header advert
Results 1 - 3 of 3
Results per page:
Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_II | Pages 301 - 302
1 May 2010
Jameson S Ramisetty N Langton D Webb J Logishetty R Nargol A
Full Access

Introduction: There are ongoing concerns regarding metal wear debris following the use of metal-on-metal (MonM) bearings for hip surface and total arthroplasty. A Type IV Hypersensitivity reaction to MonM articulations has previously been identified (aseptic lymphocyte dominated vasculitis associated lesion, ALVAL) but little is known of its incidence, diagnosis or management. Persisting groin pain in MonM patients may be undiagnosed ALVAL. At our single centre we have reviewed and compared three types of MonM articulations to examine the incidence of ALVAL and to identify trends.

Methods: The resurfacing group comprised 250 patients with the ASR prosthesis. In the resurfacing hybrid total hip replacement (THR) group there were 86 patients implanted with an ASR head on a stem. The final group comprised of 625 patients with a MonM THR using a 36mm Pinnacle head. Both the S-ROM and the Corail stems were used in the THR groups. Patients with persisting and activity-restricting groin pain had tests for infection. Patients were counselled and revision was offered if ALVAL was suspected from the clinical picture, blood results and the aspiration result. Specimens for microbiological and histological analysis were taken at the time of revision.

Results: We found 5 cases of histologically proven ALVAL in the absence of infection in 961 patients. The incidence was: 1.2% in the resurfacing group, 2.3% of Resurfacing Hybrid THR group and 0 in the 36 mm THR group. All 5 cases were in female patients. Only 1 case had any radiological abnormality. One patient was initially revised from a resurfacing to a 36mm MonM THR without clinical success. All patients have now been revised to ceramic-on-ceramic bearings with improvements in outcome.

Discussion: ALVAL may be under-diagnosed. The 5 patients we describe showed good clinical recovery following their primary procedure. However, activity levels decreased and pain increased at 6–12 months post-op. All described non-specific systemic symptoms. On examination, a painful straight leg raise was a characteristic finding. This may result from the significant effusion found around the hip at each revision. Fluid aspirated from these hips was of a characteristic colour (green grey) and viscosity. The failure of the revision of a resurfacing to a smaller MonM bearing highlights the problem of sensitisation to the metal debris. Any subsequent revision to a MonM bearing is unlikely to improve clinical outcome. This finding is consistent with previous reports in the literature. Our results suggest the incidence of ALVAL may be higher that previously thought. We suggest all patients with significant groin pain should have inflammatory markers tests and a hip aspiration performed. In the absence of infection, revision to an alternative bearing surface may be indicated.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_II | Pages 302 - 302
1 May 2010
Jameson S Webb J Langton D Ramisetty N Logishetty R Nargol A
Full Access

Introduction: Potential problems resulting from metal-on-metal hip wear debris are well known. Previous studies have shown an association between high cup angles and raised metal ion levels, but a link to clinical outcome has yet to be established. We aim to show the relationship between high cup angles, raised metal ion levels and pain following hip resurfacing.

Methods: This analysis comprised the first 250 Depuy ASR hip resurfacings performed by a single surgeon (senior author) over a 3 yr period at an independent centre. Patients were followed up, examined and x-rayed at regular intervals. We measured Harris Hip Score (HHS) and pain. The acetabular cup inclination angles were measured from standardised pelvic radiographs. A subgroup of 80 patients had serum and whole blood chromium and cobalt measurements performed. This subgroup comprised patients with a similar sex split, age, activity level, follow-up and cup angle profile to the parent group.

Results: HHS improved from 51.6 pre-operatively to 94.6 post-operatively. Males had a significantly higher post-op HHS (97.1) compared with the females (91.0). 2.1% of male patients had pain compared with 8.7% of females. High cup angles were associated with pain in females. This relationship did not occur in males. In patients with cup angles of 48° and above there were no males with pain compared to 15.4% of females. Females under 48° had a HHS of 93.7 compared to 88.0 for those over 48°. When we analysed metal ion levels there was a similar relationship. Patients with cup angles of 48° and above had significantly higher serum and whole blood chromium and cobalt levels compared with lower cup angles. Three patients with cup angles over 50° developed ALVAL (aseptic lymphocyte dominated vasculitis associated lesion) based on histological findings. All were female.

Discussion: High cup angles are associated with increased pain and elevated metal ion levels. Women seem to be more intolerant of an higher cup angle than men. This may be a consequence of the smaller prostheses in females. Cups implanted lower than 48° had a better clinical outcome and less metal ion wear. Like other 4th generation designs, the ASR cup is not a complete hemisphere (unlike the BHR) and may be susceptible to edge loading at lower inclination angles than previously thought. We believe the current recommendation of 45° +/−5° is too high based on our findings. We recommend an inclination angle that does not exceed 48°.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 2 | Pages 280 - 280
1 Feb 2007
RAMISETTY N GREISS ME