header advert
Results 1 - 4 of 4
Results per page:
Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 20 - 20
1 Nov 2021
Shimmin A Dhawan R Madurawe C Pierrepont J Baré J
Full Access

Adverse spinopelvic mobility (SPM) has been shown to increase risk of dislocation of primary total hip arthroplasty (THA). In patients undergoing THA, prevalence of adverse SPM has been shown to be as high as 41%. Stiff lumbar spine, large posterior standing pelvic tilt and severe sagittal spinal deformity have been identified as risk factors for increased hip instability. Dislocation rates for dual mobility articulations have been reported to be 0% to 1.1%. The aim of this study was to determine the early survivorship from the Australian National Joint Replacement Registry (AOANJRR) of patients with adverse SPM who received a dual mobility articulation.

A multicentre study was performed using data from 229 patients undergoing primary THA, enrolled consecutively. All the patients who had one or more adverse spine or pelvic mobility parameters had a dual mobility articulation inserted at the time of their surgery. Average age was 76 (22 to 93) years and 63% were female. At a mean of 2.1 (1 – 3.3) years post-op, the AOANJRR was analysed for follow-up. Reasons for revision and types of revision were identified.

The AOANJRR reported two revisions. One due to infection and the second due to femoral component loosening. No revisions for dislocation were reported. One patient died with the prosthesis in situ. Kaplan Meier survival was 99.3% (CI 98.3% − 100%) at 2 years.

DM bearings reduce the risk of dislocation of primary THA in patients with adverse spine and pelvic mobility.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 184 - 191
1 Feb 2017
Pierrepont J Hawdon G Miles BP Connor BO Baré J Walter LR Marel E Solomon M McMahon S Shimmin AJ

Aims

The pelvis rotates in the sagittal plane during daily activities. These rotations have a direct effect on the functional orientation of the acetabulum. The aim of this study was to quantify changes in pelvic tilt between different functional positions.

Patients and Methods

Pre-operatively, pelvic tilt was measured in 1517 patients undergoing total hip arthroplasty (THA) in three functional positions – supine, standing and flexed seated (the moment when patients initiate rising from a seated position). Supine pelvic tilt was measured from CT scans, standing and flexed seated pelvic tilts were measured from standardised lateral radiographs. Anterior pelvic tilt was assigned a positive value.


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1342 - 1346
1 Oct 2016
Spencer-Gardner L Pierrepont J Topham M Baré J McMahon S Shimmin AJ

Aims

Accurate placement of the acetabular component during total hip arthroplasty (THA) is an important factor in the success of the procedure. However, the reported accuracy varies greatly and is dependent upon whether free hand or navigated techniques are used. The aim of this study was to assess the accuracy of an instrument system that incorporates 3D printed, patient-specific guides designed to optimise the placement of the acetabular component.

Patients and Methods

A total of 100 consecutive patients were prospectively enrolled and the accuracy of placement of the acetabular component was measured using post-operative CT scans.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 160 - 165
1 Feb 2013
McDonnell SM Boyce G Baré J Young D Shimmin AJ

Noise generation has been reported with ceramic-on-ceramic articulations in total hip replacement (THR). This study evaluated 208 consecutive Delta Motion THRs at a mean follow-up of 21 months (12 to 35). There were 141 women and 67 men with a mean age of 59 years (22 to 84). Patients were reviewed clinically and radiologically, and the incidence of noise was determined using a newly described assessment method. Noise production was examined against range of movement, ligamentous laxity, patient-reported outcome scores, activity level and orientation of the acetabular component. There were 143 silent hips (69%), 22 (11%) with noises other than squeaking, 17 (8%) with unreproducible squeaking and 26 (13%) with reproducible squeaking. Hips with reproducible squeaking had a greater mean range of movement (p < 0.001) and mean ligament laxity (p = 0.004), smaller median head size (p = 0.01) and decreased mean acetabular component inclination (p = 0.02) and anteversion angle (p = 0.02) compared with the other groups. There was no relationship between squeaking and age (p = 0.13), height (p = 0.263), weight (p = 0.333), body mass index (p = 0.643), gender (p = 0.07) or patient outcome score (p = 0.422). There were no revisions during follow-up. Despite the surprisingly high incidence of squeaking, all patients remain satisfied with their hip replacement.

Cite this article: Bone Joint J 2013;95-B:160–5.