Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_5 | Pages 22 - 22
23 Apr 2024
Laufer A Frommer A Gosheger G Toporowski G Rölfing JD Antfang C Roedl R Vogt B
Full Access

Introduction

Coronal malalignment and leg length discrepancies (LLD) are frequently associated. Temporary hemiepiphysiodesis (tHED) is commonly employed for the correction of limb malalignment in skeletally immature patients. For treatment of LLD greater than 2 cm, lengthening with intramedullary legnthening nails is a safe and reliable technique. However, the combined application of these approaches in skeletally immature patients has not yet been investigated.

Materials & Methods

Retrospective radiological and clinical analysis of 25 patients (14 females, 11 males) who underwent intramedullary femoral lengthening with an antegrade PRECICE® lengthening nail as well as tHED of the distal femur and / or proximal tibia between 2014 and 2019. tHED was conducted by implantation of flexible staples (FlexTack™) either prior (n = 11), simultaneously (n = 10), or subsequently (n = 4) to femoral lengthening. The mean follow-up period was 3.7 years (±1.4).


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims

This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported.

Methods

This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed.