revision. Macro photography, an Alicona Infinite microscope and the Hood/Wasielewski scale were used to classify damage modes on the polyethylene insert. Scanning electron microscopy with energy dispersive X-ray spectroscopy was used to determine the composition of third body debris and to image the fixation surface of the tibial components.

Mean damage score was 185.4 (± 40.0 SD). Damage modes common to total knee replacements were identified on both the superior and inferior insert surfaces, these included: burnishing, scratching, pitting and abrasion. Titanium particles, hydroxyapatite fragments and bone debris were embedded in the insert surfaces. Fixation surface delamination was identified by the ongrowth of tissue between the cobalt chromium substrate and titanium alloy coating.

Damage modes indicative of high levels of wear and deformation were evident. Pitting caused by third body debris was abundant and suggested fixation surface wear and failure.