header advert
Results 1 - 3 of 3
Results per page:
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 142 - 142
1 Feb 2017
LiArno S Gopalakrishnan A Schmidig G Schmidt W Racanelli J
Full Access

INTRODUCTION

During activities of daily living (ADL), varus moments are experienced in the knee, which can result in frontal plane rotation, or liftoff, of the lateral femoral condyle with respect to the tibial plateau. An understanding of this rotation is valuable as it could potentially lead to contact between the femoral component and polyethylene post of a total knee replacement (TKR). Therefore, the purpose of this study was 1) to assess how much frontal plane rotation was achieved due to varus moments imposed on a total stabilized (TS) TKR from the stair ascent activity, and 2) to determine whether a TS TKR could withstand the contact stresses imposed by the varus loading for 1 million cycles without the post fracturing or plastically deforming.

METHODS

A PS femoral component paired with a TS polyethylene insert and baseplate (Triathlon, Stryker, Mahwah, NJ) were aligned on a multi-axis testing system (MTS Systems Corp, Eden Prairie, MN) (Figure 1). Size 1 components were used as they represented the worst-case size for testing. The femoral component was fixed at 60 degrees of flexion, representing an angle of peak varus moment during stair ascent [1]. The peak varus moment used in this study was determined by scaling the data from Orthoload.com for a 136 kg patient body weight (3 SD above average TKR patient body weight) [1, 2].

In order to evaluate the frontal plane rotation achieved due to the varus moment with minimal influence from other loads, an FEA model of the physical test setup was used to determine the lowest joint compressive load that would allow testing to be stable. Given this, testing was completed with a constant joint compressive load of 1500 N (33% of that reported by Orthoload.com) while sinusoidally applying a varus moment from 5Nm to 54.5Nm [1, 2]. The loads were applied to three samples for 1 million cycles to represent the number of stair ascent cycles experienced over 20 years [3].

Lastly, a validation test was run on a component with the polyethylene post notched at the medial distal aspect. The post fractured during testing indicating that the test could induce the clinical failure mode of interest.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 424 - 424
1 Nov 2011
Fuchs J Shields W Schmidt W Liepins I Racanelli J
Full Access

Introduction: Uncemented proximally filling porous-coated femoral components must be designed with an optimal level of press-fit. Excessive press-fit yields higher femoral stress which can result in periprosthetic femoral fracture (PPFx), whereas insufficient femoral stress can lead to a lack of initial mechanical stability, which “is necessary to achieve bone ingrowth into the porous surface” (Manley P.A. et. al., J Arthroplasty10:63–73, 1995) of the implant. An optimal press-fit design should also provide an accurate and repeatable femoral stem seating height in all patients.

A battery of cadaveric tests, physical “bench-top” tests, and finite element analyses (FEA) should be used in order to both quantitatively and qualitatively optimize a femoral press-fit design. In this study, a method is proposed to quantitatively rank candidate press-fit stem designs relative to successful predicates based on stem seating height and PPFx risk by recreating impact loading applied during surgery through a controlled “bench-top” model.

Methods: Three press fit candidate designs A, B & C and two clinically successful predicate proximal fit and fill stems (Secur-Fit™ Max (Fit & Fill 1) and Meridian® TMZF® (Fit & Fill 2), Stryker, Mahwah NJ) were evaluated. Five foam cortical shell Sawbones® femur samples (Item# 1130, Pacific Research Laboratories, Inc., Vashon, WA) were prepared for each press-fit design. A stem impactor was attached to the stem and then the stem was hand inserted in the femur. Then the construct was mounted in the drop tower using a vice and initial drop height was set to generate approximately 5500 N of impaction force when fully seated. Each stem was serially impacted until stable then step loaded until PPFx occurred. The height above/below the medial resection plane was measured after each impaction.

Results: All press-fit designs had an initial stable seating height within the desired range without causing PPFx, using an average impaction load of 5341 N. All of the press-fit designs required, on average, roughly a 200% increase in impact load (10925 N) to cause PPFx. The press-fit deign which ranked first based on seating height accuracy, defined as the design closest to zero at stable, was Design C at −0.02 mm countersunk. Design A with a standard deviation of 0.09 mm ranked first for repeatability, defined as the design with the smallest standard deviation at stable. Finally the press-fit design which ranked first for lowest PPFx risk, defined as the design that is most countersunk prior to PPFx, was Fit & Fill 1 at 6.30 mm countersunk.

Discussion: This controlled “bench-top” impact loading model successfully showed that it can quantitatively evaluate stem seating height and PPFx risk for several different femoral press-fit designs. In order to determine the optimal design, each press-fit design was ranked with equal weight given to seating height and fracture risk. Using this test method one design alternative, press-fit Design C, ranked first as the optimal combination of seating height accuracy and consistency with a low risk of PPFx. A limitation of this impaction model is that it does not directly predict PPFx rate, it only quantifies risk of fracture. Another limitation is that this model does not simulate all of the variably that is inherent to actual patient bone types. This test is one step in a battery of tests, including cadaveric evaluation and FEA, which should be used in order to optimize a femoral press-fit design.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 30 - 30
1 Mar 2009
Gillies R Hogg M Donohoo S Schmidt W Racanelli J
Full Access

Introduction: Bone resorption at the bone-implant interface is still a problem, leading to pain, poor function and the possibility of bone fracture. This loss of supporting bone tissue is due to resorption and impaired bone formation. Loosening of an implant is often not clinically or radiographically apparent for 8–10 years. It would be beneficial if these potential failures could be identified early so that revision surgery can be avoided. The aim of this study was to investigate the influence of implant material property changes and its influence on the trabecular loading patterns of the underlying supporting bone structure.

Methods: An intact and reconstructed 3D finite element (FE) model of a human femur was developed. The model was generated using PATRAN and CT scans. This was used to determine the stress, strain and interface sliding of a knee implant at heel-strike and stair climbing phases of gait. FE analysis of the model was performed using ABAQUS software. The materials properties of the bone were extracted from the CT data and applied using FORTRAN subroutines. Implant-bone interfaces were simulated using cementless fixation concepts. Sliding contact conditions were applied to simulate the immediate post-operative period.

Results: Three material property cases were analysed, with respect to the intact bone, at 100%, 25% and 2.5% of cobalt chrome’s (CoCr) Youngs modulus. At heel-strike, for the 100% case, higher stress was found at anterior flange while lower stress dominated around the pegs and intercondylar notch. For the 25% case, lower stresses were found in the intercondylar notch and higher stresses above the pegs. For the 2.5% case, stresses resembled that of intact bone, higher stresses were found above the pegs and lower stress in the intercondylar notch. In stair-climbing, for the 100% case, lower stresses were found around the pegs and in the intercondylar notch. For the 25% case, lower stresses were found in the intercondylar notch and higher stresses in areas above the pegs. For the 2.5% case, higher stresses were found at the distal condyles and lower stresses were observed in the intercondylar notch.

Discussion: The analysis presented changes in the trabecular loading and subsequently resulted in stress shielding. The general trend showed that the majority of stress shielding is occurring at the posterior flange and medial condyle while increased trabecular loading occurred at the anterior flange and lateral condyle regions. As the stiffness of the implant decreases from 100% to 25%, the differences in trabecular loading are extremely small. Both these implant material properties are very stiff in comparison to the underlying trabecular bone. However, as CoCr stiffness is decreased to 2.5% this yields a more homogenous stress distribution at the contact interfaces.