header advert
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 113 - 113
1 Apr 2019
Verstraete M Conditt M Wright T Zuckerman J Youderian A Parsons I Jones R Decerce J Goodchild G Greene A Roche C
Full Access

Introduction & Aims

Over the last decade, sensor technology has proven its benefits in total knee arthroplasty, allowing the quantitative assessment of tension in the medial and lateral compartment of the tibiofemoral joint through the range of motion (VERASENSE, OrthoSensor Inc, FL, USA). In reversal total shoulder arthroplasty, it is well understood that stability is primarily controlled by the active and passive structures surrounding the articulating surfaces. At current, assessing the tension in these stabilizing structures remains however highly subjective and relies on the surgeons’ feel and experience. In an attempt to quantify this feel and address instability as a dominant cause for revision surgery, this paper introduces an intra-articular load sensor for reverse total shoulder arthroplasty (RTSA).

Method

Using the capacitive load sensing technology embedded in instrumented tibial trays, a wireless, instrumented humeral trial has been developed. The wireless communication enables real-time display of the three-dimensional load vector and load magnitude in the glenohumeral joint during component trialing in RTSA. In an in-vitro setting, this sensor was used in two reverse total shoulder arthroplasties. The resulting load vectors were captured through the range of motion while the joint was artificially tightened by adding shims to the humeral tray.