Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 53 - 53
1 Feb 2020
Cubillos PO Santos VD Pizzolatti ALA De Mello Roesler CR
Full Access

Introduction

The release of metallic debris can promote many adverse tissue reactions, as metallosis, necrosis, pseudotumors and osteolysis 1–3. This debris is mainly generated by the fretting-corrosion mechanism due to the geometric difference in the head-stem interface 4. Retrieval and in silico analysis showed the roughness of the stem-head interface appears to play an important role in the volume of material lost and THA failure 5–7. The technical standard ISO 7206-2 recommends the measurement of average roughness (Ra) and max height of the profile (Rz) to control the quality of the surface finish of articulating surfaces on THA implants. However, despite the importance of the trunnion roughness, there is no specific requirement for this variable on the referred technical standard. The present study carried out a surface finish analysis of the trunnion of hip stems from five distinct manufacturers.

Methods

Four stems (n = 4) from five (5) distinct manufacturers (A, B, C, D, and E) were used to evaluate the roughness of the trunnion. All the stems are similar to the classical Exeter stem design, with a 12/14 taper and a polished body surface.

The roughness of trunnions was evaluated according to ISO 4287 and ISO 13565-2. The total assessment length was 4.8 mm with 0.8 mm cut-off. The first and last 8.33% of assessment length were not considered. The measurements of all samples were made in a rugosimeter with 2 µm feeler ITP (Völklingen, Germany), the velocity of 0.5 mm.s-1, and a force of 1.5 mN. The calibration was made at 20 ºC and relative humidity at 50%.

The Kruskal Wallis with post hoc Nemenyi test was used to evaluate the difference of Ra among the manufacturers. The confidence level was set at 5%.