header advert
Results 1 - 2 of 2
Results per page:
Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_16 | Pages 9 - 9
1 Oct 2014
Tomita T Futai K Iwamoto K Kii Y Kiyotomo D Murase T Yoshikawa H Sugamoto K
Full Access

Patella resection has been the least controlled element of total knee arthroplasty (TKA). We have developed an intraoperative guide system involving a custom-made surgical template designed on the basis of a three-dimensional computer simulation incorporating computed tomography (CT) data for several years. This time we have applied this intraoperative guide system for the patella resection in TKA. We investigated the accuracy of CT-based patient-specific templating (PST) for patella resection using cadaveric knee joints in vitro.

To plan the corrective patella resection, we attempted to simulate a three-dimensional patella resection with the use of computer models of the patella. From CT images of the patella we obtained three-dimensional surface models of the patella by performing a three-dimensional surface generation of the bone cortex. After the patella resection using CT-based custom-made surgical templating instrumentation, CT scan was performed again and we compared the patella shape in three-dimensional patella bone model reconstructed from pre and after cut from CT data. We compared the accuracy of patella cut using three-dimensional patella bone model reconstructed from pre and after cut from CT data. Statistical analysis was performed using paired t test.

The difference between patella cut with CT-based custom-made surgical templating instrumentation and pre-operative planning were 0.8±1.2mm (medial side) and 0.1±1.4mm (lateral side). More than 60% resulted within 2mm from the pre-operative planning. There were significant differences both in flexion/extension, external/internal rotation and bone cut depth between CT-based custom-made surgical templating instrumentation and conventional instrument.

The results in this study demonstrated the usefulness of CT-based custom-made surgical templating instrumentation for patella resection in TKA.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 98 - 98
1 Mar 2013
Kiyotomo D Sugamoto K Murase T Tomita T Kunugiza Y Kawashima K Futai K Kuramoto K Yamamoto K
Full Access

Introduction

Regarding TKA, patient specific cutting guides (PSCG), which have the same fitting surface with patient's bones or cartilages and uniquely specify the resection plane by fitting guides with bones, have been developed to assist easy, low cost and accurate surgery. They have already been used clinically in Europe and the USA. However little has been reported on clinical positioning accuracy of PSCG. Generally, the methods of making PSCG can be divided into 3 methods; construct 3D bone models with Magnetic Resonance (MR) images, construct 3D bone models with Computed Tomography (CT) images, and the last is to construct 3D bone models with both MR and CT images. In the present study, PSCG were made based on 3D bone models with CT images, examined the positioning accuracy with fresh-frozen cadavers.

Materials and Methods

Two fresh-frozen cadavers with four knees were scanned by CT. Image processing software for 3D design (Mimics Ver. 14, Marialise Inc.) was used to construct 3D bone model by image thresholding. We designed femoral cutting guides and tibial cutting guides by CAD software (NX 5.0, Siemens PLM Software Co.). CT free navigation system (VectorVision Knee, BrainLab, Inc.) was used to measure positioning error. Average absolute value of positioning error for each PSCG was derived.