Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 14 - 14
23 Jun 2023
Hooper G Thompson D Frampton C Lash N Sharr J Fulkner D Gilchrist N
Full Access

Femoral stem design affects periprosthetic bone mineral density (BMD), which may impact long term survival of cementless implants in total hip arthroplasty (THA). The aim of this study was to examine proximal femoral BMD in three morphologically different uncemented femoral stems designs to investigate whether one particular design resulted in improved preservation of BMD.

119 patients were randomized to receive either a proximally coated dual taper wedge stem, a proximally coated anatomic stem or a fully coated collarless triple tapered stem. Dual energy x-ray absorptiometry scans (Lunar iDXA, GE Healthcare, Madison, WI) assessed BMD across the seven Gruen zones pre-operatively, and post-operatively at 6-weeks, 1-year, and 2-years comparing the unoperated contralateral femur as a control.

BMD increased in zones one (2.5%), two (17.1%), three (13.0%), five (10%) and six (17.9%) for all stems. Greater preservation of BMD was measured on the lateral cortex (zone 2) for both the dual taper wedge and anatomic stems (p = 0.019). The dual taper wedge stem also demonstrated preservation of BMD in the medial calcar (zone 7) whilst the anatomic and triple taper stem declined in this region, however this was not statistically significant (p = 0.059). BMD decreased on average by 2.1% in the mid-diaphysis region, distal to the stem tip (zone 4) for all implants. All stems performed equivalently at final follow-up in all patient reported outcome measures.

Implant design was shown to impact BMD changes. All stems preserved bone at the metaphyseal level in early follow-up, which theoretically indicates favorable implant fixation. There was no significant stress shielding observed, however longer follow-up is required to elucidate the impact of this finding on implant survivorship.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 10 - 10
1 Aug 2018
Hooper G Gilchrist N Maxwell R Frampton C
Full Access

Stress shielding has been a well-recognised problem with uncemented femoral components resulting in proximal bone loss and dysfunction, but less attention has been paid to the preservation of acetabular bone stock. Uncemented acetabular components often demonstrate reduced bone density on plain radiographs in the mid-portion of the cup (zone 2), which may be due to the rigidity of the outer shell. This study compares the change in bone density around three different cups with varying moduli of elasticity at a minimum of 2 years. Our hypothesis was that less rigid cups would be associated with improved bone density and less stress shielding.

This prospective randomised controlled trial compared the bone mineral content (BMC) adjacent to three different cups with marked differences in stiffness. Cup A was an all titanium shell, cup B was a titanium coated all polyethylene implant and cup C was a tantalum backed shell. All articulations used a 32mm ceramic femoral head. Cup B used polyethylene modified by treatment with vitamin E whereas cups A and C used a liner made of irradiated cross linked polyethylene.

Five regions of interest (ROI) were established adjacent to the cup, regions 2, 3 and 4 where similar to the DeLee and Charnley regions 1, 2 and 3. Bone density was measured using IDXA preoperatively, postoperatively, 6 months, 1 and 2 years and compared for each ROI and implant.

Precision measurements showed significant reliability. All areas showed a reduction in BMC following insertion of the acetabular cup. Bone loss was less in ROI 1 and 4 in the area of rim fit for all cups and the maximal bone loss was seen in ROI 2 and 3 at the dome of the cup. The more elastic cup (Cup B) produced the least bone loss in this area (p<0.05). Cup C produced the largest bone loss at ROI 2 (40%) which continued increasing at 2 years.

Cup stiffness is related to bone loss adjacent to the acetabulum, presumably due to a similar process of stress shielding as seen in the femur. All cups produced similar changes at the periphery of the cup but the more elastic cup retained bone density beneath the cup which continued past 2 years. This improvement in bone quality is likely to be associated with better acetabular bone stock into the future and more reliable long term cup fixation.