The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 461 - 467
1 Mar 2010
Wik TS Østbyhaug PO Klaksvik J Aamodt A

The cortical strains on the femoral neck and proximal femur were measured before and after implantation of a resurfacing femoral component in 13 femurs from human cadavers. These were loaded into a hip simulator for single-leg stance and stair-climbing. After resurfacing, the mean tensile strain increased by 15% (95% confidence interval (CI) 6 to 24, p = 0.003) on the lateral femoral neck and the mean compressive strain increased by 11% (95% CI 5 to 17, p = 0.002) on the medial femoral neck during stimulation of single-leg stance. On the proximal femur the deformation pattern remained similar to that of the unoperated femurs.

The small increase of strains in the neck area alone would probably not be sufficient to cause fracture of the neck However, with patient-related and surgical factors these strain changes may contribute to the risk of early periprosthetic fracture.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 676 - 682
1 May 2009
Østbyhaug PO Klaksvik J Romundstad P Aamodt A

Hydroxyapatite-coated standard anatomical and customised femoral stems are designed to transmit load to the metaphyseal part of the proximal femur in order to avoid stress shielding and to reduce resorption of bone. In a randomised in vitro study, we compared the changes in the pattern of cortical strain after the insertion of hydroxyapatite-coated standard anatomical and customised stems in 12 pairs of human cadaver femora. A hip simulator reproduced the physiological loads on the proximal femur in single-leg stance and stair-climbing. The cortical strains were measured before and after the insertion of the stems.

Significantly higher strain shielding was seen in Gruen zones 7, 6, 5, 3 and 2 after the insertion of the anatomical stem compared with the customised stem. For the anatomical stem, the hoop strains on the femur also indicated that the load was transferred to the cortical bone at the lower metaphyseal or upper diaphyseal part of the proximal femur.

The customised stem induced a strain pattern more similar to that of the intact femur than the standard, anatomical stem.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 153 - 153
1 Mar 2008
Aamodt A Østbyhaug PO Klaksvik J Benum P
Full Access

In the past it has been widely accepted that bone remodelling of the proximal femur after cementless total hip replacement is a result of the altered mechanical environment. Usually, there is are distribution of the stresses in the bone, and subsequently bone mass, from the metaphysis to the proximal part of the diaphysis. The design rationale for some cementless stems is to transmit load to the proximal femur and thus to preserve the bone mineral content in this area. The aim of the present study was to investigate the relationship between postoperative strain shielding of the proximal femur and the bone remodelling after insertion of two different cementless femoral stems.

Experimental study: Twelve pairs of human cadaveric femurs were instrumented with strain gauge rosettes in Gruen zones2 to 7 and the cortical strains were measured during simulation of one leg stance before and after insertion of a custom stem (Unique, SCP) or an anatomic stem (ABG, Stryker-Howmedica).

Clinical study: In a prospective, randomized study including 80 patients, the same types of stems were inserted and the bone mineral density (BMD) was measured during the first two years postoperatively using DEXA. Then, the pattern of remodelling was compared with the gradient of strain shielding in each of the Gruen zones in the frontal plane.

In Gruen zone 7 the relative cortical strain shielding was45% in the femurs with a custom stem and 87% in the femurs with an anatomic stem. In zone 6 the corresponding figures were 2% and 38%, in zone 5 0% and15% and in zone 3 0% and 20%. The DEXA measurements showed a decrease in BMD in zone 7 of 22% and 23% for the two stems, respectively. In the other zones the bone loss was smaller and there was no difference between the stems.

In the proximal zones there was a highly significant difference in strain shielding between femurs receiving a customor an anatomic stem. However, there was no difference in the pattern of bone remodelling. The bone remodelling around these two stems does not seem to mirror the gradient of strain shielding.