Bone & Joint Open
Vol. 1, Issue 10 | Pages 653 - 662
20 Oct 2020
Rahman L Ibrahim MS Somerville L Teeter MG Naudie DD McCalden RW

Aims

To compare the in vivo long-term fixation achieved by two acetabular components with different porous ingrowth surfaces using radiostereometric analysis (RSA).

Methods

This was a minimum ten-year follow-up of a prospective randomized trial of 62 hips with two different porous ingrowth acetabular components. RSA exams had previously been acquired through two years of follow-up. Patients returned for RSA examination at a minimum of ten years. In addition, radiological appearance of these acetabular components was analyzed, and patient-reported outcome measures (PROMs) obtained.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 72 - 72
1 Oct 2020
Howard JL Williams HA Lanting BA Teeter MG
Full Access

Background

In recent years, the use of modern cementless implants in total knee arthroplasty has been increasing in popularity. These implants take advantage of new technologies such as additive manufacturing and potentially provide a promising alternative to cemented implant designs. The purpose of this study was to compare implant migration and tibiofemoral contact kinematics of a cementless primary total knee arthroplasty (TKA) implanted using either a gap balancing (GB) or measured resection (MR) surgical technique.

Methods

Thirty-nine patients undergoing unilateral TKA were recruited and assigned based on surgeon referral to an arthroplasty surgeon who utilizes either a GB (n = 19) or a MR (n = 20) surgical technique. All patients received an identical fixed-bearing, cruciate-retaining beaded peri-apatite coated cementless femoral component and a pegged highly porous cementless tibial baseplate with a condylar stabilizing tibial insert. Patients underwent a baseline radiostereometric analysis (RSA) exam at two weeks post-operation, with follow-up visits at six weeks, three months, six months, and one year post-operation. Migration including maximum total point motion (MTPM) of the femoral and tibial components was calculated over time. At the one year visit patients also underwent a kinematic exam using the RSA system.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_11 | Pages 57 - 57
1 Oct 2019
Broberg JS Teeter MG Lanting B Vasarhelyi EM Howard JL Yuan X Naudie DDR
Full Access

Introduction

Surgeons performing a total knee replacement (TKR) have two techniques to assist them achieve proper bone resections and ligament tension – gap balancing (GB) and measured resection (MR). GB relies on balancing ligaments prior to bony resections, whereas bony resections are made based on anatomical landmarks in MR. Many studies have been done to compare the implant migration and kinematics between the two techniques, but the results have been varied. However, these studies have not been done on modern anatomically designed prostheses using radiostereometric analysis (RSA). Anatomical designs attempt to mimic the normal knee joint structure to return more natural kinematics to the joint, with emphasis on eliminating both paradoxical anterior motion and reduced posterior femoral rollback. Given the major design differences between anatomical and non-anatomical prostheses, it is important to investigate whether one surgical technique may have advantages another. We hypothesize that there would be no difference between GB and MR techniques in implant migration, but that GB might provide better knee kinematics.

Methods

Patients were recruited to receive an anatomically designed prosthesis and randomized to groups where the GB or MR technique is used. For all patients in the study, RSA images were acquired at a 2 week baseline, as well as at 6 weeks, 3 months, and 6 months post-operatively. These images were used to collect the maximum total point motion (MTPM) of the tibial and femoral implant components relative to the bone using a model-based RSA software. A series of RSA images were also acquired at 3-months post-operatively at different knee flexion angles, ranging in 20° increments from 0° to 100°. Model-based RSA software was used to obtain the 3D positions and orientations of the femoral and tibial components, which were used to obtain the anterior-posterior (AP) contact locations for each condyle.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 52 - 52
1 Oct 2018
Naudie DD Broberg JS Howard JL Vasarhelyi EM Yuan X McCalden RW Teeter MG
Full Access

Introduction

The effectiveness of patient specific instrumentation (PSI) to perform total knee arthroplasty (TKA) remains controversial. Multiple studies have been published that reveal conflicting results on the effectiveness of PSI, but no study has analyzed the contact kinematics within knee joints replaced with the use of PSI. Since a departure from normal kinematics can lead to eccentric loading, premature wear, and component loosening, studying the kinematics in patients who have undergone TKA with PSI can provide valuable insight on the ability of PSI to improve functionality and increase longevity. The goal of the present study was to compare femoral and tibial component migration (predictive of long-term loosening and revision) and contact kinematics following TKA using conventional instruments (CI) and PSI based surgical techniques.

Methods

The study was designed as a prospective, randomized controlled trial of 50 patients, with 25 patients each in the PSI and CI groups, powered for radiostereometric analysis (RSA). Patients in the PSI group received an MRI and standing 3-foot x-rays to construct patient-specific cut-through surgical guides for the femur and tibia with a mechanical limb alignment. All patients received the same posterior-stabilized implant with marker beads inserted in the bone around the implants to enable RSA imaging. Patients underwent supine RSA exams at multiple time points (two and six weeks, three and six months, and one and two years). At 2 years post-op, a series of RSA radiographs were acquired at different knee flexion angles, ranging in 20° increments from 0° to 120°, to measure the tibiofemoral contact kinematics. Migrations of the femoral and tibial components were calculated using model-based RSA software. Kinematics were measured for each condyle for magnitude of excursion, contact location, and stability.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 10 - 10
1 Oct 2018
Howard JL Aljurayyan A Somerville L Teeter MG Vasarhelyi E Lanting B
Full Access

Introduction

Early functional recovery following total hip arthroplasty (THA) has the potential to increase patient satisfaction and reduce resource utilization. The direct anterior approach (DA) has been shown to provide earlier recovery compared to the direct lateral (DL) approach based on functional tests and outcome scores. There are limited studies that objectively evaluate functional recovery comparing the two approaches in the early post-operative period. Activity trackers have emerged as a valid tool to objectively quantify physical activity levels and potentially better assess functional status compared to commonly reported functional questionnaires. The purpose of this study is to measure physical activity levels in patients undergoing THA with the DA approach and compare these to THA with the direct lateral approach in the immediate postoperative period.

Methods

In a tertiary academic center we prospectively enrolled patients with primary OA that were eligible for a primary THA undergoing either the DA or the DL approach using the same prosthesis. Patients with comorbidities precluding them from ambulation, diagnoses of AVN or RA or undergoing bilateral THA were excluded. The number of steps walked per day were measured using wristband activity tracking technology for one week preoperatively, the first 2 weeks postoperatively and for 1 week leading up to their 6-week follow-up appointment. The University of California, Los Angeles (UCLA) activity score was also collected at the same two time points. Demographics were analyzed with descriptive statistics. A non-parametric Mann Whitney U test was used to determine whether a difference in physical activity levels exist between the DA and DL approach groups in the first 2 weeks and 6 weeks postoperatively.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 6 - 6
1 Oct 2018
Naudie DD Paish AD Nikolov HN Chmiel T El-Warrak AO Welch ID Teeter MG Holdsworth DW
Full Access

Introduction

As new innovations are developed to improve the longevity of joint replacement components, preclinical testing is necessary in the early stages of research into areas such as osseointegration, metal-cartilage wear and periprosthetic joint infection (PJI). Large-animal studies that test load-bearing components are expensive, however, requiring that animals be housed in special facilities that are not available at all institutions. Comparably, small animal models, such as the rat, offer several advantages including lower cost. Load-bearing implants remain difficult to manufacture via traditional methods in the sizes required for small-animal testing. Recent advances in additive manufacturing (3D metal-printing) have allowed for the creation of miniature joint replacement components in a variety of medical-grade metal alloys. The objective of this work is to create and optimize an image-based 3D-printed rat hip implant system that will allow in vivo testing of functional implant properties in a rat model.

Methods

A database of n=25 previously-acquired, 154μm micro-CT volumes (eXplore Locus Ultra, GE Medical) of male Sprague-Dawley rats (390–610g) were analyzed to obtain spatial and angular relationships between several anatomical features of the proximal rat femora. Mean measurements were used to guide the creation of a femoral implant template in computer-aided design software (Solidworks, Dassault Systemes). Several different variations were created, including collarless and collared designs, in a range of sizes to accommodate rats of various weights. Initial prototypes were 3D-printed 316L stainless steel with subsequent iterations printed in Ti6Al4V titanium and F75 cobalt-chrome. Implants were post-processed via sandblasting, hand-polishing, ultrasonic bath, and sterilization in an autoclave. Innate surface texturing was left on manufactured stems to promote osseointegration. Surgical implantation was performed in three live Sprague-Dawley rats (900g, 500g, 750g) with preservation of muscle attachments to the greater trochanter. Micro-CT imaging and X-ray fluoroscopy were performed post-operatively on each animal at 1 day, and 1, 3, 9 and 12 weeks to evaluate gait and component positioning.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 558 - 559
1 Nov 2011
Teeter MG Naudie DD McErlain DD Brandt J Yuan X MacDonald SJ Holdsworth DW
Full Access

Purpose: This study develops and validates a technique to quantify polyethylene wear in tibial inserts using micro-computed tomography (micro-CT), a nondestructive high resolution imaging technique that provides detailed images of surface geometry in addition to volumetric measurements.

Method: Six unworn and six wear-simulated Anatomic Modular Knee (DePuy Inc, Warsaw, IN) tibial inserts were evaluated. Each insert was scanned three times using micro-CT at a resolution of 50 μm. The insert surface was reconstructed for each scan through automatic segmentation and the insert volume was calculated. Gravimetric analysis was also performed for all inserts, and the micro-CT and gravimetric volumes were compared to determine accuracy. The utility of surface deviation maps derived from micro-CT was demonstrated by co-registering a worn and unworn insert. 3D deviations were measured continuously across the entire insert surface, including the articular and backside surfaces.

Results: The mean percent volume difference between the micro-CT and gravimetric techniques was 0.04% for the unworn inserts and 0.03% for the worn inserts. No significant difference was found between the micro-CT and gravimetric volumes for the unworn or worn inserts (P = 0.237 and P = 0.135, respectively). The mean coefficient of variation for volume between scans was 0.07% for both unworn and worn inserts. The map of surface deviations between the worn and unworn insert revealed focal deviations exceeding 750 μm due to wear.

Conclusion: Micro-CT provides precise and accurate volumetric measurements of polyethylene tibial inserts. Quantifiable 3D articular and backside surface deviation maps can be created from the detailed geometry provided by the technique. Compared to coordinate mapping, micro-CT provides 10 times greater surface sampling resolution (50 μm vs 500 μm) across the entire insert surface. Micro-CT is a useful analysis tool for wear simulator and retrieval studies of the polyethylene components used in total knee replacement.