Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 33 - 33
1 Mar 2017
Mueller U Kretzer J
Full Access

Introduction

In total hip arthroplasty the femoral head is connected to the stem based on a taper connection (Fig. 1). Implant manufacturers proclaim that the modular tapers are not standardized and can vary from manufacturer to manufacturer. The combination of different implant components from different manufacturers (Mix & Match) is not permitted. However, in case of revision surgery, where the stem is still well fixed, the surgeon may decide to use a femoral head of a different manufacturer (Mix & Match). This decision may be related to a limited availability of a manufacturer-identical head or the manufacturer can't be identified. In this study, different taper combinations were experimentally investigated to assess the effect of Mix & Match on taper strength.

Methods

Hip stems and metal heads (n=3) of four different implant manufacturers (Biomet, DePuy, Smith&Nephew, Zimmer) were used. Firstly, torque-off tests similar to the ISO draft (ISO/DIS 7206-13) were performed without mixing the components. Subsequently, the stems were combined with metal heads from other manufacturers. All taper connections were impacted with 3 kN and then an increasing torque was applied until the head disconnected using a material testing machine (MTS Mini Bionix), (Fig. 2). The maximal torque off value was used as a measure for the taper strength.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 36 - 36
1 Mar 2017
Mueller U Kretzer J
Full Access

Introduction

Taper corrosion and fretting has been identified to be a major problem in total hip replacement during the past years. Taper design and manufacturing are not been standardised, and therefore it can be assumed that the tapers vary among different implant manufacturers. This can lead to variable contact situations and stresses in the taper junction depending on the combination. It can be assumed that the taper strength will influence the occurrence and magnitude of micromotions which are known to influence corrosion. Therefore, the aim of this study was to assess the influence of the taper angle clearance on the taper connection strength.

Material & Methods

For the investigation stem dummys with different taper angles were used that were manufactured from titanium alloy. The stem dummys were combined with ceramic heads with identically taper angles. Out of this, there were seven groups ranging from distal contact through full contact up to proximal contact. Three samples were used in each group and five repetitive measurements per samples were performed.

All taper connections were impacted with different forces (1 kN, 3 kN, 6 kN and 10 kN) and afterwards an increasing torque was applied until the head disconnected. The maximal torque off value was used as a measure for the taper strength.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 107 - 107
1 Mar 2017
Reiner T Bader N Panzram B Kretzer J Zeifang F
Full Access

Purpose

Total shoulder arthroplasty (TSA) has become a successful treatment option for degenerative shoulder disease. With the increasing incidence in primary TSA procedures during the last decades, strategies to improve implant longevity become more relevant. Implant failure is mainly associated with mechanical or biological causes. Chronic inflammation as a response to wear particle exposure is regarded as a main biological mechanism leading to implant failure. Metal ions released by fretting and corrosion at modular taper connections of orthopedic implants can cause cell-mediated hypersensitivity reactions and might lead to aseptic loosening. Modularity is also commonly used in total shoulder replacement. However, little is known about metal ion exposure in patients following TSA. The objective of this study was to determine in-vivo blood metal ion levels in patients after TSA and to compare blood metal ion levels to control subjects without metal implants.

Methods

A total of 19 patients with anatomical total shoulder prosthesis (TSA group) and 20 patients with reverse total shoulder prosthesis (RSA group) who underwent unilateral total shoulder replacement at our hospital between March 2011 and December 2014 with no other metal implant or history of environmental metal ion exposure were recruited for analysis of blood metal ion concentrations of cobalt (Co), chromium (Cr) and titanium (Ti) at a mean follow-up period of 2.3 years (0.7–4.3). For comparison of metal ion concentrations blood samples were obtained in a healthy control group of 23 subjects without metal implants. Ethical approval and informed consent of each patient were obtained for this study.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 80 - 80
1 Mar 2017
Paulus A Ebinger K Hasselt S Jansson V Bader R Kretzer J Utzschneider S
Full Access

Introduction

Metal on metal bearings are used especially in hip resurfacing. On the one hand, small bone preserving implants can be used. On the other hand recent studies found a variety of local and systemic side effects, for instance the appearance of pseudotumors, that are explained by pathologic biological reaction of the metal wear debris. The detailed mechanisms are still not understood until now. Thus it was the aim of this study to investigate the local reaction of metal wear particles and metal ions in a murine model. The hypothesis was that mainly metal ions provoke adverse histopathological reactions in vivo.

Material and Methods

Three groups, each with 10 Balb / c mice were generated. Group A: injection of a 50 µl metal ion suspension at a concentration of 200 µg / l in the left knee. Group B: injection of a 50 µl 0,1 vol% metal particle suspension into the left knee joint. Group C (control group): injection of a 50 µl of 0,1 vol% PBS-suspension in the left knee. Incubation for 7 days, followed by euthanasia of the animals by intracardiac pentobarbital. The left and right knee, the lungs, kidneys, liver and spleen were removed. Histologic paraffin sections in 2 microns thickness were made, followed by HE (overview staining) and Movat (Pentachrom staining) staining. The histologic analysis was a done by a light microscopic evaluation of the subdivided visual fields at 200× magnification.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 13 - 13
1 Mar 2017
Sonntag R Al-Salehi L Braun S Mueller U Reinders J Kretzer J
Full Access

Introduction

Wear plays a key role in the clinical outcome of total hip replacements (THR). In addition, increased frictional moment can stress the implant interfaces which may lead to high torsional loadings in the intermodular taper junction (fretting) and cup loosening and to the development of noise (squeaking). Against the background of larger head diameters (increased range of motion and decreased risk of dislocation), the friction induced by the joint articulation is of particular interest. As of now, the investigation of friction with the use of relevant joint kinematics and loadings are limited to numerical studies. Experimental approaches use simplified models which do not take into consideration complex activities. Thus, with the aim of this study is the identification of articular frictional moments that consider critical in vivo loading conditions and kinematics as well as the clinical cup inclination, head size and clearance of ceramic-on-ceramic hip bearings.

Materials and Methods

A standard hip simulator (Minibionix 852 with 4 DOF Hip setup, MTS, Eden Prairie, USA) was modified in order to allow for high-precision friction measurements during head-insert articulation in all 6 DOF (MC2.5D-500, AMTI, Boston, USA). Disturbing systemic effects have been minimized by using quasi frictionless aerostatic lateral force compensation (Eitzenberger, Wessobrunn, Germany) and cross talk compensation. Beside the standard protocoll for in vitro wear assessment (ISO 14242-1), more complex profiles from in vivo patient data (Heidelberg Motion Lab and Orthoload database) have been used: normal walking with different walking speeds and patient's weights, stairs up/down and start-stop conditions. All-ceramic bearings (Biolox delta, Ceramtec, Plochingen, Germany) have been orientated in clinically relevant cup inclinations (30, 45, 60 and 75 deg). For each head diameter (28, 36 and 48 mm) n=8 specimens have been devided in two groups: small and large clearance according to the manufacturer's specification. All tests were run at 37°C in diluted bovine serum (20 g/l protein content).


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 34 - 34
1 Mar 2017
Mueller U Lee C Thomsen M Heisel C Kretzer J
Full Access

Introduction

This study was performed to investigate the failure mechanism of one specific hip arthroplasty cup design that has shown a high clinical failure rate. The aim of this study was to identify general design problems of this polyethylene inlay.

Material and Methods

55 consecutive retrievals of a cementless screw ring (Mecron) were collected. In any case a 32 mm ceramic head was used. All implants failed due to aseptic loosening. The follow-up of the implants was 3 to 16 years. We recorded backside wear, fatigue of the polyethylene at the flanges on the outer rim and at the cup opening (32 mm inner diameter). To assess the deformation of the inlay, the smallest and the median diameter of the cup opening were measured using a 3 dimensional coordinate measuring machine (Multisensor, Mahr, Germany).


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 85 - 85
1 Feb 2017
Kretzer J Schroeder M Mueller U Sonntag R Braun S
Full Access

The numbers of anatomic total shoulder joint replacements (ATSR) is increasing during the past years with encouraging clinical results. However, the survivorship of ATSR is lower as compared to total knee and hip replacements. Although the reasons for revision surgery are multifactorial, wear-associated problems like loosening are well-known causes for long-term failure of ATSR. Furthermore there is lack of valid experimental wear tests for ATSR. Therefore the purpose of this study was to define experimental wear testing parameters for ATSR and to perform a wear study comparing ceramic and metallic humeral heads.

Kinetic and kinematic data were adopted from in-vivo loading measurements of the shoulder joint (orthoload.com) and from several clinical studies on shoulder joint kinematics. As activity an ab/adduction motion of 0 to 90° in combination with an ante/retroversion while lifting a load of 2 kg has been chosen. Also a superior-inferior translation of the humeral head has been considered. The wear assessment was performed using a force controlled AMTI joint simulator for 3×106 cycles (Fig. 1) and polyethylene wear has been assed gravimetrically.

The studied ATSR (TuronTM, DJO Surgical, USA) resulted in a polyethylene wear rate of 62.75 ± 1.60 mg/106 cycles in combination with metallic heads. The ceramic heads significantly reduced the wear rate by 26.7 % to 45.99 ± 1.31 mg/106 (p<0.01). The wear scars dimensions were in good agreement to clinical retrievals.

This study is the first that experimentally studied the wear behavior of ATSR based on clinical and biomechanical data under load controlled conditions. In term of wear the analyzed ATSR could clearly benefit from ceramic humeral heads. However, in comparison to experimental wear studies of total knee and hip replacements the wear rate of the studied ATSR was relatively high. Therefore further research may focus on optimized wear conditions of ATSR and the hereby described method may serve as a tool to evaluate a wear optimization process.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 34 - 34
1 May 2016
Beckmann N Gotterbarm T Innmann M Merle C Kretzer J Streit M
Full Access

Introduction

The optimal bearing for hip arthroplasty is still a matter of debate. in younger and more active patients ceramic-on-polyethylene (CoP) bearings are frequently chosen over metal-on-polyethylene (MoP) bearings to reduce wear and increase biocompatibility. However, the fracture risk of ceramic heads is higher than that of metal heads. This can cause serious issue, as ceramic fractures pose a serious complication often necessitating major revision surgery – a complication more frequently seen in ceramic-on-ceramic bearings. To date, there are no long-term data (> 20 years of follow-up) reporting fracture rates of the ceramic femoral heads in CoP bearings.

Patients and Methods

We retrospectively evaluated the clinical and radiographic results of 348 cementless THAs treated with 2nd generation Biolox® Al2O3 Ceramic-on-Polyethylene (CoP) bearings, which had been consecutively implanted between January 1985 and December 1989. At implantation the mean patient age was 57 years. The cohort was subsequently followed for a minimum of 20 years. At the final follow-up 111 patients had died, and 5 were lost to follow-up (Fig. 1). A Kaplan-Meier survivorship analysis was used to estimate the cumulative incidence of ceramic head fractures over the long-term.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 61 - 61
1 May 2016
Mueller U Kretzer J
Full Access

Introduction

Taper corrosion has been identified to be major problem in total hip replacement during the past years. Patients may suffer from adverse local tissue reactions (ALTR) due to corrosion products that are released from modular taper connection. So far, the mechanism that leads to taper corrosion in taper connections is not fully understood. Some retrieval studies tried to correlate implant related design parameters to the incidence and the severeness of taper corrosion. For example Kocagöz et al.[1] have not seen an influence of the taper clearance to taper corrosion. Hothi et al.[2] showed that shorter and rougher tapers increase taper corrosion. One limitation of retrieval studies may be that the analysed tapers are used and may have been altered during in vivo service. Beside the effect of taper corrosion many surgeons are not aware that the tapers may vary among different manufactors. With our study we want to provide taper related data that may be used for comparison and correlation (e.g. retrieval studies). Therefore we aimed to assess and compare geometric and topographic design parameters of currently available hip stem tapers from different manufacturers.

Material

For comparison well established cementless hip stems made of titanium alloy were choosen. All of them have a 12/14-taper. The analysed implants are shown in Fig. 1. As geometrical parameters the taper angle, the opening taper diameter and the taper length were measured using a coordinate measuring machine (CMM) (MarVision MS 222, Mahr, Göttingen, Germany; accuracy: ±2.3 µm). Several topographical parameters (e.g. Ra, Rz, etc.) were determined using a tactil roughness measurement instrument (Perthometer M2, Mahr, Göttingen, Germany; accuracy: 12 nm). Three independent samples of each taper were measured five times.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 60 - 60
1 May 2016
Mueller U Reinders J Kretzer J
Full Access

Introduction

Temporary use of antibiotic-impregnated polymethylmethacrylate (PMMA) bone cement spacers in two-stage revisions is considered to be standard of care for patients with a chronic infection of a joint replacement. Spacers should be wear resistant and load-bearing to avoid prolonged immobilisation of the patient and to reduce morbidity.

Most cement spacers contain barium sulphate or zirconium dioxide as radio-opaque substrate. Both are quite hard materials that may negatively influence the wear behaviour of the spacer.

Calcium carbonate is another radio-opaque substrate with lower hardness potentially increasing the wear resistance of the spacer materials.

The purpose of the study was to compare a prototype PMMA knee spacer (calcium carbonate loaded) with a commercially available spacer (containing barium sulphate) regarding the wear performance and particle release in a knee wear simulator.

Material and Methods

Spacer K (TECRES, Italy) was used as barium sulphate (10%) containing spacer material. A prototype material (Heraeaus Medical, Germany) with 15% calcium carbonate was compared. Both were gentamicin impregnated, ready-made for clinical application (preformed) and consist of a tibial and a femoral component.

Force-controlled simulation was carried out on an AMTI knee simulator. The test parameters were in accordance to ISO 14243–1 with a 50% reduced axial force (partial weight bearing). Tests were carried out at 37 °C in closed chambers filled with calf serum. Tests were run for 500,000 cycles at a frequency of 1 Hz. For wear analysis, gravimetric wear measurements according to ISO 14243–2 and wear particle analysis according to ASTM F1877–05 were performed.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 72 - 72
1 May 2016
Nadorf J Kinkel S Kretzer J
Full Access

INTRODUCTION

Modular knee implants are used to manage large bone defects in revision total knee arthroplasty. These implants are confronted with varying fixation characteristics, changes in load transfer or stiffen the bone. In spite of their current clinical use, the influence of modularity on the biomechanical implant-bone behavior (e.g. implant fixation, flexibility, etc.) still is inadequately investigated.

Aim of this study is to analyze, if the modularity of a tibial implant could change the biomechanical implant fixation behavior and the implant-bone flexibility.

MATERIAL & METHODS

Nine different stem and sleeve combinations of the clinically used tibial revision system Sigma TC3 (DePuy) were compared, each implanted standardized with n=4 in a total of 36 synthetic tibial bones. Four additional un-implanted bones served as reference. Two different cyclic load situations were applied on the implant: 1. Axial torque of ±7Nm around the longitudinal stem axis to determine the rotational implant stability. 2. Varus-valgus-torque of ±3,5Nm to determine the bending behavior of the stem. A high precision optical 3D measurement system allowed simultaneous measuring of spatial micromotions of implant and bone. Based on these micromotions, relative motions at the implant-bone-interface and implant flexibility could be calculated.