Posterior-stabilized versus mid-level constraint polyethylene components in total knee arthroplasty

a biomechanical cadaveric analysis of laxity and collateral ligament forces



Download PDF

Abstract

Aims

Mid-level constraint designs for total knee arthroplasty (TKA) are intended to reduce coronal plane laxity. Our aims were to compare kinematics and ligament forces of the Zimmer Biomet Persona posterior-stabilized (PS) and mid-level designs in the coronal, sagittal, and axial planes under loads simulating clinical exams of the knee in a cadaver model.

Methods

We performed TKA on eight cadaveric knees and loaded them using a robotic manipulator. We tested both PS and mid-level designs under loads simulating clinical exams via applied varus and valgus moments, internal-external (IE) rotation moments, and anteroposterior forces at 0°, 30°, and 90° of flexion. We measured the resulting tibiofemoral angulations and translations. We also quantified the forces carried by the medial and lateral collateral ligaments (MCL/LCL) via serial sectioning of these structures and use of the principle of superposition.

Results

Mid-level inserts reduced varus angulations compared to PS inserts by a median of 0.4°, 0.9°, and 1.5° at 0°, 30°, and 90° of flexion, respectively, and reduced valgus angulations by a median of 0.3°, 1.0°, and 1.2° (p ≤ 0.027 for all comparisons). Mid-level inserts reduced net IE rotations by a median of 5.6°, 14.7°, and 17.5° at 0°, 30°, and 90°, respectively (p = 0.012). Mid-level inserts reduced anterior tibial translation only at 90° of flexion by a median of 3.0 millimetres (p = 0.036). With an applied varus moment, the mid-level insert decreased LCL force compared to the PS insert at all three flexion angles that were tested (p ≤ 0.036). In contrast, with a valgus moment the mid-level insert did not reduce MCL force. With an applied internal rotation moment, the mid-level insert decreased LCL force at 30° and 90° by a median of 25.7 N and 31.7 N, respectively (p = 0.017 and p = 0.012). With an external rotation moment, the mid-level insert decreased MCL force at 30° and 90° by a median of 45.7 N and 20.0 N, respectively (p ≤ 0.017 for all comparisons). With an applied anterior load, MCL and LCL forces showed no differences between the two inserts at 30° and 90° of flexion.

Conclusion

The mid-level insert used in this study decreased coronal and axial plane laxities compared to the PS insert, but its stabilizing benefit in the sagittal plane was limited. Both mid-level and PS inserts depended on the MCL to resist anterior loads during a simulated clinical exam of anterior laxity.

Cite this article: Bone Jt Open 2023;4(6):432–441.

Take home message

The mid-level constraint insert used in this study decreased axial and coronal laxity compared to the posterior-stabilized insert; however, it did not decrease sagittal plane laxity.

In the sagittal plane, anterior laxity following posterior-stabilized and mid-level total knee arthroplasty is dependent on the medial collateral ligament, which carries supraphysiological loads compared to the native knee.

Introduction

Up to 20% of patients who undergo total knee arthroplasty (TKA) are dissatisfied with the outcome.1,2 The feeling of knee instability is a common cause of patient dissatisfaction, which often leads to revision surgery.3-6 Surgical factors including bone resection techniques, ligament balancing, and implant design all contribute to TKA stability and are active areas of research to improve outcomes.6,7

Regarding implant design, many TKA systems have introduced mid-level constraint tibial inserts to manage increased intraoperative coronal plane laxity that cannot be resolved using standard ligament balancing techniques and a posterior-stabilized (PS) insert.8-10 Mid-level inserts are useful for patients with severe deformity and ligamentous laxity and have a low incidence of post-TKA instability or aseptic loosening at mid-term follow-up.8 Mid-level constraint varies by manufacturer but it is typically defined as an insert that has a wider, taller post than a PS insert, articulates with a primary femoral component, and resists coronal plane moments via post contact with the femoral box.8-11 Despite increased utilization of mid-level constraint, limited objective biomechanical data guiding rationale for their use are available.

While mid-level inserts provide constraint in the coronal plane,8,9 achieving stability requires consideration of kinematics and ligament loading in multiple planes.6,12,13 Unfortunately, the impact of PS and mid-level inserts on sagittal plane (anteroposterior; AP) laxity, axial plane (internal and external rotation; IE) laxity, and collateral ligament forces are not well understood. For example, excessive anterior tibial translation (ATT) of the tibia in midflexion may produce recurrent knee effusions, a sensation of ‘giving way’, difficulty with stair descent, and pain in the anterior periarticular soft-tissues.14 Moreover, in PS TKA, loading of the collateral ligaments is not well understood despite their importance in maintaining knee stability in the absence of the resected cruciate ligaments.15-20

Our goal was to quantify the impact of a mid-level constraint tibial insert on coronal, sagittal, and axial plane laxities and collateral ligament loading compared to a PS insert design. We asked two questions: 1) Does laxity in the coronal, axial, and sagittal planes differ between a PS and mid-level insert with simulated clinical exams under minimal compressive load?; and 2) How much force is carried by the medial and lateral collateral ligaments (MCL and LCL) in these three planes with PS and mid-level tibial inserts?

Methods

Surgery

We performed TKAs on eight independent cadaveric legs (five male, three female; mean age 63 years (standard deviation (SD) 12; 47 to 79); all left). All cadavers were fresh frozen, stored at -20°C, and then thawed at room temperature for 24 hours prior to TKA installation. Axial CT scans were then performed with 0.625 mm slice thickness and 0.5 × 0.5 mm2 in-plane resolution (Biograph mCT; Siemens, USA). The CT scans, together with medical history, physical examination, and direct inspection were used to ensure the absence of chondral, ligament, or meniscal injury, prior surgery, osteoarthritis, gross limb malalignment, and flexion contracture. One implant system (Persona; Zimmer Biomet, USA) was used. In this design, the mid-level insert has identical articular geometry as the PS insert but has a wider, taller post (Figure 1).

Fig. 1 
            Overlay of 3D scans of the posterior-stabilized insert (multicolored) and mid-level tibial insert (semitransparent) (Persona; Zimmer Biomet, USA). Inserts were imaged via CT (Biograph mCT; Siemens, Germany) and overlayed using 3D shape matching (Geomagic Wrap; 3D Systems, USA). The heat map indicates differences in insert geometries in millimetres. Hotter colors indicate distances and locations where the PS insert geometry is smaller than the mid-level insert geometry. The maximum difference at the articular surface was < 0.1 mm.

Fig. 1

Overlay of 3D scans of the posterior-stabilized insert (multicolored) and mid-level tibial insert (semitransparent) (Persona; Zimmer Biomet, USA). Inserts were imaged via CT (Biograph mCT; Siemens, Germany) and overlayed using 3D shape matching (Geomagic Wrap; 3D Systems, USA). The heat map indicates differences in insert geometries in millimetres. Hotter colors indicate distances and locations where the PS insert geometry is smaller than the mid-level insert geometry. The maximum difference at the articular surface was < 0.1 mm.

Subsequent TKA surgery utilized the measured resection technique targetting neutral mechanical alignment via manual instrumentation through a medial parapatellar approach. A typical dissection was performed with subperiosteal release of the deep MCL as is characteristic given the neutral limb alignment of our cohort.21 After removal of the cruciate ligaments, a standard distal femoral resection (10 mm from the distal medial femoral condyle) was performed. A tibial cutting jig was placed between the middle and medial thirds of tibial tubercle, in line with the centre of the ankle joint, and with 3° of posterior slope. Nine millimetres of bone was measured from the lateral tibial plateau to set resection depth. Femoral sizing was conducted, and a 4-in-1 cutting jig was aligned with the transepicondylar and anteroposterior axes. Trialling was performed, and the PS insert size was chosen based on examination by a fellowship-trained arthroplasty surgeon as would be carried out in the operating room. A partial release of the posterior capsule was utilized to achieve full extension in two knees. No other releases were performed.

Robotic testing

After TKA installation, specimens were prepared for robotic testing (ZX165U; Kawasaki Robotics, USA). First, the femur and tibia were sectioned 15 cm proximal and distal to the joint line, respectively, and set in potting cement (Bondo; 3M, USA). Subsequently, the potted femur was rigidly fixed to the ground through a pedestal, and the tibia was fixed to the robotic manipulator at full extension (Figure 2).22 The robot end effector was instrumented with a six-axis force/torque sensor (Theta; ATI Industrial Automation, USA). We performed repeated measures testing of the PS and mid-level inserts and alternated the order of insert testing from knee-to-knee to account for bias due to ligament stretching.

Fig. 2 
            Robotic manipulator with six-axis force/torque sensor (load cell) with cadaveric knee specimen mounted in place. Each specimen was tested with the posterior-stabilized and mid-level tibial inserts (Persona; Zimmer Biomet, USA) shown to the right.

Fig. 2

Robotic manipulator with six-axis force/torque sensor (load cell) with cadaveric knee specimen mounted in place. Each specimen was tested with the posterior-stabilized and mid-level tibial inserts (Persona; Zimmer Biomet, USA) shown to the right.

Next, the knee coordinate system was defined. Tibial rotations and translations relative to the femur were described using a coordinate system based on geometrical features of the tibial tray and femoral component (Supplementary Material, Knee Coordinate System Definition).23 Specifically, the axis of internal-external (IE) rotation was oriented perpendicular to the tray, and the AP axis was aligned parallel to both the tibial tray and to the femoral box and centred within the tibial tray. Varus and valgus (VV) and IE moments and AP forces were applied, and the resulting rotations and translations were measured about and along these axes.22,24

After defining the coordinate system, the knee was flexed from 0° to 90° in 1° increments under 10 N of compression with forces and moments in the remaining directions minimized. Convergence tolerances for the loading algorithms were resultant forces and moments within 5 N and 0.4 Newton-metres (Nm), respectively, of the target loads.25

Subsequently, the collateral and capsular tissues were preconditioned at full extension via five cycles each of varus and valgus loading to 6 Nm in each direction. After preconditioning, laxity assessments were conducted. All assessments were carried out under minimal compressive load (10 N) to emphasize the contribution of the collateral ligaments,11 and since surgeons focus on insert selection and soft-tissue tensioning in the operative setting by examining the knee with minimal compressive loads. Coronal laxity was assessed by measuring varus and valgus angulations (in degrees) in response to varus and valgus moments of 6 Nm with the knee at 0° (full extension), 30° (midflexion), and 90° of flexion, as in previous work.11 Anterior laxity was assessed by quantifying ATT relative to the fixed femur with 30 N anterior force applied at 0° and 30° of flexion and a 20 N anterior force applied at 90°. The 30 N anterior force was selected because it avoided tibial dislocation and excessive loading of the collateral ligaments. ATT was referenced to an applied posterior force of 20 N to ensure contact between the femoral cam and tibial post. IE laxity was assessed at the same three flexion angles by quantifying the net IE tibial rotation relative to the femur under applied moments of ± 5 Nm. Finally, coupled tibial axial rotation was defined as the net change in IE rotation with the applied anterior force.16,26

Finally, serial ligament sectioning and the principle of superposition were used to determine the resultant in situ MCL and LCL forces at the peak applied VV, anterior, and IE loads.27 To calculate ligament force, the tibial insert was removed from the knee joint and then tibiofemoral kinematics corresponding to each knee condition were repeated immediately before and after sectioning each ligament. The collaterals were identified and sectioned at the joint line. Ligament force was then calculated as the resultant of the vector difference in force measured across the knee before and after sectioning each ligament. The order of sectioning and of replaying the kinematics was randomized. When replaying the kinematics, the velocity of the robot averaged 0.8 mm/second in the AP direction and 0.13°/second in the varus and valgus directions.

This study was approved by our organization’s institutional review board.

Statistical analysis

We conducted a power analysis based on work by Luyckx et al,28 who reported a mean of 7.5° (SD 1.9°) of coronal laxity at 90° of flexion. Using these data and defining a minimal clinically significant difference of 2.5° revealed that a sample size of seven knees would achieve 80% power to detect a difference between groups with two-sided significance level (α) of 0.050 in a repeated measures study design.

Our data were not normally distributed as determined via Kolmogrov-Smirnov tests (p < 0.05). Therefore, all outcome measures were reported as medians and interquartile ranges (IQRs). Varus and valgus angulations, ATT, IE rotation, coupled axial rotation, and collateral ligament forces were compared across PS and mid-level inserts via individual Wilcoxon signed-rank tests. Signficance was set at p < 0.05.

Results

Regarding our first research question, the mid-level constraint insert reduced varus angulation by a median of 0.4° (IQR 0.1 to 0.8), 0.9° (IQR 0.4 to 2.0), and 1.5° (IQR 0.2 to 2.0) at 0°, 30°, and 90° of flexion, respectively (p ≤ 0.018 for all comparisons; Table I; Figure 3). Similarly, the mid-level constraint reduced valgus angulation by a median of 0.3° (IQR 0.2 to 0.5), 1.0° (IQR 0.4 to 1.3) and 1.2° (IQR 0.4 to 1.8) at 0°, 30°, and 90°, respectively (p ≤ 0.027 for all comparisons; Table I; Figure 3). In the axial plane, the mid-level inserts reduced net IE rotations by 5.6° (IQR 3.2 to 9.3), 14.7° (IQR 9.0 to 21.2), and 17.5° (IQR 13.0 to 19.7) at 0°, 30°, and 90°, respectively (p = 0.012 for all comparisons) (Table I, Figure 4). In the sagittal plane, knees with the mid-level insert showed less ATT than the PS insert; however, the decrease was only statistically significant at 90°of flexion, with a median reduction of 3.0 mm (IQR 0.9 to 3.7) (p = 0.036) (Figure 5). With the applied anterior tibial load, the mid-level insert decreased coupled axial rotation at 30° and 90° by a median of 2.2° (IQR 1.7 to 4.2) (p = 0.012) and 4.1° (IQR 3.0 to 6.5) (p = 0.025), respectively (Figure 6).

Table I.

Summary of kinematic outcome measurements for posterior-stabilized and mid-level tibial inserts.

Applied load Outcome measure Flexion angle (°) Median (IQR) p-value*
PS Mid-level
Varus 6 Nm Varus angulation (°) 0 2.0 (1.2 to 2.7) 1.5 (0.7 to 2.1) 0.017
30 4.2 (1.7 to 4.9) 1.9 (1.3 to 2.9) 0.018
90 3.5 (3.1 to 6.8) 2.9 (2 to 4.4) 0.017
Valgus 6 Nm Valgus angulation (°) 0 2.2 (1 to 3.2) 1.9 (0.8 to 2.8) 0.020
30 3.2 (2.0 to 4.0) 2.3 (1.3 to 3.1) 0.027
90 3.4 (2.7 to 4.2) 2.4 (1.8 to 2.8) 0.025
Anterior 30 N Anterior translation (mm) 0 3.1 (2.3 to 3.4) 2.4 (1.8 to 3.4) 0.063
30 9.7 (5.4 to 20) 7.0 (3.5 to 16.9) 0.050
Anterior 20 N 90 5.2 (4.6 to 5.9) 3.3 (0.9 to 5.2) 0.036
IE ± 5 Nm Axial rotation (°) 0 19.3 (10.7 to 27.8) 13.1 (7.5 to 18.4) 0.012
30 30.2 (22.1 to 47.3) 15.8 (12.1 to 22.9) 0.012
90 34.5 (31.3 to 40.7) 18.7 (12.8 to 26.6) 0.012
  1. *

    Wilcoxon signed-rank test.

  1. IE, internal-external rotation; IQR, interquartile range; N, Newton; Nm, Newton-metres; PS, posterior-stabilized.

Fig. 3 
          Varus and valgus angulations in degrees in response to varus and valgus moments of 6 Nm at 0°, 30°, and 90° of flexion for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and 5th and 95th percentiles, respectively. The horizontal black line within each box is the median. *p < 0.05.

Fig. 3

Varus and valgus angulations in degrees in response to varus and valgus moments of 6 Nm at 0°, 30°, and 90° of flexion for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and 5th and 95th percentiles, respectively. The horizontal black line within each box is the median. *p < 0.05.

Fig. 4 
          Net internal-external (axial) rotation in degrees of the tibia in response to an applied moment of ± 5 Newton-metres at 0°, 30°, and 90° of flexion for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. *p < 0.05.

Fig. 4

Net internal-external (axial) rotation in degrees of the tibia in response to an applied moment of ± 5 Newton-metres at 0°, 30°, and 90° of flexion for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. *p < 0.05.

Fig. 5 
          Anterior tibial translation in miilimeters (mm) in response to a 30 N anterior force at 0° and 30° of flexion and a 20 N anterior force at 90° for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. Circles indicate outliers. *p < 0.05.

Fig. 5

Anterior tibial translation in miilimeters (mm) in response to a 30 N anterior force at 0° and 30° of flexion and a 20 N anterior force at 90° for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. Circles indicate outliers. *p < 0.05.

Fig. 6 
          Magnitude of coupled axial rotation in degrees during anteroposterior testing in response to a 30 N anterior force at 0° and 30° of flexion and a 20 N anterior force at 90° for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. Circles indicate outliers. *p < 0.05.

Fig. 6

Magnitude of coupled axial rotation in degrees during anteroposterior testing in response to a 30 N anterior force at 0° and 30° of flexion and a 20 N anterior force at 90° for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. Circles indicate outliers. *p < 0.05.

Regarding our second research question, under an applied varus moment the mid-level insert significantly decreased LCL force compared to the PS insert by a median of 4.4 N (IQR 1.8 to 8.6), 22.3 N (IQR 13.8 to 30.2), and 32.1 N (IQR 11.4 to 37.7 ) at 0°, 30°, and 90° of flexion, respectively (p ≤ 0.036 for all comparisons). In contrast, the mid-level insert did not reduce MCL force with an applied valgus moment (Table II, Figure 7). With an applied internal rotation moment, the mid-level insert significantly decreased LCL force at 30° and 90° of flexion by a median of 25.7 N (IQR 15.1 to 40.0) and 31.7 N (IQR 8.8 to 48.7), respectively (p ≤ 0.017 for all comparisons) (Table II). With an applied external rotation moment, the mid-level insert significantly decreased MCL force at 30° and 90° by a median of 45.7 N (IQR 26.4 to 82.4) and 20.0 N (IQR 6.7 to 50.0), respectively (p ≤ 0.017 for all comparisons). With an applied anterior load, MCL and LCL forces showed no statistical differences between PS versus mid-level inserts at both 30° and 90° (Figure 8).

Table II.

Summary of measured ligament forces for posterior-stabilized and mid-level tibial inserts.

Applied load Ligament force (N) Flexion angle (°) Median (IQR) p-value*
PS Mid-level
Varus 6 Nm LCL 0 45 (22 to 81) 40 (18 to 61) 0.036
30 63 (46 to 75) 40 (22 to 53) 0.017
90 40 (36 to 54) 8 (3 to 17) 0.012
Valgus 6 Nm MCL 0 69 (56 to 112) 49 (40 to 110) 0.327
30 82 (69 to 115) 70 (46 to 105) 0.123
90 55 (52 to 75) 50 (46 to 69) 0.161
Anterior 30 N LCL 0 13 (3 to 30) 13 (2 to 26) 0.779
30 9 (5 to 13) 6 (3 to 13) 0.263
Anterior 20 N 90 6 (2 to 6) 1 (0 to 2) 0.093
Anterior 30 N MCL 0 19 (5 to 59) 26 (7 to 82) 0.017
30 47 (30 to 72) 44 (29 to 73) 0.401
Anterior 20 N 90 10 (1 to 27) 12 (9 to 37) 0.050
Internal rotation 5 Nm LCL 0 38 (8 to 50) 17 (6 to 27) 0.093
30 37 (23 to 48) 2 (1 to 18) 0.017
90 38 (13 to 56) 1 (1 to 7) 0.012
Internal rotation 5 Nm MCL 0 57 (31 to 94) 35 (19 to 103) 0.484
30 70 (37 to 113) 60 (34 to 89) 0.263
90 39 (33 to 69) 19 (5 to 58) 0.017
External rotation 5 Nm LCL 0 42 (17 to 78) 31 (12 to 47) 0.012
30 28 (10 to 68) 20 (9 to 49) 0.093
90 2 (1 to 4) 2 (1 to 3) 0.327
External rotation 5 Nm MCL 0 45 (23 to 106) 31 (7 to 101) 0.208
30 102 (88 to 123) 48 (3 to 84) 0.012
90 67 (53 to 78) 41 (22 to 62) 0.017
  1. *

    Wilcoxon signed-rank test.

  1. IQR, interquartile range; LCL, lateral collateral ligament; MCL, medial collateral ligament; Nm, Newton-metres; PS, posterior-stabilized.

Fig. 7 
          a) Medial collateral ligament (MCL) force in Newtons (N) and b) lateral collateral ligament (LCL) force in response to applied moments of 6 Nm valgus and 6 Nm varus, respectively, at 0°, 30°, and 90° of flexion for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. Circles indicate outliers. *p < 0.05.

Fig. 7

a) Medial collateral ligament (MCL) force in Newtons (N) and b) lateral collateral ligament (LCL) force in response to applied moments of 6 Nm valgus and 6 Nm varus, respectively, at 0°, 30°, and 90° of flexion for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. Circles indicate outliers. *p < 0.05.

Fig. 8 
          a) Medial collateral ligament (MCL) force in Newtons (N) and b) lateral collateral ligament (LCL) force in response to a 30 N anterior force at 0° and 30° of flexion and a 20 N anterior force at 90° for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. Circles indicate outliers. *p < 0.05.

Fig. 8

a) Medial collateral ligament (MCL) force in Newtons (N) and b) lateral collateral ligament (LCL) force in response to a 30 N anterior force at 0° and 30° of flexion and a 20 N anterior force at 90° for posterior-stabilized (PS) and mid-level (ML) tibial inserts. Boxes and whiskers correspond to the quartiles and fifth and 95th percentiles, respectively. The horizontal black line within each box is the median. Circles indicate outliers. *p < 0.05.

Discussion

Mid-level constraint inserts reduced varus, valgus, and IE rotational laxities compared to PS inserts with a smaller reduction in varus and valgus (median reductions ≤ 1.2° in valgus and ≤ 1.5° in varus) compared to axial rotation (median reductions ≤ 11.9°) (Table I). Mid-level inserts had minimal effect on AP laxity. The minor decreases in ATT are probably due to the observed reduction in coupled axial rotation. Mid-level inserts decreased LCL forces under a varus moment, but they did not reduce MCL forces under a valgus moment, likely because valgus angulations were lower and therefore not as dependent on the tibial post (Figure 3). The mid-level insert also decreased collateral ligament loads with applied IE moments (Table II). Finally, the mid-level insert did not decrease the forces on the collateral ligaments with an anterior applied load (Figure 8). Importantly, the MCL was the primary ligamentous restraint to ATT in midflexion for both inserts (Table II). The key clinical message of this study is that since mid-level inserts are often selected intraoperatively in the setting of lax collateral ligaments for coronal stability, surgeons must be aware that mid-level inserts still depend on the MCL for anterior stability. Therefore, mid-level inserts would confer limited anterior restraint in the setting of MCL insufficiency, which, we speculate, may contribute to patient perceptions of knee instability.

Biomechanical data describing the impact of mid-level inserts on laxity and ligament loading are limited. In a robotic study of a constrained condylar knee (CCK) design, Wang et al11 demonstrated varus angulations that increased from an average of 1.0° in extension to 1.5° at 90° of flexion and valgus angulation that increased slightly from 1.0° in extension to 1.7° in flexion. Our coronal plane testing of the mid-level insert also showed increasing VV angulations with increasing knee flexion and, as expected, showed greater varus and valgus laxity compared to the previously reported data for the CCK design (Table I). The net VV angulations of the mid-level inserts in our study were also larger than the manufacturer-reported 1.5°.9 This discrepancy could be due to differences in testing conditions, which were not specified by the manufacturer; it highlights the importance of rigorously describing testing parameters such as the applied loads, coordinate system definitions, and allowable directions of tibiofemoral motion to enable comparison of implant behavior.

Mid-level inserts reduced IE laxity to a greater extent than VV laxity (Table I). We speculate that the increased axial constraint may have implications for aseptic loosening as the reduced motions could lead to increased loading at the fixation interface. Fortunately, this concern has not manifested in early clinical outcome studies of inserts with mid-level constraint.8,10

Our findings call into question the ability of mid-level inserts to address laxity in the sagittal plane. The mid-level insert used in this study, which has an identical articular surface profile as the PS insert (Figure 1), decreased ATT only at 90° of flexion by a median of 3 mm (Figure 5). Similar to a CCK insert,11 this decrease in ATT is likely related to the reduced coupled axial rotation caused by increased conformity between the femoral box and the wider, taller tibial post in the mid-level compared to the PS insert (Figure 6). However, we speculate that the additional anterior restraint afforded by the mid-level insert is likely inadequate to prevent instability. Clinically, in cases of MCL laxity, a surgeon choosing a mid-level insert should be aware that although the increased constraint could decrease laxity in the coronal and axial planes, the knee will still be left with increased ATT given the lax condition of the MCL. Given the dual role of the MCL in resisting coronal and sagittal loading in PS TKA, future investigation in mid-level polyethylene designs should also address AP laxity, particularly in midflexion.

Under applied VV moments, the mid-level insert offloaded the LCL with varus but not the MCL with valgus (Figure 7). This finding may be related to our coronal plane laxity data. Specifically, the maximum varus angulation for the PS insert was up to 94% greater than the maximum valgus angulation at 30° of flexion (9.7° varus vs 5.0° valgus) (Figure 3). Thus, the MCL likely engages (i.e. carries load) earlier with valgus loading than the LCL does with varus loading. This earlier engagement of the MCL likely causes it to carry load before the tibial post contacts the femoral box, thereby preventing the post from offloading the MCL. In contrast, the larger varus angulations enable the post and box to make contact and to share load with the LCL.

In the sagittal plane, the mid-level insert did not offload the collateral ligaments compared to the PS insert. The collateral ligaments, primarily the MCL, carried high forces at both 30° (midflexion) and 90° with both inserts (Table II). We speculate that this finding has important implications for using mid-level inserts for multiplanar ligamentous laxity, as this insert may not provide the desired sagittal plane constraint, potentially contributing to symptomatic flexion or mid-flexion instability and to postoperative discomfort around the MCL. The MCL of the replaced knee carries several times greater force than MCL in the native knee with an applied anterior load in midflexion.29,30 We speculate that increased loading of the MCL following TKA could contribute to the commonly observed clinical scenario in which the MCL ‘stretches out’ over time, leading to a patient’s perception that the TKA does not feel ‘normal’, potentially contributing to patient dissatisfaction.

Regarding limitations, our study evaluated the PS and mid-level inserts of only one commercial implant system. Variations in PS and mid-level designs among manufacturers, including designs with taller posts and with single radius congruent designs, could limit the broader applicability of our results. Studies are needed to determine the extent to which our findings apply to other PS TKA designs. In addition, our study was performed in a cadaveric model without an extensor mechanism or hamstring forces. Thus, our study focused on the stabilizing role of the passive restraints (i.e. articular surfaces and ligaments). We also used minimal compressive load to simulate the intraoperative testing scenario, which maximizes the contribution of the soft-tissues to joint stability and minimizes the role of the bearing surfaces. Also, mechanical cutting jigs were used to implant the knee prosthesis with neither robotic assistance nor computer navigation, which reflects the most common clinical scenario. This approach likely increased variability in ligament balance among specimens. In addition, midflexion testing was limited to 30° of flexion to reduce the number of loading cycles and to mitigate soft-tissue stretching. We focused on this flexion angle because large increases in frontal plane laxity occur in early midflexion with PS TKA,28,31 and our surgeon team prefers this angle of intraoperative assessment. Finally, testing was performed on knees with intact collateral ligaments, which differs from the clinical setting in which mid-level inserts are most used.

Mid-level inserts decreased coronal and axial laxities under loads simulating a clinical knee exam, but they were less effective in reducing sagittal plane laxity, particularly in midflexion. Thus, surgeons should be aware that mid-level inserts provide axial and coronal restraint with the potential to offload the collateral ligaments; however, they are not as effective in resisting anterior loads in the sagittal plane. In the sagittal plane, ATT with the PS and mid-level insert designs used in this study is dependent on the MCL, which carries supraphysiological loads relative to the native knee.


Correspondence should be sent to Cynthia A. Kahlenberg. E-mail:

References

1. Bourne RB , Chesworth BM , Davis AM , Mahomed NN , Charron KDJ . Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res . 2010 ; 468 ( 1 ): 57 63 . Crossref , PubMed Google Scholar

2. Kahlenberg CA , Nwachukwu BU , McLawhorn AS , Cross MB , Cornell CN , Padgett DE . Patient satisfaction after total knee replacement: a systematic review . HSS J . 2018 ; 14 ( 2 ): 192 201 . Crossref , PubMed Google Scholar

3. Paxton EW , Namba RS , Maletis GB , et al. A prospective study of 80,000 total joint and 5000 anterior cruciate ligament reconstruction procedures in a community-based registry in the united states . J Bone Joint Surg Am . 2010 ; 92 Suppl 2 : 117 132 . Crossref , PubMed Google Scholar

4. Lombardi AV , Berend KR , Adams JB . Why knee replacements fail in 2013: patient, surgeon, or implant? Bone Joint J . 2014 ; 96-B ( 11 Supple A ): 101 104 . Crossref , PubMed Google Scholar

5. Parratte S , Pagnano MW . Instability after total knee arthroplasty . J Bone Joint Surg Am . 2008 ; 90 ( 1 ): 184 194 . PubMed Google Scholar

6. Abdel MP , Haas SB . The unstable knee: wobble and buckle . Bone Joint J . 2014 ; 96-B ( 11 Supple A ): 112 114 . Crossref , PubMed Google Scholar

7. Wright TM . Joint stability in total knee arthroplasty: what is the target for a stable knee? J Am Acad Orthop Surg . 2017 ; 25 Suppl 1 : S25 S28 . Crossref , PubMed Google Scholar

8. Crawford DA , Law JI , Lombardi AV , Berend KR . Midlevel constraint without stem extensions in primary total knee arthroplasty provides stability without compromising fixation . J Arthroplasty . 2018 ; 33 ( 9 ): 2800 2803 . Crossref , PubMed Google Scholar

9. Kahlenberg CA , Chalmers B , Sun HJ , et al. Polyethylene components in primary total knee arthroplasty: a comprehensive overview of fixed bearing design options . J Knee Surg . 2022 ; 35 ( 13 ): 1401 1408 . Crossref , PubMed Google Scholar

10. Dubin JA , Westrich GH . Mid-level constraint may correct coronal plane imbalance without compromising patient function in patients with severe osteoarthritis . J Orthop . 2020 ; 21 : 84 87 . Crossref , PubMed Google Scholar

11. Wang X , Malik A , Bartel DL , Wright TM , Padgett DE . Load sharing among collateral ligaments, articular surfaces, and the tibial post in constrained condylar knee arthroplasty . J Biomech Eng . 2016 ; 138 ( 8 ). Crossref , PubMed Google Scholar

12. Vince KG , Abdeen A , Sugimori T . The unstable total knee arthroplasty: causes and cures . J Arthroplasty . 2006 ; 21 ( 4 Suppl 1 ): 44 49 . Crossref , PubMed Google Scholar

13. Whiteside LA . Soft tissue balancing: the knee . J Arthroplasty . 2002 ; 17 ( 4 Suppl 1 ): 23 27 . Crossref , PubMed Google Scholar

14. Schwab JH , Haidukewych GJ , Hanssen AD , Jacofsky DJ , Pagnano MW . Flexion instability without dislocation after posterior stabilized total knees . Clin Orthop Relat Res . 2005 ; 440 ( NA ): 96 100 . Crossref , PubMed Google Scholar

15. Butler DL , Noyes FR , Grood ES . Ligamentous restraints to anterior-posterior drawer in the human knee. A biomechanical study . J Bone Joint Surg Am . 1980 ; 62 ( 2 ): 259 270 . PubMed Google Scholar

16. Fukubayashi T , Torzilli PA , Sherman MF , Warren RF . An in vitro biomechanical evaluation of anterior-posterior motion of the knee. Tibial displacement, rotation, and torque . J Bone Joint Surg Am . 1982 ; 64 ( 2 ): 258 264 . PubMed Google Scholar

17. Sakane M , Fox RJ , Woo SL , Livesay GA , Li G , Fu FH . In situ forces in the anterior cruciate ligament and its bundles in response to anterior tibial loads . J Orthop Res . 1997 ; 15 ( 2 ): 285 293 . Crossref , PubMed Google Scholar

18. Markolf KL , Mensch JS , Amstutz HC . Stiffness and laxity of the knee--the contributions of the supporting structures. A quantitative in vitro study . J Bone Joint Surg Am . 1976 ; 58 ( 5 ): 583 594 . PubMed Google Scholar

19. Markolf KL , Graff-Radford A , Amstutz HC . In vivo knee stability. A quantitative assessment using an instrumented clinical testing apparatus . J Bone Joint Surg Am . 1978 ; 60 ( 5 ): 664 674 . Crossref , PubMed Google Scholar

20. Hsieh HH , Walker PS . Stabilizing mechanisms of the loaded and unloaded knee joint . J Bone Joint Surg Am . 1976 ; 58 ( 1 ): 87 93 . PubMed Google Scholar

21. Cheung A , Yan CH , Chan PK , Chiu KY . The medial collateral ligament in primary total knee arthroplasty: anatomy, biomechanics, and injury . J Am Acad Orthop Surg . 2020 ; 28 ( 12 ): e510 e516 . Crossref , PubMed Google Scholar

22. Marom N , Ouanezar H , Jahandar H , et al. Lateral extra-articular tenodesis reduces anterior cruciate ligament graft force and anterior tibial translation in response to applied pivoting and anterior drawer loads . Am J Sports Med . 2020 ; 48 ( 13 ): 3183 3193 . Crossref , PubMed Google Scholar

23. Grood ES , Suntay WJ . A joint coordinate system for the clinical description of three-dimensional motions: application to the knee . J Biomech Eng . 1983 ; 105 ( 2 ): 136 144 . Crossref , PubMed Google Scholar

24. Camp CL , Jahandar H , Sinatro AM , Imhauser CW , Altchek DW , Dines JS . Quantitative anatomic analysis of the medial ulnar collateral ligament complex of the elbow . Orthop J Sports Med . 2018 ; 6 ( 3 ): 2325967118762751 . Crossref , PubMed Google Scholar

25. Imhauser C , Mauro C , Choi D , et al. Abnormal tibiofemoral contact stress and its association with altered kinematics after center-center anterior cruciate ligament reconstruction: an in vitro study . Am J Sports Med . 2013 ; 41 ( 4 ): 815 825 . Crossref , PubMed Google Scholar

26. Livesay GA , Rudy TW , Woo SL , et al. Evaluation of the effect of joint constraints on the in situ force distribution in the anterior cruciate ligament . J Orthop Res . 1997 ; 15 ( 2 ): 278 284 . Crossref , PubMed Google Scholar

27. Woo SL-Y , Fox RJ , Sakane M , Livesay GA , Rudy TW , Fu FH . Biomechanics of the ACL: measurements of in situ force in the ACL and knee kinematics . The Knee . 1998 ; 5 ( 4 ): 267 288 . Crossref Google Scholar

28. Luyckx T , Vandenneucker H , Ing LS , Vereecke E , Ing AV , Victor J . Raising the joint line in TKA is associated with mid-flexion laxity: a study in cadaver knees . Clin Orthop Relat Res . 2018 ; 476 ( 3 ): 601 611 . Crossref , PubMed Google Scholar

29. Kanamori A , Sakane M , Zeminski J , Rudy TW , Woo SL . In-situ force in the medial and lateral structures of intact and ACL-deficient knees . J Orthop Sci . 2000 ; 5 ( 6 ): 567 571 . Crossref , PubMed Google Scholar

30. Sakane M , Livesay GA , Fox RJ , Rudy TW , Runco TJ , Woo SL . Relative contribution of the ACL, MCL, and bony contact to the anterior stability of the knee . Knee Surg Sports Traumatol Arthrosc . 1999 ; 7 ( 2 ): 93 97 . Crossref , PubMed Google Scholar

31. Chalmers BP , Elmasry SS , Kahlenberg CA , et al. Additional distal femoral resection increases mid-flexion coronal laxity in posterior-stabilized total knee arthroplasty with flexion contracture: a computational study . Bone Joint J . 2021 ; 103-B ( 6 Supple A ): 87 93 . Crossref , PubMed Google Scholar

Author contributions

C. A. Kahlenberg: Conceptualization, Methodology, Project administration, Investigation, Data curation, Formal analysis, Writing – original draft, Writing – review & editing.

E. E. Berube: Software, Investigation, Data curation, Formal analysis, Visualization, Writing – original draft, Writing – review & editing.

W. Xiang: Methodology, Investigation, Data curation, Formal analysis.

J. E. Manzi: Investigation, Data curation.

H. Jahandar: Conceptualization, Methodology, Investigation, Data curation.

B. P. Chalmers: Conceptualization, Formal analysis, Writing – review & editing.

M. Cross: Conceptualization, Formal analysis, Writing – review & editing.

D. J. Mayman: Conceptualization, Formal analysis, Writing – review & editing.

T. M. Wright: Conceptualization, Formal analysis, Writing – review & editing.

G. Westrich: Conceptualization, Formal analysis, Writing – review & editing.

C. W. Imhauser: Conceptualization, Methodology, Investigation, Data curation, Formal analysis, Writing – original draft, Writing – review & editing.

P. K. Sculco: Conceptualization, Methodology, Investigation, Data curation, Formal analysis, Writing – original draft, Writing – review & editing.

Funding statement

The authors disclose receipt of the following financial or material support for the research, authorship, and/or publication of this article: Zimmer Biomet provided the implants used in this study. Funding from the Clark Foundation and Kirby Foundation

ICMJE COI statement

M. B. Cross reports royalties or licenses from Depuy Synthes and PSI, consulting fees from Depuy Synthes, Exactech, and 3M, payment or honoraria for lectures, presentations, speakers bureaus, manuscript writing or educational events from Depuy Synthes and 3M, and stock or stock options in Ospitek, PSI, and Intellijoint, all of which are unrelated to this article. D. J. Mayman reports royalties or licenses from Orthalign and Smith & Nephew, and stock or stock options in Imagen, Orthalign, Smith & Nephew, Wishbone, Micare Path, and Cymedica, all of which are unrelated to this article. C. W. Imhauser reports payment or honoraria for lectures, presentations, speakers bureaus, manuscript writing or educational events from Corin Medical, unrelated to this study. P. K. Sculco reports consulting fees from Zimmer Biomet, Intellijoint Surgical, Lima Corporate, Depuy Synthes, and ATEC, payment or honoraria for lectures, presentations, speakers bureaus, manuscript writing or educational events from Zimmer Biomet, stock or stock options in Parvizi Surgical Innovation and Intellijoint Surgical, and receipt of equipment, materials, drugs, medical writing, gifts or other services, all of which are unrelated to this study.

Data sharing

The data that support the findings for this study are available to other researchers from the corresponding author upon reasonable request.

Ethical review statement

This study was approved by our institution’s institutional review board (IRB # 2016-0071).

Open access funding

Publication fees were paid for by Hospital for Special Surgery.

Supplementary material

Knee Coordinate System Definition.

© 2023 Author(s) et al. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (CC BY-NC-ND 4.0) licence, which permits the copying and redistribution of the work only, and provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc-nd/4.0/