Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 104 - 104
1 Feb 2020
Zarei M Hamlin B Urish K Anderst W
Full Access

INTRODUCTION

Controversy exists regarding the ability of unicompartmental knee arthroplasty (UKA) to restore native knee kinematics, with some studies suggesting native kinematics are restored in most or all patients after UKA1–3, while others indicate UKA fails to restore native knee kinematics4,5. Previous analysis of UKA articular contact kinematics focused on the replaced compartment2,5, neglecting to assess the effects of the arthroplasty on the contralateral compartment which may provide insight to future pathology such as accelerated degeneration due to overload6 or a change in the location of cartilage contact7. The purpose of this study was to assess the ability of medial UKA to restore native knee kinematics, contact patterns, and lateral compartment dynamic joint space. We hypothesized that medial UKA restores knee kinematics, compartmental contact patterns, and lateral compartment dynamic joint space.

METHODS

Six patients who received fixed-bearing medial UKA consented to participate in this IRB-approved study. All patients (4 M, 2 F; average age 62 ± 6 years) completed pre-surgical (3 weeks before) and post-surgical (7±2 months) testing. Synchronized biplane radiographs were collected at 100 images per second during three repetitions of a chair rise movement (Figure 1). Motion of the femur, tibia, and implants were tracked using an automated volumetric model-based tracking process that matches subject-specific 3D models of the bones and prostheses to the biplane radiographs with sub-millimeter accuracy8. Anatomic coordinate systems were created within the femur and tibia9 and used to calculate tibiofemoral kinematics10. Additional outcome measures included the center of contact in the medial and lateral compartments, and the lateral compartment dynamic joint space (i.e. the distance between subchondral bone surfaces)11. The results of the three movement trials were averaged for each knee in each test session. All outcome measures were interpolated at 5° increments of knee extension (Figure 2). The average differences between knees at corresponding flexion angles were analyzed using paired t-tests with significance set at p < 0.05.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 58 - 58
1 Mar 2017
Urish K Hamlin B Plakseychuk A Levison T Digioia A
Full Access

Introduction

There have been increased concerns with trunnion fretting and corrosion and adverse local tissue reactions (ALTR) in total hip arthroplasty. We report on 11 catastrophic trunnion failures associated with severe ALTR requiring urgent revision arthroplasty.

Methods

We retrospectively reviewed 10 patients with gross trunnion failure (n=11) and an additional 3 patients with impending trunnion failure.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 10 - 10
1 May 2016
MacDonald D Schachtner J Chen A Cates H Klein G Mont M Kraay M Malkani A Lee G Hamlin B Rimnac C Kurtz S
Full Access

Introduction

Highly crosslinked polyethylene (HXLPE) was clinically introduced approximately a decade and a half ago to reduce polyethylene wear rates and subsequent osteolysis. Clinical and radiographic studies have repeatedly shown increased wear resistance, however concerns of rim oxidation and fatigue fracture remain. Although short to intermediate term retrieval studies of these materials are available, the long-term behavior of these materials remains unclear.

Methods

Between 2000 and 2015, 115 1st generation HXLPE acetabular liners implanted for 5 or more years were collected and analyzed as part of an ongoing, multi-institutional orthopaedic implant retrieval program. There were two material cohorts based on thermal processing (annealed (n=45) and remelted (n=70)). Each cohort was stratified into two more cohorts based on implantation time (5 – 10 years and >10 years). For annealed components, the intermediate-term liners (n=30) were implanted on average (±SD) for 7.3 ± 1.7 years while the long-term liners (n=15) were implanted for 11.3 ± 1.8 years. For remelted components, the intermediate-term liners (n=59) were implanted on average (±SD) for 7.2 ± 1.3 years while the long-term liners (n=11) were implanted for 11.3 ± 1.2 years. For each cohort, the predominant revision reasons were loosening, instability, and infection (Figure 1). Short-term liners (in-vivo <5ys) from previous studies were analyzed using the same protocol for use as a reference.

For oxidation analysis, thin slices (∼200 μm) were taken from the superior/inferior axis and subsequently boiled in heptane for 6 hours to remove absorbed lipids that may interfere with the oxidation analysis. 3mm line profiles (in 100μm increments) were taken perpendicular to the surface at each region of interest. Oxidation indices were calculated according to ASTM 2102. Penetration was measured directly using a calibrated micrometer (accuracy=0.001mm).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 129 - 129
1 May 2016
Kurtz S Arnholt C MacDonald D Higgs G Underwood R Chen A Klein G Hamlin B Lee G Mont M Cates H Malkani A Kraay M Rimnac C
Full Access

Introduction

Previous studies of retrieved CoCr alloy femoral heads have identified imprinting of the stem taper surface features onto the interior head bore, leading researchers to hypothesize that stem taper microgrooves may influence taper corrosion. However, little is known about the role of stem taper surface morphology on the magnitude of in vivo corrosion damage. We designed a matched cohort retrieval study to examine this issue.

Methods

From a multi-institutional retrieval collection of over 3,000 THAs, 120 femoral head-stem pairs were analyzed for evidence of fretting and corrosion using a visual scoring technique based on the severity and extent of fretting and corrosion damage observed at the taper. A matched cohort design was used in which 60 CoCr head-stem pairs with a smooth stem taper were matched with 60 CoCr head-stem pairs having a micro-grooved surface, based on implantation time, flexural rigidity, apparent length of taper engagement, and head size. This study was adequately powered to detect a difference of 0.5 in corrosion scores between the two cohorts, with a power of 82% and 95% confidence. Both cohorts included CoCr and Ti-6-4 alloy femoral stems. A high precision roundness machine (Talyrond 585, Taylor Hobson, UK) was used to measure surface morphology and categorize the stem tapers into smooth vs. micro-grooved categories. Fretting and corrosion damage at the head/neck junction was characterized using a modified semi-quantitative adapted from the Goldberg method by three independent observers. This method separated corrosion damage into four visually determined categories: minimal, mild, moderate and severe damage.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_16 | Pages 12 - 12
1 Oct 2014
Smith J Picard F Lonner J Hamlin B Rowe P Riches P Deakin A
Full Access

Knee osteoarthritis results in pain and functional limitations. In cases where the arthritis is limited to one compartment of the knee joint then a unicondylar knee arthroplasty (UKA) is successful, bone preserving option. UKA have been shown to result in superior clinical and functional outcomes compared to TKA patients. However, utilisation of this procedure has been limited due primarily to the high revision rates reported in joint registers. Robotic assisted devices have recently been introduced to the market for use in UKA. They have limited follow up periods but have reported good implant accuracy when compared to the pre-operative planned implant placement.

UKA was completed on 25 cadaver specimens (hip to toe) using an image-free approach with infrared optical navigation system with a hand held robotically assisted cutting tool. Therefore, no CT scan or MRI was required. The surface of the condylar was mapped intra operatively using a probe to record the 3 dimensional surface of the area of the knee joint to be resurfaced. Based on this data the size and orientation of the implant was planned. The user was able to rotate and translate the implant in all three planes. The system also displays the predicted gap balance graph through flexion as well as the predicted contact points on the femoral and tibial component through flexion. The required bone was removed using a bur. The depth of the cut was controlled by the robotically controlled freehand sculpting tool.

Four users (3 consultant orthopaedic surgeon and a post-doctoral research associate) who had been trained on the system prior to the cadaveric study carried out the procedures. The aim of this study was to quantify the differences between the ‘planned’ and ‘achieved’ cuts. A 3D image of the ‘actual’ implant position was overlaid on the ‘planned’ implant image. The errors between the ‘actual’ and the ‘planned’ implant placement were calculated in three planes and the three rotations. The maximum femoral RMS angular error was 2.34°. The maximum femoral RMS translational error across all directions was up to 1.61mm. The maximum tibial RMS angular error was 2.60°. The maximum tibial RMS translational error across all directions was up to 1.67mm.

In conclusion, the results of this cadaver study reported low RMS errors in implant position placement compared to the plan. The results were comparable with those published from clinical studies investigating other robotic orthopaedic devices. Therefore, the freehand sculpting tool was shown to be a reliable tool for cutting bone in UKA and the system allows the surgeon to plan the placement of the implant intra operatively and then execute the plan successfully.