Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 93 - 93
1 Feb 2020
Cipparone N Robinson M Chen J Muir J Shah R
Full Access

Acetabular cup positioning remains a real challenge and component malpositioning after total hip arthroplasty (THA) can lead to increased rates of dislocation and wear. It is a common cause for revision THA. A novel 3D imageless mini-optical navigation system was used during THA to provide accurate, intraoperative, real-time, and non-fluoroscopic data including component positioning to the surgeon. This retrospective comparative single surgeon and single approach study examined acetabular component positioning between traditional mini-posterolateral THA and mini-posterolateral THA using the 3D mini-optical navigation system. A retrospective chart review was conducted of 157 consecutive (78 3D mini-optical navigation and 79 traditional non-navigation methods) THAs performed by the senior author using a mini-posterolateral approach at an ambulatory surgery center and hospital setting. Two independent reviewers analyzed postoperative radiographs in a standardized fashion to measure acetabular component positioning. Demographic, clinical, surgical, and radiographic data were analyzed.

These groups were found to have no statistical difference in age, gender, and BMI (Table I). There was no difference between groups in acetabular components in the Lewinnek safe zone, 31.2% vs 26.6% (p = 0.53). Cup anteversion within the safe zone did not differ, 35.1% vs 40.5% (p = 0.48); while cup inclination within the safe zone differed, with more in the navigation group, 77.9% vs 51.9% (p < 0.01). Change in leg length was significantly different with the navigation group's leg length at 1.9 ± 6.3, less than the traditional at 5.4 ± 7.0 (p < 0.01). There was no difference in mean change in offset between groups (4.5 ± 5.9 vs 6.2 ± 7.9, p = 0.12); navigation, traditional) (Table II). The 3D mini-optical navigation group did have significantly longer operative time (98.4 ± 17.5 vs 89.3 ± 15.5 p < 0.01). Use of the novel 3D Mini-optical Navigation System significantly improved cup inclination compared to traditional methods while increasing operative time.

For any figures or tables, please contact the authors directly.