Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 356 - 356
1 Sep 2012
Philippot R Boyer B Schneider L Farizon F
Full Access

Dual mobility significantly reduces the risk of prosthetic instability. This mechanical complication occurs when the prosthetic head moves out from the retentive polyethylene liner, such phenomenon called intra-prosthetic dislocation reports a ten-year incidence of 2% in the literature.

We prospectively analysed all intra-prosthetic dislocations having occurred since 1985 in our department in order to investigate patient- and implant-related risk factors.

91 intra-prosthetic dislocations occurred with NOVAE (SERF) cups in 85 patients of mean age 50.7 years. Intra-prosthetic dislocation occurred after a mean period of 8.8 years.

A PRO (SERF) stem was implanted in 56 cases and a PF (SERF) stem in 35. The stems were different from one another in their neck diameter and material: 13 mm titanium neck and 16 mm stainless steel neck respectively.

When taking both prosthetic features into account, no significant difference could be established regarding the time between implantation and dislocation.

Comparison between these two prosthetic features was performed by means of two continuous homogeneous series which included 240 patients implanted with PF stems and 382 patients with PRO stems. At a mean 15-year follow-up, the two series reported a non-statistically different intra-prosthetic dislocation rate of 4%.

In both series, young age and large diameter cups were considered predisposing factors for intra-prosthetic dislocation.

Therefore, unlike suggested by several authors, prosthetic neck material and diameter do not appear as the main predictors for intra-prosthetic dislocation which is highly promoted by patient-related features.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 355 - 355
1 Sep 2012
Philippot R Camilleri JP Boyer B Farizon F
Full Access

The concept of stainless steel dual mobility cups in total hip arthroplasty has demonstrated very low long-term instability rates and a 98% survival rate after 12 years. We systematically implanted titanium alloy acetabular cups during a one year period. The purpose of our retrospective study was to report the 18-year clinical outcome data in a homogeneous and continuous series of 103 primary total hip replacements after implantation of a cementless titanium cup.

All patients were implanted with NOVAE Ti (SERF) cups made of titanium alloy combined with a retentive polyethylene liner and a 22.2 mm cobalt chrome prosthetic head. Mean patient age at the time of surgery was 53 years. All patients were clinically and radiographically evaluated.

The overall 18-year actuarial cup survival rate with a 95% confidence interval was 87.4%. At last follow-up, there was no evidence of implant instability whereas acetabular aseptic loosening was reported in one case and high wear of the retentive liner in 9.

The results of this investigation confirmed the long-term stability of dual-mobility implants. The main limitation of this system was early wear of the polyethylene liner in contact with the titanium metal back and reaction with third body along with loss of liner retentivity. In our study, titanium demonstrated favourable osteointegration properties but poor tribologic characteristics, therefore suggesting its interest at the bone-cup interface only.