Satisfactory medium- to long-term outcomes of cemented rotating hinge prosthesis in revision total knee arthroplasty

a specialist centre study with minimum four years’ follow-up



Download PDF

Abstract

Aims

The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes.

Methods

A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.

Results

Overall, 69 consecutive patients (69 knees) were included with a median age of 78 years (interquartile range 69 to 84), and there were 46 females (66.7%). Indications were septic revisions in 26 (37.7%), and aseptic aetiology in the remining 43 (62.3%). The mean follow-up was 9.7 years (4 to 18), and the overall complication was rate was 7.24%, all with patellofemoral complications. Failure rate with ‘any cause revision’ was 5.8%. There was one case of aseptic loosening of the femoral component. At ten years, 17/69 patients (24.63%) had died, and implant survivorship was 92.2%.

Conclusion

In our experience, the SMILES rotating hinge prosthesis achieves satisfactory long-term outcomes with ten-year implant survivorship of 92.2% and a patellofemoral complication rate of 7.24%.

Cite this article: Bone Jt Open 2023;4(10):776–781.

Take home message

The Stanmore Modular Individualized Lower Limb System (SMILES) cemented rotating hinge prothesis has 92% survivorship in revision total knee arthroplasty patients.

Aseptic loosening is usually seen more on the femoral side, and a well-fixed tibia can be revised to SMILES distal femoral replacement, keeping the same tibial component.

Introduction

The demand for primary total knee arthroplasty (TKA) continues to rise worldwide with projected increase in revision TKA (rTKA).1-3 In its 19th annual report, the UK National Joint Registry (NJR) reported re-revision rate of 3.54% at one year, rising to 15.4% at ten years, and 18.98% at 15 years following first rTKA, with aseptic loosening, infection, and instability accounting for the majority of re-revisions.1 With re-revisions, the complexity increases, as does the use of rotating-hinge prostheses.4 In a recent systematic review of rotating-hinge implants in rTKA, survivorship was reported from 51% to 92.5% up to ten years’ follow-up, with complication rates ranging from 9.2% to 63%.5 Notably, patellofemoral complications remain high, with patella subluxation reported in 29.6% of cases in a recent contemporary series.6

Since the introduction of the rotating-hinge mechanism, contemporary hinge implants have seen improved outcomes and survivorship. However, common complications with hinges remain high with early loosening and patellofemoral issues.7-9 The indications to use a rotating-hinge implant in rTKA practice are well-established, including collateral ligament deficiency, severe bone loss that compromises ligament attachments, gross flexion-extension mismatch, recurvatum, and gross multidirectional instability.10

Patellofemoral instability with a hinge construct is multifactorial with patients’, surgical and implant-related factors. Based on the specific hinge design, the mobile bearings allow rotation with varying degrees of constraint.4 The Stanmore Modular Individualized Lower Limb System (SMILES) was first introduced in 1990. In this design, there is some rotational control built in the under-surface of the tibial component, so as it rotates it rides upwards providing some resistance to rotation which is advantageous for the patellofemoral mechanics (Figure 1). This system is fully cemented and of relatively low cost.

Fig. 1 
          Photograph of the Stanmore Modular Individualized Lower Limb System (SMILES) prosthesis; the polyethylene insert is bevelled placed on the tibial plate and limits rotational motion to ± 5°.

Fig. 1

Photograph of the Stanmore Modular Individualized Lower Limb System (SMILES) prosthesis; the polyethylene insert is bevelled placed on the tibial plate and limits rotational motion to ± 5°.

In this study, we aim to evaluate the medium- to long-term outcomes of SMILES rotating hinge prosthesis in rTKA at our tertiary unit. We hypothesize that this fully cemented system leads to satisfactory clinical outcomes.

Methods

This was a retrospective consecutive study of all patients who underwent a rTKA using SMILES rotating hinge knee system between 2005 to 2018. Local study ethical approval was obtained. We use this system in elderly patients with collateral ligament failure, posterior capsular failure, or Anderson Orthopaedic Research Institute classification system III (AORI-III),11 with significant femoral condylar bone loss where augmented condylar revisions are inadequate to provide durable fixation. Patients were identified using a local prospective database and linkable data obtained from the NJR for rTKA. Demographic, clinical, and surgical data were collected from patients’ electronic health records. We excluded patients that required endoprostheses with distal femoral replacements.12,13 All patients underwent routine preoperative anaesthetic assessment and received a spinal anaesthetic with upper thigh sterile tourniquet and perioperative prophylactic antibiotics.

Implant

SMILES is a fully cemented hinged knee made from a cast cobalt-chromium-molybdenum and titanium alloy. The rotating-hinge articulation includes a bevelled polyethylene bearing surface placed on the tibial plate and limits rotational motion to ± 5°. Hyperextension is constrained by a bumper that acts as a secondary bearing surface (Figure 1). This system offers three tibial options in two sizes; a rotating-hinge all-polyethylene tibia, a rotating-hinge metal cased tibia with short (140 mm), and long stem (180 mm) options, and a fixed hinge tibia with short and long stems. In our practice, we use the rotating-hinge metal cased tibia. The femoral component is either small or standard size with a 140 mm long femoral curved titanium stem of 13 mm diameter for standard components and 12 mm for small components. The hinge mechanism is assembled with an axle, a pair of polyethylene bushes, and a titanium circlip.

Operative technique

Knees were approached through a standard medial parapatellar arthrotomy with subluxation of the patella following complete synovectomy. Components were then removed in the standard fashion. The knee was then reconstructed with flexion first approach.14 Tibial preparation was then performed, and the canal reamed to accept a cemented stem. Attention was then turned to femoral preparation. A trial was then assembled and the joint line level was restored in flexion and extension and checked using a combination of anatomical markers and soft-tissue tension, particularly extensor apparatus and length measurements, including patellofemoral articulation.15 Once satisfactory trial positioning was obtained, definitive implants were assembled and cemented using Palacos R+G cement (Hereaus Medical, Germany). Routine closure was then performed in layers over a drain which was removed in 24 hours. Full weightbearing was encouraged, as tolerated with routine physiotherapy. Follow-up was performed regularly at six weeks, three months, and 12 months thereafter.

Outcome measures

The primary outcome measure was implant survivorship and mechanical failure, which was defined as the presence of progressive radiolucent lines around either component on serial anteroposterior and lateral radiographs assessed by two authors (HEM, BVB), or the need for revision surgery for aseptic loosening. The secondary outcome measure was complications, particularly patellofemoral complications. Reoperations and revision for any cause were also collected. Death was identified through both local hospital electronic databases and linked data from the NJR/NHS Personal Demographic Service.

Statistical analysis

Values of all parameters are presented as percentages and Kaplan-Meier survival analysis were performed for implant survivorship using SPSS 16.0 software (SPSS, USA).

Results

There were 69 consecutive patients (69 knees) during the study period, which were all included in the analysis. These included 46 females (66.75%) and 23 males (33.25%) with a median age of 78 years (interquartile range (IQR) 70 to 82), and a mean BMI of 30 kg/m2 (27 to 33). The majority of patients (88.4%) had American Society of Anesthesiology (ASA) grade II/III. The indications were second stage reconstruction for infection (37.7%), and aseptic aetiology in the remining patients (62.3%) (Table I).

Table I.

Patients’ baseline characteristics and indication for surgery.

Variable Data
Patients/knees, n 69
Median age, yrs (IQR) 78 (70 to 82)
Female sex, n (%) 46 (66.7)
BMI, kg/m 2
Median (IQR) 30 (27 to 33)
Mean, (SD; range) 30.7 (5.3; 20 to 51)
ASA grade, n (%)
I 4 (5.8)
II 23 (33.3)
III 38 (55.0)
IV 4 (5.9)
Indications, n (%)
Septic 26 (37.7)
Two-stage 15 (21.7)
Single-stage 11 (16)
Aseptic 43 (62.3)
Loosening with ligamentous instability 22 (31.9)
Instability (incompetent MCL) 14 (20.3)
Subluxation/dislocation (posterior capsular failure) 6 (8.7)
Stiffness 1 (1.4)
  1. ASA, American Society of Anesthesiology; IQR, interquartile range; MCL, medical collateral ligament; SD, standard deviation.

The mean follow-up was 9.7 years with a median 8.4 years (4 to 18). The overall complication rate was 7.24%, all with patellofemoral complications (Table II), and have not had operative interventions. These were one case of extensor failure that required extensor mechanism reconstruction in an elderly frail patient who opted for nonoperative management. Similarly, three cases of patella instabilities were managed nonoperatively. Only four cases were revised (5.8%); infection in two cases, one periprosthetic fracture, and one case for mechanical failure and aseptic loosening of the femoral component (Figures 2 to 4; Table III). There were no other cases of mechanical failure or progressive radiological loosening around either the tibial or femoral components.

Table II.

Outcomes and ten-year implant survivorship.

Variables Data
Patients/knees, n 69
Mean follow-up, yrs (range) 9.7 (4 to 18)
Median follow-up, yrs (IQR) 8.4 (4.5 to 10.7)
Any cause revision, n (%) 4 (5.8)
Patellofemoral complications, n (%) 5 (7.24)
Patella dislocation/subluxation, n 4
Extensor mechanism failure, n 1
Ten-year implant survivorship (any cause revision), % 92.2
  1. IQR, interquartile range.

Fig. 2 
          a) Preoperative anteroposterior and lateral radiographs of left knee in a 79-year-old female with a fusion nail following first-stage revision for infection. b) Anteroposterior and lateral radiographs at three-year follow-up using Stanmore Modular Individualized Lower Limb System (SMILES) rotating hinge, with no loosening and satisfactory clinical outcomes.

Fig. 2

a) Preoperative anteroposterior and lateral radiographs of left knee in a 79-year-old female with a fusion nail following first-stage revision for infection. b) Anteroposterior and lateral radiographs at three-year follow-up using Stanmore Modular Individualized Lower Limb System (SMILES) rotating hinge, with no loosening and satisfactory clinical outcomes.

Fig. 3 
          a) Preoperative anteroposterior and lateral radiographs of a failed infected and dislocated total knee arthroplasty (TKA) in an 83-year-old female. b) Immediate postoperative, a) and two years follow-up; b-c) following revision TKA with a rotating hinge Stanmore Modular Individualized Lower Limb System (SMILES) prosthesis and satisfactory clinical and radiological outcomes.

Fig. 3

a) Preoperative anteroposterior and lateral radiographs of a failed infected and dislocated total knee arthroplasty (TKA) in an 83-year-old female. b) Immediate postoperative, a) and two years follow-up; b-c) following revision TKA with a rotating hinge Stanmore Modular Individualized Lower Limb System (SMILES) prosthesis and satisfactory clinical and radiological outcomes.

Fig. 4 
          a, b) Preoperative anteroposterior and lateral radiographs of right revision knee in a 75-year-old female with aseptic loosening and instability. c, d) Postoperative radiographs following re-revision total knee arthoplasty with a rotating hinge Stanmore Modular Individualized Lower Limb System (SMILES) prosthesis. b) Anteroposterior and lateral radiographs at 5.6 years follow-up with loosening around the femoral component with well-fixed tibial component. c, d) Anteroposterior and lateral radiographs following single component revision to a distal femoral replacement of the same system retaining the tibial component.

Fig. 4

a, b) Preoperative anteroposterior and lateral radiographs of right revision knee in a 75-year-old female with aseptic loosening and instability. c, d) Postoperative radiographs following re-revision total knee arthoplasty with a rotating hinge Stanmore Modular Individualized Lower Limb System (SMILES) prosthesis. b) Anteroposterior and lateral radiographs at 5.6 years follow-up with loosening around the femoral component with well-fixed tibial component. c, d) Anteroposterior and lateral radiographs following single component revision to a distal femoral replacement of the same system retaining the tibial component.

Table III.

Revision for any cause.

Patient no. Age, yrs;

M/F
BMI, kg/m2 ASA grade Indication Time to revision, yrs Cause of reoperation Outcome following re-revision
1 76; M 35 III Aseptic loosening, instability 1.1 Periprosthetic fracture 4.3 yrs/ no further surgery
2 60; M 43 III Second stage; PJI 1.5 PJI, repeat two-stage Implant-arthrodesis
3 72; M 29 III Second stage; PJI 7.9 PJI, repeat two-stage 4.6 yrs/ no further surgery
4 75; F 25 III Aseptic loosening of rTKA (Figure 4) 5.6 Femoral component aseptic loosening and failure 2 yrs/ no further surgery
  1. ASA, American Society of Anesthesiologists; DFR, distal femoral replacement; PJI, periprosthetic joint infection; rTKA, Revision total knee arthroplasty.

Survivorship analysis

At ten years, 17/69 patients (24.63%) had died. Implant survivorship analysis, using “revision for any cause” as an end point, was 92.2% at ten years, with estimated mean implant survivorship of 9.2 years (95% confidence interval (CI) 8.97 to 9.97) (Figure 5).

Fig. 5 
            Kaplan-Meier implant survivorship plot at ten years.

Fig. 5

Kaplan-Meier implant survivorship plot at ten years.

Discussion

In this study, we report satisfactory ten-year survivorship and low complication rate of SMILES rotating hinge prosthesis in rTKA patients. Although this implant has been in use for many years, little has been published on the long-term clinical outcomes in non-tumour patients. Here, rTKA patients are often elderly, female, and invariably with osteoporotic bone. A fully cemented prosthesis is therefore beneficial, offering the ability to immediately weightbear and mobilize.

In their meta-analysis of 17 studies of condylar rTKA, Wang et al16 found no significant differences in failure for any reason, reoperation, aseptic loosening, or infection between rTKA with cemented or cementless stem fixation. However, little is known about the effects of stem fixation on long-term outcomes of rotating hinge implants.17 To our knowledge, this is the largest series of this prosthesis in rTKA patients. Back et al18 reported on their series of SMILES prosthesis (29 patients/29 knees) over 4.5 years to 11 years. The authors reported one patient with evidence of aseptic loosening on radiographs and an overall failure rate of 13.8%. However, they reported a low rate of patellofemoral complications of 6%.

We also report a low patellofemoral complication rate of 7.24% over the entire 18-year study period, which were managed nonoperatively. In contrast, in their recent series of the S-ROM (DePuy Synthes, USA) rotating hinge in rTKA, Panesar et al19 reported their outcomes at a mean seven years’ follow-up. They reported a 26% complication rate, particularly with patellofemoral disorders and 19% revision rate. Notably, the S-ROM design has no inherent resistance to rotation, as the rotating hinge is delivered with a flat poly on a flat metal tibial tray, compared to the SMILES system, which does attempt to limit the degree of rotation, which can occur to ± 5° (Figure 1). It is possible that the increased rotational freedom, while having benefits for protecting tibial fixation, may increase the risk of patellofemoral instability.20 Further, the one case of mechanical failure in this study was around the femoral component in a previously multiply revised knee. Here, it appears that cement interdigitation into a sclerotic femoral bone was not achieved which led to early failure. However, the tibial component was well-fixed, which allowed for a single component revision into a distal femoral replacement with good outcome.

A number of studies on contemporary rotating-hinge implants in rTKA have been published, although most with short- to medium-term outcomes with various implants.9,20-25 Our study has long-term follow-up with low complication and revision rates for any cause. However, our study is limited by retrospective nature of its design and lack of clinical scores. In addition, our data was collected prospectively for all patients in a mandated national registry adding to its internal validity. Furthermore, our unit is a specialist tertiary centre with a multidisciplinary team approach ensuring standardization of care.

To conclude, in our experience, the SMILES rotating hinge knee achieves satisfactory long-term outcomes with ten-year implant survivorship of 92.2% and a low patellofemoral complication rate of 7.24%. The latter is achieved by ensuring central tracking of the patella with appropriate soft-tissue tension. There appears to be a protective role of the rotating hinge design in the SMIELS system against patellofemoral complications with its inherent resistance to rotation. However, further comparative studies are needed to further clarify role of insert design in rotating-hinge implants.


Correspondence should be sent to Hosam E. Matar. E-mail:

References

1. 18th UK National Joint Registry . 2021 . https://reports.njrcentre.org.uk/Portals/0/PDFdownloads/NJR%2018th%20Annual%20Report%202021.pdf ( date last accessed 18 September 2023 ). Google Scholar

2. Singh JA , Yu S , Chen L , Cleveland JD . Rates of total joint replacement in the United States: future projections to 2020-2040 using the national inpatient sample . J Rheumatol . 2019 ; 46 ( 9 ): 1134 1140 . Crossref PubMed Google Scholar

3. Matharu GS , Culliford DJ , Blom AW , Judge A . Projections for primary hip and knee replacement surgery up to the year 2060: an analysis based on data from the national joint registry for England, Wales, Northern Ireland and the Isle of Man . Ann R Coll Surg Engl . 2022 ; 104 ( 6 ): 443 448 . Crossref PubMed Google Scholar

4. Pasquier G , Ehlinger M , Mainard D . The role of rotating hinge implants in revision total knee arthroplasty . EFORT Open Rev . 2019 ; 4 ( 6 ): 269 278 . Crossref PubMed Google Scholar

5. Kouk S , Rathod PA , Maheshwari AV , Deshmukh AJ . Rotating hinge prosthesis for complex revision total knee arthroplasty: a review of the literature . J Clin Orthop Trauma . 2018 ; 9 ( 1 ): 29 33 . Crossref PubMed Google Scholar

6. Hintze J von , Niemeläinen M , Sintonen H , Nieminen J , Eskelinen A . Good mid-term outcome of the rotating hinge knee in primary total knee arthroplasty - results of a single center cohort of 106 knees with a median follow-up of 6.3 years . Knee . 2021 ; 28 : 273 281 . Crossref PubMed Google Scholar

7. Lettin AW , Deliss LJ , Blackburne JS , Scales JT . The Stanmore hinged knee arthroplasty . J Bone Joint Surg Br . 1978 ; 60-B ( 3 ): 327 332 . Crossref PubMed Google Scholar

8. Korim MT , Esler CNA , Reddy VRM , Ashford RU . A systematic review of endoprosthetic replacement for non-tumour indications around the knee joint . Knee . 2013 ; 20 ( 6 ): 367 375 . Crossref PubMed Google Scholar

9. Smith TH , Gad BV , Klika AK , Styron JF , Joyce TA , Barsoum WK . Comparison of mechanical and nonmechanical failure rates associated with rotating hinged total knee arthroplasty in nontumor patients . J Arthroplasty . 2013 ; 28 ( 1 ): 62 67 . Crossref PubMed Google Scholar

10. Rodríguez-Merchán EC . Total knee arthroplasty using hinge joints: indications and results . EFORT Open Rev . 2019 ; 4 ( 4 ): 121 132 . Crossref PubMed Google Scholar

11. Engh GA , Ammeen DJ . Bone loss with revision total knee arthroplasty: defect classification and alternatives for reconstruction . Instr Course Lect . 1999 ; 48 : 167 175 . PubMed Google Scholar

12. Matar HE , Bloch BV , James PJ . Outcomes of salvage endoprostheses in revision total knee arthroplasty for infection and aseptic loosening: experience of a specialist centre . Knee . 2021 ; 29 : 547 556 . Crossref PubMed Google Scholar

13. Matar HE , Bloch BV , James PJ . Distal femoral replacements for acute comminuted periprosthetic knee fractures: satisfactory clinical outcomes at medium-term follow-up . Arthroplast Today . 2021 ; 7 : 37 42 . Crossref PubMed Google Scholar

14. Matar HE , Bloch BV , Cameron HU , James PJ . Revision Total Knee Arthroplasty . In : Principles of Surgical Reconstruction: Back to the Beginning… Again. Revision Total Knee Arthroplasty: A Practical Guide . Cham : Springer Publishing , 2021 : 123 137 . Crossref Google Scholar

15. Hirschmann MT , Becker R . The Unhappy Total Knee Replacement . In : Hirschmann MT , Becker R . 52 Joint Line Restoration in Revision Total Knee Replacement . Cham : Springer International Publishing , 2015 : 631 638 . Crossref Google Scholar

16. Wang C , Pfitzner T , von Roth P , Mayr HO , Sostheim M , Hube R . Fixation of stem in revision of total knee arthroplasty: cemented versus cementless-a meta-analysis . Knee Surg Sports Traumatol Arthrosc . 2016 ; 24 ( 10 ): 3200 3211 . Crossref PubMed Google Scholar

17. Ogura K , Fujiwara T , Morris CD , Boland PJ , Healey JH . Long-term competing risks for overall and cause-specific failure of rotating-hinge distal femoral arthroplasty for tumour reconstruction . Bone Joint J . 2021 ; 103-B ( 8 ): 1405 1413 . Crossref PubMed Google Scholar

18. Back DL , David L , Hilton A , Blunn G , Briggs TWR , Cannon SR . The SMILES prosthesis in salvage revision knee surgery . Knee . 2008 ; 15 ( 1 ): 40 44 . Crossref PubMed Google Scholar

19. Panesar K , Al-Mouazzen L , Nessa L , Jonas SC , Agarwal S , Morgan-Jones R . Revision total knee arthroplasty using an uncemented metaphyseal sleeve, rotating hinge prosthesis: a case series of 99 patients . J Arthroplasty . 2021 ; 36 ( 6 ): 2121 2125 . Crossref PubMed Google Scholar

20. Matar HE , Bloch BV , James PJ . High ten-year implant survivorship and low patellofemoral complication rate for S-ROM rotating-hinge implants in revision total knee arthroplasty: a single-centre study . Bone Jt Open . 2022 ; 3 ( 3 ): 205 210 . Crossref PubMed Google Scholar

21. Baier C , Lüring C , Schaumburger J , et al. Assessing patient-oriented results after revision total knee arthroplasty . J Orthop Sci . 2013 ; 18 ( 6 ): 955 961 . Crossref PubMed Google Scholar

22. Cottino U , Abdel MP , Perry KI , Mara KC , Lewallen DG , Hanssen AD . Long-term results after total knee arthroplasty with contemporary rotating-hinge prostheses . J Bone Joint Surg Am . 2017 ; 99-A ( 4 ): 324 330 . Crossref PubMed Google Scholar

23. Farid YR , Thakral R , Finn HA . Intermediate-term results of 142 single-design, rotating-hinge implants: frequent complications may not preclude salvage of severely affected knees . J Arthroplasty . 2015 ; 30 ( 12 ): 2173 2180 . Crossref PubMed Google Scholar

24. Hossain F , Patel S , Haddad FS . Midterm assessment of causes and results of revision total knee arthroplasty . Clin Orthop Relat Res . 2010 ; 468 ( 5 ): 1221 1228 . Crossref PubMed Google Scholar

25. Joshi N , Navarro-Quilis A . Is there a place for rotating-hinge arthroplasty in knee revision surgery for aseptic loosening? J Arthroplasty . 2008 ; 23 ( 8 ): 1204 1211 . Crossref PubMed Google Scholar

Author contributions

H.E Matar: Conceptualization, Formal analysis, Methodology, Project administration, Writing – review & editing.

B. V. Bloch: Conceptualization, Methodology, Writing – review & editing.

P. J. James: Conceptualization, Methodology, Writing – review & editing.

Funding statement

The author(s) received no financial or material support for the research, authorship, and/or publication of this article.

ICMJE COI statement

B. V. Bloch discloses consulting fees from DePuy Synthes, Zimmer Biomet, and Ethicon; and payment or honoraria for lectures, presentations, speakers bureaus, manuscript writing, or educational events from DePuy Synthes and Ethicon, all of which is unrelated to this work.

Data sharing

All data generated or analyzed during this study are included in the published article and/or in the supplementary material.

Ethical review statement

Ethics approval was not required for this study.

Open access funding

The authors report that they received open access funding for this manuscript from Nottingham Hospital charities, UK.

Twitter

Follow B. V. Bloch @Bloch_ortho

Follow Nottingham Hip and Knee @NottsHipKnee

© 2023 Author(s) et al. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (CC BY-NC-ND 4.0) licence, which permits the copying and redistribution of the work only, and provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc-nd/4.0/