The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 54 - 58
1 May 2024
Wassilew GI Zimmerer A Fischer M Nonnenmacher L O'Hara L Hube R

Aims

The use of a porous metal shell supported by two augments with the ‘footing’ technique is one solution to manage Paprosky IIIB acetabular defects in revision total hip arthroplasty. The aim of this study was to assess the medium-term implant survival and radiological and clinical outcomes of this technique.

Methods

We undertook a retrospective, two-centre series of 39 hips in 39 patients (15 male, 24 female) treated with the ‘footing’ technique for Paprosky IIIB acetabular defects between 2007 and 2020. The median age at the time of surgery was 64.4 years (interquartile range (IQR) 54.4 to 71.0). The median follow-up was 3.9 years (IQR 3.1 to 7.0).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 17 - 17
1 Dec 2013
Bolland B Gardner E Roques A Maul C Culliford D Zeineh N O'Hara L
Full Access

This study reports the mid-term results of a large bearing uncemented metal on metal total hip replacement (MOMHTHR) matched series using the Synergy stem and Birmingham modular head in 36 hips (mean follow up 61 months). All patients underwent clinical, metal ion and MRI assessment. Wear analysis was performed on retrieved heads using Redlux non-contact optical profilometry.

Seven patients (19%) have undergone revision surgery. All revisions had two or more of either symptoms, high metal ions or an MRI suggestive of an adverse reaction to metal debris (ARMD). There was no evidence of component malposition or impingement. Frank staining of tissues together with high volume dark brown fluid collections were found in all cases. All stems and cups were well fixed. In 4 cases pubic and ischial lysis (adjacent to the inferior fins) was observed. All 7 cases had radiological, intraoperative and histological evidence of ARMD (Figure 1). The failure cohort had significantly higher whole blood cobalt ion levels and OHS (p = 0.001), but no significant difference in cup size (p = 0.77), gender predominance, stem offset or cup position (p = 0.12). Sleeves had been used in all revision cases

Wear analysis (n = 4) demonstrated increased wear at the trunnion/sleeve interface in a distribution compatible with micromotion (Figure 2). There was normal wear at the articulating surface.

This series further demonstrates unacceptable failure rates in LHMOMTHR in a series where a compatible stem for the BHR modular head was used. Use of a CoCr sleeve within a CoCr head taper appears to contribute to abnormal wear and therefore potential ARMD and subsequent failure.