Background

Calipered kinematically aligned (KA) total knee arthroplasty (TKA) restores the in vitro internal-external (I-E) rotation laxities at 0° and 90° of the native knee. Although increasing and decreasing the thickness of the insert in 1 mm increments loosens and tightens the flexion space, there are little data on how this might adversely affect the screw-home mechanism and I-E rotational laxity. The present study determined the differences in the I-E range of rotation and I-E positions at maximum extension and at 90° of flexion that result from the use of insert thicknesses that deviate ± 1mm in thickness from the implanted insert.

Methods

20 patients were treated with a calipered KA and a PCL retaining implant with a 1:1 medial ball-in-socket constraint and a non-constrained lateral flat articular insert surface. Verification checks, that are validated to restore native tibial compartment forces without release of healthy ligaments including the PCL, were used to select the optimal insert thickness. Trial inserts with thicknesses ranging from 10 to 13 mm were 3-D printed with medial goniometric markings that record rotation from 20° external to −20° internal with respect to a sagittal line laser marked on center of the medial condyle of the trial femoral component at maximum extension and 90° of flexion (Figure 1).