Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 111 - 111
1 Apr 2019
Verstraete M Conditt M Lieffort D Hazin W Trousdale J Roche M
Full Access

Introduction and Aims

Sensor technology is seeing increased utility in joint arthroplasty, guiding surgeons in assessing the soft tissue envelope intra-operatively (OrthoSensor, FL, USA). Meanwhile, surgical navigation systems are also transforming, with the recent introduction of inertial measurement unit (IMU) based systems no longer requiring optical trackers and infrared camera systems in the operating room (i.e. OrthAlign, CA, USA). Both approaches have now been combined by embedding an IMU into an intercompartmental load sensor. As a result, the alignment of the tibial varus/valgus cut is now measured concurrently with the mediolateral tibiofemoral contact load magnitudes and locations. The wireless sensor is geometrically identical to the tibial insert trial and is placed on the tibial cutting plane after completing the proximal tibial cut. Subsequently, the knee is moved through a simple calibration maneuver, rotating the tibia around the heel. As a result, the sensor provides a direct assessment of the obtained tibial varus/valgus alignment. This study presents the validation of this measurement.

Method

In an in-vitro setting, sensor-based alignment measurements were repeated for several simulated conditions. First, the tibia was cut in near-neutral alignment as guided by a traditional, marker-based surgical navigation system (Stryker, MI, USA). Subsequently, the sensor was inserted and a minimum of five repeated sensor measurements were performed.

Following these measurements, a 3D printed shim was inserted between the sensor and the tibial cutting plane, introducing an additional 2 or 4 degrees of varus or valgus, with the measurements then being repeated. Again, for each condition, a minimum of five sensor measurements were performed. Following completion of the tests, a computed tomography (CT) scan of the tibia was obtained and reconstructed using open source software (3DSlicer).