Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 102 - 102
1 Feb 2020
DiGeorgio C Yegres J VanDeven J Stroud N Cheung E Grey S Yoo J Deshmukh R Crosby L Roche C
Full Access

Introduction

Little guidance exists regarding the minimum screw length and number necessary to achieve fixation with reverse shoulder arthroplasty (rTSA). The goal of this study is to quantify the pre- and post-cyclic baseplate displacements associated with two baseplate designs of different sizes using multiple screw lengths and numbers in a low density polyurethane bone substitute model.

Methods

The test was conducted according to ASTM F 2028–17. The baseplate displacements of standard and small reverse shoulder constructs (Equinoxe, Exactech, Inc.) were quantified in a 15pcf polyurethane block (Pacific Research, Inc.) before and after cyclic testing with an applied load of 750N for 10,000 cycles. Baseplates were constructed using 2 or 4 screws with 3 different poly-axial locking compression screw lengths: 4.5×18mm, 4.5×30mm, and 4.5×46mm. Five of each configuration were tested for a total of 30 specimens for each baseplate. A two-tailed, unpaired student's t-test (p<0.05) compared baseplate displacements before and after cyclic loading in both the superior-inferior (S/I) and anterior-posterior (A/P) directions. The standard and small results were then compared.