header advert
Results 1 - 2 of 2
Results per page:
Full Access

We report the outcome of 320 primary Total Hip Arthroplasties (THA) with minimum 10-year follow-up (range 10-17 years, mean 12.6 years), performed by a single surgeon in Tauranga New Zealand, with the Exeter Contemporary Flanged all-polyethylene cup and Exeter femoral stem via a posterior approach. The aim of the study is to compare the results with the published results from the design centre and create a baseline cohort for further outcomes research in this centre.

All patients were prospectively followed at 6 weeks, 1 year, 5 years, 10 years, (and 15 years when available). Of 333 cases that matched the inclusion criteria, 13 procedures in 12 patents were excluded because of concomitant bone grafting and/or supplementary cage fixation, leaving 320 primary THA procedures in 280 patients, including 26 bilateral procedures in 13 patients.

Mean follow-up of the surviving cases was 12.6 (range 5.0-17.1) years. There were 12 revisions – 2 for fracture, 5 for instability, 1 for impingement pain and 4 for infection. There were no revisions for aseptic cup loosening. Kaplan-Meier survivorship with revision for aseptic loosening as the endpoint was 100% at 15.0 years (with minimum 40 cases remaining at risk). All-cause acetabular revision in 12 cases result in a Kaplan-Meier survival of 95.9% (95% CI: 93.5 to 98.3%).

Cemented THA with the Exeter Contemporary Flanged cup and the Exeter stem is a durable combination with results that can be replicated outside of the design centre. The Exeter Contemporary Flanged cup has excellent survivorship at 15 years when used with the Exeter stem. Cemented THA with well-proven components should be considered the benchmark against which newer designs and materials should be compared.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 12 - 12
1 Mar 2008
Lamberton TD Charity J Kenny P Timperley AJ Gie G
Full Access

Impaction bone grafting with a cemented polished double-taper stem as a technique for revision of the femoral component was introduced in 1987 at our institution.

As at January 2000, 540 cases in 487 patients had been performed. All procedures have been studied prospectively and there are no patients lost to follow-up. We present the survivorship and outcome data for these patients.

Survivorship at 15 years is 90.6 percent [95 percent confidence interval: 88–93 percent]. Clinical scores show marked and sustained improvement.

There have been 45 failures [8.3 percent]. Technical error contributed to 13 of the 24 non-infective complications, but with improved technique plus the addition of long stemmed impaction grafting, there have been no technical errors since 1996.

Our results show that revision of the femoral component with impaction bone grafting is a reliable and durable technique with an acceptably low complication rate and with excellent survivorship at 15 years.