Viscosupplementation is the current treatment modality for early stage arthritis and in some cases for delaying joint replacement procedures. Rheological properties similar to that of synovial fluid and high molecular weight have been recognized as the determining factors in hyaluronic acids (HA) therapeutic and analgesic value (
There is currently much interest in the wear of metal-on-metal THRs and potential concerns about elevated metal ion levels. Generally, wear of metal-on-metal THR’s has been low in simulator studies. Slightly higher and more variable wear has been found clinically. Variations in surgical approach, technique and fixation method may influence the level of force applied across the prosthesis during gait. It is hypothesised that increased joint tensioning may increase loading of THR’s during the swing-phase; leading to elevated wear and friction due to depleted fluid film lubrication. This study aimed to assess the effect of swing-phase load on the friction, lubrication and wear of metal-on-metal THR’s. Cobalt-chrome 28mm metal-on-metal THR’s were tested in a physiological hip simulator, loading was modified to provide; (1) ISO swing-phase load (280N, as per ISO 14242-1) and (2) low swing-phase load (<
100N). Friction testing was conducted using a pendulum friction simulator, with 280N and 100N swing-phase loads. Theoretical lubrication modelling was carried out using elastohydrodynamic lubrication theory. The overall mean volumetric wear rates was 10-times greater in THR’s tested with an ISO swing-phase load in comparison to THR’s tested with low swing-phase loads (0.58±0.49 compared to 0.06±0.039mm3/million cycles). The friction factors were 0.129 and 0.173 respectively under low and ISO swing-phase conditions. A decrease in the predicted lubricant film thickness when the swing-phase load was increased was observed; at the start of stance phase this was 0.12microns and 0.07microns under low and ISO swing-phase conditions respectively. The results demonstrate that the performance of metal-on-metal THR’s is highly dependent on swing-phase load conditions. It is postulated that fixation method and surgical technique can affect the swing-phase load. This study has demonstrated that over-tensioning of the tissues may also accelerate wear. These observations may explain some of the variations reported clinically.
Low positive swing phase load (<
100N) Positive swing phase load (300N, as per standard ISO 14242–1) Negative swing phase load, leading to microseparation and joint laxity. All tests were carried out in 25% (v/v) new-born bovine serum, with gravimetric wear measurements completed every million cycles.
Ultra high molecular weight polyethylene (UHMWPE) wear debris induced osteolysis is a major cause of long term failure of total hip replacements. Particles in the 0.1–1.0_m size range are believed to have greater osteolytic potential than larger wear debris. Crosslinked polyethylenes have been shown to have improved wear resistance compared to non-crosslinked materials on smooth counterfaces, however wear debris from cross-linked UHMWPE has been shown to be smaller than that produced from non-crosslinked materials. The aim of this study was to compare the wear, wear debris and biological activity of non-crosslinked and crosslinked polyethylenes when worn against smooth and scratched counterfaces.
Following total hip replacement surgery, fluroscopy studies have shown that a mean separation of 2 mm can occur between CoCr femoral heads and UHMWPE acetabular cups during the swing phase of gait [ A physiological hip simulator was used, loads and motions were applied to approximate in vivo conditions. The alumina ceramic heads and polyethylene cups were 28 mm in diameter and were tested for 5 million cycles in 25% new born calf serum at 1 Hertz. Microseparation was achieved by displacing the femoral head inferiorly during swing phase, where the head contacted the inferior cup rim and was laterally displaced. On heel strike the head contacted the superior cup rim prior to relocation. The volumetric wear of the polyethylene inserts was approximately four times less under microseparation conditions (5.6 ± 5.3 mm3 per million cycles), in comparison to standard conditions (25.6 ± 5.3 mm3 per million cycles). Deformation of the cup rim was observed, but some of this was attributed to creep. It is postulated that this reduction in wear was due to the separation of the components in swing phase improving the entrapment of lubricant, hence reducing wear via a squeeze film lubrication mechanism. In conclusion, surgical procedures that produce a small and controlled amount of joint laxity and microseparation may lead to a reduction in wear of the polyethylene acetabular cups.
Proponents of the biological theory of aseptic loosening have in recent years tended to concentrate on the production and distribution of particulate ultra-high-molecular-weight polyethylene (UHMWPE) debris around the potential joint space. However, mechanical loading of cemented implants with the differing elastic moduli of metal stems, polymethylmethacrylate (PMMA) cement and bone can result in relative micromotion, implying the potential for production of metal and PMMA particles from the stem-cement interface by fretting wear. In order to investigate the production and biological reactivity of debris from this interface, PMMA and metal particulate debris was produced by sliding wear of PMMA pins containing barium sulphate and zirconium dioxide against a Vaquasheened stainless steel counterface. This debris was characterised by SEM, energy-dispersive analysis by X-ray (EDAX) and image analysis, then added to cell cultures of a human monocytic cell line, U937, and stimulation of pro-osteolytic cytokines measured by ELISA. Large quantities of PMMA cement debris were generated by the sliding wear of PMMA pins against Vaquasheened stainless steel plates in the method developed for this study. Both cements stimulated the release of pro-osteolytic TNFα from the U937 monocytic cell line, in a dose-dependent fashion. There was a trend towards greater TNFα release with Palacos cement than CMW cement at the same dose. Palacos particles also caused significant release of IL-6, another pro-osteolytic cytokine, while CMW did not. The particulate cement debris produced did not stimulate the release of GM-CSF or IL1β from the U937 cells. These results may explain the cytokine pathway responsible for bone resorption caused by particulate PMMA debris. Radio-opaque additives are of value in surgical practice and clinical studies to quantify the relevance of these in vitro findings are required before the use of cement containing radio-opacifier is constrained.
A physiological hip simulator was used, loads and motions were applied to approximate
When considering all the subjects, no correlation was found between the response of their cells to polyethylene particles and LPS stimulation. However the cells of four subjects gave a much higher response to LPS than the rest and when these where excluded the correlation between the response to LPS and PE particles was significant with an R2 value of 0.9076.
TNFα
levels Particle stimulation
LPS stimulation
Control
0.043–0.059
0.097–0.208
Patients
0–1.1
0.03–17.693
We have examined 26 retrieved, failed titanium-alloy femoral stems. The clinical details, radiological appearances and the histology of the surrounding soft tissues in each patient were also investigated. The stems were predominantly of the flanged design and had a characteristic pattern of wear. A review of the radiographs showed a series of changes, progressive with time. The first was lateral debonding with subsidence of the stem. This was followed by calcar resorption and fragmentation or fracture of the cement. Finally, osteolysis was seen, starting with a radiolucency at the cement-bone interface and progressing to endosteal cavitation. Three histological appearances were noted: granulomatous, necrobiotic and necrotic. We suggest that an unknown factor, possibly related to the design of the stem, caused it to move early. After this, micromovement at the cement-stem interface led to the generation of particulate debris and fracture of the cement. A soft-tissue reaction to the debris resulted in osteolysis and failure of fixation of the prostheses.
We compared and quantified the modes of failure and patterns of wear of 11 Mittelmeier and 11 Ceraver-Ostal retrieved alumina-alumina hip prostheses with reference to the corresponding clinical and radiological histories. Macroscopic wear was assessed using a three-dimensional co-ordinate measuring machine. Talysurf contacting profilometry was used to measure surface roughness on a microscopic scale and SEM to determine mechanisms of wear at the submicron level. The components were classified into one of three categories of wear: low (no visible/measurable wear), stripe (elliptical wear stripe on the heads and larger worn areas on the cups) and severe (macroscopic wear, large volumes of material lost). Overall, the volumetric wear of the alumina-alumina prostheses was substantially less than the widely used metal and ceramic-on-polyethylene combinations. By identifying and eliminating the factors which accelerate wear, it is expected that the lifetime of these devices can be further increased.
We examined stainless-steel, cobalt-chrome, titanium and alumina and zirconia ceramic femoral heads retrieved at revision surgery. All the heads had articulated against ultra-high-molecular-weight-polyethylene (UHMWPE) acetabular cups. We studied the simulation of third-body damage and the wear of UHMWPE against the various materials used for the heads. The surfaces of the retrieved heads were analysed using a two-dimensional contacting profilometer. Third-body damage was characterised by the mean height of the scratches above the mean line (Rpm). The alumina ceramic and zirconia ceramic retrieved heads were found to have significantly less damage. In laboratory studies the ceramics were also more resistant to simulated third-body damage than the metal alloys. We studied the wear of UHMWPE against the damaged counterfaces in simple configuration tests. The damaged ceramics produced less polyethylene wear than the damaged metal counterfaces. The wear factor of UHMWPE against the damaged materials was dependent on the amount of damage to the counterface (Rp). Our study has shown the benefit of using the harder and more damage-resistant ceramic materials for femoral heads.
Ultra-high-molecular-weight polyethylene (UHMWPE) components for total joint replacement generate wear particles which cause adverse biological tissue reactions leading to osteolysis and loosening. Sterilisation of UHMWPE components by gamma irradiation in air causes chain scissions which initiate a long-term oxidative process that degrades the chemical and mechanical properties of the polyethylene. Using a tri-pin-on-disc tribometer we studied the effect of ageing for ten years after gamma irradiation in air on the volumetric wear, particle size distribution and the number of particles produced by UHMWPE when sliding against a stainless-steel counterface. The aged and irradiated material produced six times more volumetric wear and 34 times more wear particles per unit load per unit sliding distance than non-sterilised UHMWPE. Our findings indicate that oxidative degradation of polyethylene after gamma irradiation in air with ageing produces more wear.