Early migration of a medially stabilized total knee arthroplasty

a radiostereometric analysis study up to two years



Download PDF

Abstract

Aims

Medial pivot (MP) total knee arthroplasties (TKAs) were designed to mimic native knee kinematics with their deep medial congruent fitting of the tibia to the femur almost like a ball-on-socket, and a flat lateral part. GMK Sphere is a novel MP implant. Our primary aim was to study the migration pattern of the tibial tray of this TKA.

Methods

A total of 31 patients were recruited to this single-group radiostereometric analysis (RSA) study and received a medial pivot GMK Sphere TKA. The distributions of male patients versus female patients and right versus left knees were 21:10 and 17:14, respectively. Mean BMI was 29 kg/m2 (95% confidence interval (CI) 27 to 30) and mean age at surgery was 63 years (95% CI 61 to 66). Maximum total point motions (MTPMs), medial, proximal, and anterior translations and transversal, internal, and varus rotations were calculated at three, 12, and 24 months. Patient-reported outcome measure data were also retrieved.

Results

MTPMs at three, 12, and 24 months were 1.0 mm (95% CI 0.8 to 1.2), 1.3 mm (95% CI 0.9 to 1.7), and 1.4 mm (0.8 to 2.0), respectively. The Forgotten Joint Score was 79 (95% CI 39 to 95) and Knee Injury and Osteoarthritis Outcome Score obtained at two years was 94 (95% CI 81 to 100), 86 (95% CI 75 to 93), 94 (95% CI 88 to 100), 69 (95% CI 48 to 88), and 81 (95% CI59 to 100) for Pain, Symptoms, Activities of Daily Living, Sport & Recreation, and Quality of Life, respectively.

Conclusion

In conclusion, we found that the mean increase in MTPM was lower than 0.2 mm between 12 and 24 months and thus apparently stable. Yet the GMK Sphere had higher migration at one and two years than anticipated. Based on current RSA data, we therefore cannot conclude on the long-term performance of the implant, pending further assessment.

Cite this article: Bone Jt Open 2021;2(9):737–744.

Take home message

The GMK Sphere showed good clinical scores, but had a higher short-term migration than anticipated.

Based on the radiostereometric analysis data we cannot conclude on the long-term performance yet, pending further assessment.

Introduction

Total knee arthroplasty (TKA) is a very common procedure for treating osteoarthritis of the knees in most countries, and incidence is expected to continue rising.1,2 However, not all patients are satisfied after a TKA.3 Several new implants have therefore been introduced to the market in recent decades, to meet patients’ increasing functional demands. The medial pivot (MP) category was introduced in the 1990s to mimic the kinematics of the native knee.4 The native knee is tight in the medial compartment, with a concave medial tibial plateau, and a circular medial femoral condyle fitting almost like a ball-on-socket. The lateral tibial plateau is rather flat. This, in addition to the laxity of the lateral collateral ligament (unlike the tightness of the medial collateral ligament), facilitates medial pivoting, lateral sliding, and a rolling motion of the joint during flexion and extension.5,6 The first generation of MPs consisted of The Medial Rotation Knee (1994, MRK; MatOrtho, UK) and the Advance Medial-Pivot Knee (1998, Wright Medical Group, USA).7-9 A second generation was later introduced with the SAIPH Knee System (2009, Matortho), Evolution Medial-Pivot Knee (2010, MicroPort Orthopaedics, USA), and the Global Medacta Knee Sphere (GMK Sphere) (2011, Medacta International, Switzerland)10-12 (Figures 1a and 1b). The latter uses an ultra high molecular weight polyethylene (UHMWPE) insert,13 made from Granular-UHMWPE-Ruhrchemie (GUR) 1020 and sterilized with ethylene oxide (EtO).14 Several smaller studies have shown good clinical results in terms of function, kinematics, and longevity of these implants.15-18 Our primary aim was to assess the migration pattern of the tibial tray of the GMK Sphere using radiostereometric analysis (RSA), and to compare this with previously known limits of safe migration patterns with respect to aseptic loosening. Secondary aims included wear, alignment, and clinical performance.

Fig. 1 
          a) Global Medacta Knee Sphere (GMK Sphere, anterior view) displaying the congruent medial and flat lateral parts of the insert designed to mimic the native knee motion during flexion and extension. Blue dots indicate position of feature points. b) The posterior view of the same implant displaying the congruent medial and flat lateral parts of the insert designed to mimic the native knee motion during flexion and extension.

Fig. 1

a) Global Medacta Knee Sphere (GMK Sphere, anterior view) displaying the congruent medial and flat lateral parts of the insert designed to mimic the native knee motion during flexion and extension. Blue dots indicate position of feature points. b) The posterior view of the same implant displaying the congruent medial and flat lateral parts of the insert designed to mimic the native knee motion during flexion and extension.

Methods

A single series of 31 consecutive patients was recruited at Oslo University Hospital, Ullevaal, Oslo, Norway. The study protocol is shown in Table I and Figure 2. All patients received a cemented GMK Sphere TKA using Refobacin Bone Cement R (Zimmer Biomet, USA). One of two experienced surgeons performed all surgeries between April 2016 and February 2018. All patients underwent the same operative procedure and postoperative protocol including a medial parapatellar approach, without tourniquets. During surgery, five to eight tantalum 1 mm beads (RSA Biomedical, Sweden) were inserted in tibial bone with a fair geometrical spread.

Table I.

Inclusion and exclusion criteria of the study.

Inclusion criteria
Patients with knee osteoarthritis
Exclusion criteria
Preoperative flexion contracture more than 15°
Preoperative limited range of motion under anaesthetics (less than 110°)
Less than 50 or more than 75 years of age at the time of surgery
Use of walking aids because of other musculoskeletal and neuromuscular problems
Preoperative diagnosis other than osteoarthritis and avascular necrosis (e.g. rheumatoid arthritis, tumours)
Revision arthroplasty
Obesity with BMI > 35
Impaired collateral ligaments
Postoperative revision surgery due to deep wound infection
Fig. 2 
          Flow chart of the study. *Unrelated to the study; this patient already had condition number (CN) > 120. ME, mean error; RSA, radiostereometric analysis.

Fig. 2

Flow chart of the study. *Unrelated to the study; this patient already had condition number (CN) > 120. ME, mean error; RSA, radiostereometric analysis.

Clinical evaluation

Baseline data such as age, sex, and BMI were retrieved. For clinical assessment, we used the Knee Injury and Osteoarthritis Outcome Score (KOOS)19 at baseline and all timepoints. The Forgotten Joint Score 12 (FJS-12)20 was retrieved at three and 24 months postoperatively. Degrees of flexion and valgus/varus alignment were recorded postoperatively using a manual goniometer. All complications were accounted for.

Conventional radiology

All patients had preoperative plain standing radiographs, including the hip-knee-ankle (HKA) exposures. These were repeated at three months postoperatively, together with a CT scan of the artificial joint. Valgus and varus knee angles were defined as positive and negative values respectively. Tibial tray rotation was evaluated using Berger’s method.21

RSA

Supine RSA radiographs were taken postoperatively within a week and thus before discharge, and at three, 12, and 24 months using fixed ceiling-mounted x-ray tubes (Proteus XR/A, GE Healthcare, USA and Canon Triathlon T3, Japan) and knee cage number 10 (UmRSA; RSA Biomedical). All patients had double RSA examinations once for precision purposes. All RSA images were analyzed using RSAcore (v. 4.1, the Netherlands) Model Based RSA software. The first author analyzed all images and migration was reported for translations and rotations in all planes, feature point motions, as well as maximum total point motion (MTPM). Left-sided RSA knees were converted to right-sided by multiplying the segmental x-translations and y-z-rotations by a factor of -1,22,23 while x-, y-, and z-translations and rotations were reported with signed values and categorized as medial, proximal, and anterior translations and transversal, internal, and varus rotations, respectively.

Our upper limits for condition number (CN) and mean error (ME) were 120 and 0.35 respectively.22,23 Computer-aided design models (CAD) of both femoral and tibial components for all sizes were obtained from the manufacturer. They were implemented in the RSAcore software with the feature points positioned as anteriorly, medially, laterally, posteriorly, tip (all tibia model), and centre of the medial condyle (femoral model only) (Figures 1a and 1b). Feature point translations of the tibia were calculated. The change in sagittal distance of the centre of the femoral condyle and medial tibia feature points over time was calculated in terms of wear of the polyethylene of the medial part of the insert. As the RSA radiographs were performed in supine position, only implants regarded as stable in the mediolateral direction, i.e. < 5° movement in any position, were included in this analysis.24

Statistical analysis

We used SPSS for Windows v. 25 (IBM, USA) for statistical analysis. Data were normally distributed unless otherwise stated and presented with means and 95% confidence intervals (CIs). When appropriate, a paired t-test was used, presenting p-values with a significance level < 0.05. Non-normally distributed data were presented with median and interquartile range (IQR), and p-values calculated using the Wilcoxon signed-rank test.

We regarded a within-group change of 0.2 mm in MTPM from 12 to 24 months as clinically relevant.25 With an α of 0.05 and a power of 80% and the assumption of a standard deviation (SD) of 0.3, we calculated that we needed a minimum of 20 patients.26 To account for loss to follow-up and the exclusion of some patients, we originally included 30 patients. One patient died unrelated to the study within one year, so one extra patient was included.

Precision of the RSA analyses was reported as SDs of the absolute mean of the difference of repeated measurements with 95% CI and thus reported as SD 1.96.23

Results

Clinical evaluation

Table II shows patient demographics. KOOS improved from preoperatively to two years and FJS-12 from three months to two years (Table III and Figure 3). KOOS scores of patients identified as high-risk, based on the RSA analyses, are shown in Table IV. There was no difference in flexion of the knees from preoperatively to two years (Table V).

Table II.

Demographic data.

Variable Total
Total, patients (knees) 31 (31)
Sex, M:F 21:10
Side, R:L 17:14
Smokers, n 6
Mean BMI, kg/m2 (95% CI) 29 (27 to 30)
Mean age, yrs (95% CI) 63 (61 to 66)
Mean operation time, mins (95% CI)* 118 (114 to 124)
Mean LOS, days (95% CI) 4.0 (3.6 to 4.3)
  1. *

    Data missing for one patient.

  1. CI, confidence interval; LOS, length of stay.

Table III.

Patient-reported outcome measures.

PROM Preop median (IQR) 3-mth median (IQR) 1-yr median (IQR) 2-yr median (IQR) Ceiling effect %* p-value
KOOS
Pain 44 (33 to 56) 72 (64 to 97) 94 (74 to 100) 94 (816 to 100) 38 < 0.001
Symptoms 50 (46 to 64) 68 (57 to 84) 89 (6 to 93) 86 (75 to 93) 3 < 0.001
ADL 52 (34 to 59) 84 (64 to 96) 94 (82 to 100) 94 (88 to 100) 8 < 0.001
Sport&Rec 10 (5 to 26) 55 (36 to 75) 70 (55 to 82) 69 (48 to 88) 3 < 0.001
QoL 189 (13 to 31) 63 (47 to 78) 88 (63 to 97) 81 (59 to 100) 38 < 0.001
FJS-12 N/A 60 (27 to 83) N/A 79 (39 to 95) 10% 0.002
  1. *

    Ceiling effect at two years.

  1. Wilcoxon signed-rank test comparing change from preoperative to two years.

  1. ADL, activities of daily living; FJS, Forgotten Joint Score; IQR, interquartile range; KOOS, Knee injury and Osteoarthritis Outcome Score; N/A, not applicable.

Fig. 3 
            Median Knee injury and Osteoarthritis Outcome Score (KOOS) from baseline to 24 months. ADL, activities of daily living, QoL, quality of life.

Fig. 3

Median Knee injury and Osteoarthritis Outcome Score (KOOS) from baseline to 24 months. ADL, activities of daily living, QoL, quality of life.

Table IV.

Clinical and radiological data of high-risk patients. preoperative and postoperative hip-knee-ankle angle and CT rotation (Berger) > 18° means that the implant is internally rotated. Varus and valgus knee hip-knee-ankle angles defined as negative and positive values, respectively.

Patient Pain Symptoms ADL Sport & Recreation QoL Preop HKA, ° Postop HKA, ° CT rotation, °
10 64 54 87 60 44 0.1 6.2 29.6
11 47 39 51 5 38 -7.4 -1.8 11.6
20 100 100 97 90 10 -9.7 -1.7 23.1
25 89 100 92 83 100 12.4 1.6 14.3
28 100 89 100 95 100 -7.2 1.6 9.4
38 100 75 100 90 100 -6.0 -2.7 3.6
  1. ADL, activities of daily living; HKA, hip-knee-ankle angle; QoL, quality of life.

Table V.

Flexion from preoperatively to two years.

Variable n Preop mean, ° (95% CI) 3mth mean, ° (95% CI) 1-yr mean, ° (95% CI) 2-yr mean, ° (95% CI) p-value*
Flexion 29 123 (118 to 128) 115 (109 to 120) 122 (119 to 127) 120 (117 to 124) 0.250
  1. *

    Paired t-test comparing change from preoperative to two years.

  1. CI, confidence interval.

Conventional radiology

Mean preoperative and postoperative HKA angles were -6° (95% CI -8.5 to -3) and -1° (95% CI -2.3 to -0.25) respectively (p < 0.001, paired t-test). Mean postoperative tibial rotation using the Berger technique21 was 15.1° (5% CI 12.0 to 18.1).

RSA

Precision calculated by repeated RSA examinations is displayed in Table VI. The mean difference and SD of MTPM were 0.01 and 0.39 respectively. Because not all of these were suitable for analysis, the total number of double examinations was only 14. The mean CN and ME for all examinations were 63.8 and 0.15. Segmental rotations and translations and mean and individual MTPMs are seen in Table VI and Figure 4 respectively. The mean MTPM increased the most before three months and towards one year. Subsequently, it seems stable. Otherwise, segmental mean translations and rotations all seem to be within the range of their respective precisions. The wear data are also shown in Table VI.

Table VI.

Migration, rotation, and wear of GMK Sphere.

Variable 3mths 1 yr 2 yrs Precision
Mean MTPM, mm (95% CI) 1.00 (0.78 to 1.21) 1.30 (0.94 to 1.67) 1.40 (0.84 to 1.96) 0.01 (0 to 0.77)
Translation, mm (95% CI)
Medial 0.01 (-0.08 to 0.09) 0.04 (-0.13 to 0.21) 0.05 (-0.14 to 0.24) 0.01 (-0.18 to 0.20)
Proximal -0.03(-0.10 to 0.04) -0.03(-0.16 to 0.1) -0.10(-0.34 to 0.13) 0.03 (-0.07 to 0.13)
Anterior 0.04 (-0.16 to 0.23) -0.01(-0.24 to 0.21) 0.06 (-0.11 to 0.22) 0.00 (-0.41 to 0.41)
Rotation,°(95% CI)
Transversal 0.09 (-0.19 to 0.38) -0.10(-0.33 to 0.12) -0.32(-0.80 to 0.15) 0.04 (-0.97 to 1.04)
Internal -0.37 (-0.71 to 0.02) 0.04 (-0.57 to 0.66) -0.39(-0.81 to 0.03) 0.06 (-1.19 to 1.3)
Varus 0.09 (-0.08 to 0.26) 0.09 (-0.21 to 0.39) 0.16 (-0.45 to 0.78) 0.01 (-0.24 to 0.27)
Median wear, mm (IQR) 0.03 (-0.23 to 0.21) -0.13(-0.48 to 0.25) 0.09 (-0.10 to 0.55) 0.116*
  1. *

    Wilcoxon signed-rank test comparing change in wear from three to 24 months.

  1. CI, confidence interval; IQR, interquartile range; MTPM, maximum total point motion.

Fig. 4 
            Individual and mean maximum total point motions (MTPMs) with error bars showing 95% confidence intervals.

Fig. 4

Individual and mean maximum total point motions (MTPMs) with error bars showing 95% confidence intervals.

Four of the patients had a higher than anticipated migration or transversal rotation (Table VII).

Table VII.

High-risk patients qualifying based on Ryd et al25 or Gudnason et al.27

Thresholds 10* 11 20 25 28 38
> 0.2 mm 12 to 24 months X X X X
Transversal rotation 24 months X
Peripheral distal translation > 0.6 mm X X X X X
Peripheral proximal translation > 0.9 mm X X
  1. *

    Revised after 32 months.

  1. Based on Ryd et al.

  1. Based on Gudnason et al.

Sagittal point movement of the tibial tray is depicted in Figure 5. An analysis of the peripheral distal or proximal translations did not reveal any specific migration pattern. However, we identified five implants with peripheral distal or proximal translation above 0.6 mm or 0.9 mm respectively (Table VII).27

Fig. 5 
            Peripheral distal and proximal sagittal translations of the different feature points over time.

Fig. 5

Peripheral distal and proximal sagittal translations of the different feature points over time.

Adverse events

One death occurred within one year, not study-related. One patient was revised due to aseptic loosening.

Discussion

Our main finding is that this implant migrated initially and then stabilized after three to 12 months. This concurs with the literature on early migration of cemented implants.25 A MTPM of 1.3 mm at 12 months puts the implant in the “at risk” category of Pijls et al.28 In our study, based on the one-year MTPM results, we cannot state whether there is a higher or lower risk of revision due to aseptic loosening of 5% at ten years.28 However, several implants with good long-term survivorship also fall into this category.29,30

Segmental motion, measuring the movement of the centre of the implant, often underestimates real migration. Peripheral feature points (Figure 1) give a better impression of the real implant movement as the dominant failure mechanism for tibial baseplates is tilting (rotation) rather than general subsidence.31 Furthermore, we could identify some individual high-risk implants based on previous studies by Ryd et al,25 using the strict continuous migration criteria, and by Gudnason et al,27 using the transversal rotation or proximal or distal peripheral translation of the feature points of the tibial tray. One of these implants was revised due to aseptic loosening of the tibial tray 32 months after surgery (Patient 10). The postoperative HKA angle of this patient revealed a valgus alignment of 6°. We therefore attribute the failure to surgical reasons rather than implant-related reasons. Another characteristic feature of the patients with high-risk implants was inferior clinical scores (Patient 11). This implant was well aligned in the coronal plane but was 7° externally rotated on the CT scan. This would probably not cause any clinical problems.32 The other patients had excellent KOOS scores at two years, implying no symptoms of aseptic loosening. Nevertheless, we know from the literature that symptoms of loose implants can take up to ten years to become apparent.25

It has been debated whether the use of a tourniquet is important for good fixation of implants. Several RSA studies have, however, proven that this is not the case.33-35 Another explanation of the somewhat high MTPM could be the cement used in the study. Refobacin Bone Cement R has been used for several years at our hospital, and several studies including registry and RSA studies suggest that this cement gives good fixation and long-term survivorship.36-38 The GMK Sphere has a shorter but wider wing of the keel than for instance the NexGen CR (Zimmer Biomet) and the Triathlon CR (Stryker, USA), both known for their excellent survivorship.39 In theory, this could increase the rotation of the implant. We found that the mean internal and transversal rotation of the implant is lower than the precision measured by double examinations, so we conclude that for that rotation it is unlikely.

It has been stated that the combination of early and continuous migration defines a specific migration pattern of each implant.40 Most cemented implants seem to have a migration pattern with a lower mean MTPM than the GMK Sphere.41 A recent study by van Hamersveld et al42 shows, however, that PS implants have a migration pattern with a higher mean MTPM than CR implants. MP implants are actually constrained implants medially. No authors have studied their natural migration pattern, but they may have a higher initial migration before stabilizing. If so, this could partly explain the somewhat higher migration found in the current data.

Several studies indicate good mid- to long-term results of medial pivot implants.16,43 One review article found similar or even better survivorship of the Advance MP, compared to other TKAs.44 Another article found no difference in survivorship at 13 years between MP and central mobile-bearing TKAs.45 In a recent review article, Cacciola et al46 found that primary MP implants in general provide overall mid-term survivorship comparable to standard cruciate-retaining and posterior-stabilized implants, according to the available data, and yield better high-end function than standard implants.

Most studies on the GMK Sphere focus on the implant’s kinematics,47,48 rather than survival. We did, however, publish a registry study in 2020 on implants from the Norwegian Arthroplasty Registry and the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR).49 In that study we found a hazard ratio (HR) of 2.0 (95% CI 1.5 to 2.6; p < 0.001) for revision of any cause of the GMK Sphere compared to the three most used minimally stabilized TKAs in the AOANJRR. There was also a higher HR for revision of the MP category due to malalignment, instability, and patella erosion in the AOANJRR, but we could not stratify this by brand.

Although our study is not powered to evaluate clinical results, the scores on the FJS50 and KOOS30,51 are consistent with other TKAs. The polyethylene wear over two years was not clinically relevant (p = 0.116, Wilcoxon signed-rank test).

According to the registries, the GMK Sphere has a cumulative five-year all-cause revision rate of 3.5% (95% CI 3.0 to 4.0) and 3.1% (95% CI 2.1 to 4.6) in the AOANJRR52 and the National Joint Registry53 respectively. This is higher than the < 3% revision thresholds set by the registries,28 and could be supported by our findings.

We did not find any static RSA studies on any implant from the MP design nor the GMK Sphere. Since this design has been on the market for over two decades, this is somewhat surprising. We agree with previous scholars that there should have been a phased or stepwise introduction of novel implants to the market.40,54,55 This study therefore fills a significant gap of knowledge in the literature.

One weakness is the number of patients included, as the small number does not account for the random distribution in the baseline data in the general population. This could affect the external validity of the study. RSA studies are costly, yet provide high precision.22 Our power calculation shows that the number of implants is sufficient to study the migration of the implant over time, as do several other previously published RSA studies.22,26 Although we had some dropouts, the final number of patients was sufficient with respect to the power calculation at all timepoints. Because our RSA radiographs were taken in supine position, and thus without weightbearing, we could only use the images in patients with knees regarded as stable in the mediolateral direction to assess polyethylene wear. The study by van Ijsseldijk et al24 suggests that this could be done, as they found no difference in wear between non-weightbearing examinations of stable knees and weightbearing examinations.

This study has several strengths. Firstly, we assessed both the MTMP and all six degrees of freedom, which complies with the ISO standard.23 Secondly, because of the feature points in the CAD models, we could study the peripheral point motions of the implants. We could thus identify implants at risk of mechanical loosening that would otherwise be regarded as stable. Thirdly, all the surgeries were performed by two experienced surgeons only. Given that this is a novel implant, with an anticipated learning curve, we believe this is a strength. Fourthly, numerous RSA studies have been published from this hospital.56-58 The staff are therefore experienced in using RSA technology.

In our study, one of 31 patients showed a clear migration pattern for mechanical loosening and was revised. This was probably due to non-implant-specific technical difficulties during the primary surgery, and we believe that the malalignment of the implant was a possible reason for early loosening.59

In conclusion, we found that the mean increase in MTPM was lower than 0.2 mm between 12 and 24 months and thus seems stable. However, the GMK Sphere had a higher total migration at one and two years than anticipated. Based on current RSA data, we therefore cannot conclude on the long-term performance of the implant, pending further assessment.


Correspondence should be sent to Frank-David Øhrn. E-mail:

References

1. Zhang Y , Jordan JM . Epidemiology of osteoarthritis . Clin Geriatr Med . 2010 ; 26 ( 3 ): 355 369 . Crossref PubMed Google Scholar

2. Klug A , Gramlich Y , Rudert M , et al. The projected volume of primary and revision total knee arthroplasty will place an immense burden on future health care systems over the next 30 years . Knee Surg Sports Traumatol Arthrosc . 2020 . Google Scholar

3. Gunaratne R , Pratt DN , Banda J , Fick DP , Khan RJK , Robertson BW , et al. Patient dissatisfaction following total knee arthroplasty: A systematic review of the literature . J Arthroplasty . 2017 ; 32 ( 12 ): 3854 3860 . Crossref PubMed Google Scholar

4. Hossain F , Patel S , Rhee S-J , Haddad FS , et al. Knee arthroplasty with a medially conforming ball-and-socket tibiofemoral articulation provides better function . Clin Orthop Relat Res . 2011 ; 469 ( 1 ): 55 63 . Crossref PubMed Google Scholar

5. Freeman MA , Pinskerova V . The movement of the normal tibio-femoral joint . J Biomech . 2005 ; 38 ( 2 ): 197 208 . Crossref PubMed Google Scholar

6. Pinskerova V , Johal P , Nakagawa S , et al. Does the femur roll-back with flexion? J Bone Joint Surg Br . 2004 ; 86-B ( 6 ): 925 931 . Crossref PubMed Google Scholar

7. No authors listed . Matortho: MRK CEM CS Fixed STD PE PAT UNCEM. Orthopaedic Data Evaluation Panel . https://www.odep.org.uk/product.aspx?pid=3302 (date last assessed August 15, 2021) Google Scholar

8. Anderson MJ , Becker DL , Kieckbusch T . Patellofemoral complications after posterior-stabilized total knee arthroplasty: a comparison of 2 different implant designs . J Arthroplasty . 2002 ; 17 ( 4 ): 422 426 . Crossref PubMed Google Scholar

9. No authors listed . Advance Medial-Pivot Knee System . Microport . https://www.microportortho.com/products/knees/primary-total-knee-systems/advance-knee-system ( date last accessed 12 August 2021 ). Google Scholar

10. No authors listed . MatOrtho: Saiph Cem CS fixed Std PE without patella . Orthopaedic Data Evaluation Panel . http://www.odep.org.uk/product.aspx?pid=9470 ( date last accessed 12 August 2021 ). Google Scholar

11. No authors listed . Medacta: GMK Sphere . Orthopaedic Data Evaluation Panel . http://www.odep.org.uk/product.aspx?pid=9545 ( date last accessed 12 August 2021 ). Google Scholar

12. No authors listed . Advance medial-pivot knee system: the design rationale . Wright Medical Technology . http://www.pacificsurgical.ph/uploads/Products/Advance%20Medial%20Pivot%20Total%20Knee%20Replacement/Advance%20Design%20Rationale%20(new)%20EMEA.pdf ( date last accessed 12 August 2021 ). Google Scholar

13. No authors listed . GMK Sphere: Stability for life . Mecdata . http://www.medacta.ca/EN/gmk-sphere-ca ( date last accessed 12 August 2021 ). Google Scholar

14. Cheppalli N , Metikala S , Albertson BS , Yaw K , et al. Plastics in total knee replacement: Processing to performance . Cureus . 2021 ; 13 ( 1 ): e12969 . Crossref PubMed Google Scholar

15. Dehl M , Bulaid Y , Chelli M , et al. Total knee arthroplasty with the Medial-Pivot knee system: Clinical and radiological outcomes at 9.5 years’ mean follow-up . Orthop Traumatol Surg Res . 2017 ; 104 ( 2 ): 185 191 . Google Scholar

16. Karachalios T , Varitimidis S , Bargiotas K , et al. An 11- to 15-year clinical outcome study of the Advance Medial Pivot total knee arthroplasty: pivot knee arthroplasty . Bone Joint J . 2016 ; 98-b ( 8 ): 1050 1055 . Crossref PubMed Google Scholar

17. Schmidt R , Ogden S , Blaha JD , et al. Midterm clinical and radiographic results of the medial pivot total knee system . Int Orthop . 2014 ; 38 ( 12 ): 2495 2498 . Crossref PubMed Google Scholar

18. Schmidt R , Komistek RD , Blaha JD , et al. Fluoroscopic analyses of cruciate-retaining and medial pivot knee implants . Clin Orthop Relat Res . 2003 ; 410 : 139 147 . Crossref PubMed Google Scholar

19. Roos EM , Roos HP , Lohmander LS , Ekdahl C , Beynnon BD . Knee Injury and Osteoarthritis Outcome Score (KOOS)--development of a self-administered outcome measure . J Orthop Sports Phys Ther . 1998 ; 28 ( 2 ): 88 96 . Crossref PubMed Google Scholar

20. Behrend H , Giesinger K , Giesinger JM , Kuster MS . the “forgotten joint” as the ultimate goal in joint arthroplasty . The Journal of Arthroplasty . 2012 ; 27 ( 3 ). Google Scholar

21. Berger RA , Crossett LS , Jacobs JJ , et al. Malrotation causing patellofemoral complications after total knee arthroplasty . Clin Orthop Relat Res . 1998 ; 356 : 144 153 . Crossref PubMed Google Scholar

22. Valstar ER , Gill R , Ryd L , et al. Guidelines for standardization of radiostereometry (RSA) of implants . Acta Orthop . 2005 ; 76 ( 4 ): 563 572 . Crossref PubMed Google Scholar

23. Implants for Surgery-Roentgen stereophotogrammetric analysis for the assessment of migration of orthopaedic implants . 2013 . https://www.iso.org/obp/ui/#iso:std:iso:16087:ed-1:v1:en (date last accessed August 15, 2021) Google Scholar

24. van Ijsseldijk EA , Valstar ER , Stoel BC , et al. Measuring polyethylene wear in total knee arthroplasty by RSA: differences between weight-bearing and non-weight-bearing positioning . J Orthop Res . 2014 ; 32 ( 4 ): 613 617 . Crossref PubMed Google Scholar

25. Ryd L , Albrektsson BE , Carlsson L , et al. Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses . J Bone Joint Surg Br . 1995 ; 77 ( 3 ): 377 383 . PubMed Google Scholar

26. Turgeon TR , Gascoyne TC , Laende EK , et al. The assessment of the stability of the tibial component of a novel knee arthroplasty system using radiostereometric analysis . Bone Joint J . 2018 ; 100-b ( 12 ): 1579 1584 . Crossref PubMed Google Scholar

27. Gudnason A , Adalberth G , Nilsson KG , et al. Tibial component rotation around the transverse axis measured by radiostereometry predicts aseptic loosening better than maximal total point motion . Acta Orthop . 2017 . Crossref PubMed Google Scholar

28. Pijls BG , Valstar ER , Nouta KA , et al. Early migration of tibial components is associated with late revision: a systematic review and meta-analysis of 21,000 knee arthroplasties . Acta Orthop . 2012 ; 83 ( 6 ): 614 624 . Crossref PubMed Google Scholar

29. Molt M , Ryd L , Toksvig-Larsen S . A randomized RSA study concentrating especially on continuous migration . Acta Orthop . 2016 ; 87 ( 3 ): 262 267 . Crossref PubMed Google Scholar

30. Kaptein BL , den Hollander P , Thomassen B , et al. A randomized controlled trial comparing tibial migration of the ATTUNE cemented cruciate-retaining knee prosthesis with the PFC-sigma design . Bone Joint J . 2020 ; 102-b ( 9 ): 1158 1166 . Crossref PubMed Google Scholar

31. Wojtowicz R , Henricson A , Nilsson KG , et al. Uncemented monoblock trabecular metal posterior stabilized high-flex total knee arthroplasty: similar pattern of migration to the cruciate-retaining design - a prospective radiostereometric analysis (RSA) and clinical evaluation of 40 patients (49 knees) 60 years or younger with 9 years’ follow-up . Acta Orthop . 2019 ; 90 ( 5 ): 460 466 . Google Scholar

32. Steinbrück A , Schröder C , Woiczinski M , et al. Influence of tibial rotation in total knee arthroplasty on knee kinematics and retropatellar pressure: an in vitro study . Knee Surg Sports Traumatol Arthrosc . 2016 ; 24 ( 8 ): 2395 2401 . Crossref PubMed Google Scholar

33. Ledin H , Aspenberg P , Good L . Tourniquet use in total knee replacement does not improve fixation, but appears to reduce final range of motion . Acta Orthop . 2012 ; 83 ( 5 ): 499 503 . Google Scholar

34. Ejaz A , Laursen AC , Jakobsen T , Rasmussen S , Nielsen PT , Laursen MB , et al. Absence of a tourniquet does not affect fixation of cemented TKA: A randomized RSA study of 70 patients . J Arthroplasty . 2015 ; 30 ( 12 ): 2128 2132 . Google Scholar

35. Molt M , Harsten A , Toksvig-Larsen S . The effect of tourniquet use on fixation quality in cemented total knee arthroplasty a prospective randomized clinical controlled RSA trial . Knee . 2014 ; 21 ( 2 ): 396 401 . Crossref PubMed Google Scholar

36. Ø B , Espehaug B , Havelin LI , et al. Bone cement product and failure in total knee arthroplasty . Acta Orthop . 2017 ; 88 ( 1 ): 75 81 . Crossref PubMed Google Scholar

37. Olerud F , Olsson C , Flivik G . Comparison of Refobacin bone cement and palacos with gentamicin in total hip arthroplasty: an RSA study with two years follow-up . Hip Int . 2014 ; 24 ( 1 ): 56 62 . Crossref PubMed Google Scholar

38. van der Voort P , Valstar ER , Kaptein BL , et al. Comparison of femoral component migration between Refobacin bone cement R and Palacos R + G in cemented total hip arthroplasty: A randomised controlled roentgen stereophotogrammetric analysis and clinical study . Bone Joint J . 2016 ; 98-b ( 10 ): 1333 1341 . Crossref PubMed Google Scholar

39. Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) . Hip, Knee & Shoulder arthroplasty: 2020 annual report, Adelaide; AOA . 2020 . https://aoanjrr.sahmri.com/annual-reports-2020 Google Scholar

40. Pijls BG , Nelissen RG . The era of phased introduction of new implants . Bone Joint Res . 2016 ; 5 ( 6 ): 215 217 . Crossref PubMed Google Scholar

41. Pijls BG , Plevier JWM , Nelissen R . RSA migration of total knee replacements . Acta Orthop . 2018 ; 89 ( 3 ): 320 328 . Crossref PubMed Google Scholar

42. van Hamersveld KT , Marang-van de Mheen PJ , Tsonaka R , Nilsson KG , Toksvig-Larsen S , Nelissen RGHH , et al. Risk factors for tibial component loosening: A meta-analysis of long-term follow-up radiostereometric analysis data . J Bone Joint Surg Am . 2021 ; 103 ( 12 ): 1115 1124 . Crossref PubMed Google Scholar

43. Bordini B , Ancarani C , Fitch DA . Long-term survivorship of a medial-pivot total knee system compared with other cemented designs in an arthroplasty registry . J Orthop Surg Res . 2016 ; 11 : 44 . Crossref PubMed Google Scholar

44. Fitch DA , Sedacki K , Yang Y . Mid- to long-term outcomes of a medial-pivot system for primary total knee replacement: a systematic review and meta-analysis . Bone Joint Res . 2014 ; 3 ( 10 ): 297 304 . Crossref PubMed Google Scholar

45. Jenny JY , Bercovy M , Cazenave A , et al. No difference in 13-year survival after medial pivot or central pivot mobile bearing total knee arthroplasty. A propensity matched comparative analysis . Knee Surg Sports Traumatol Arthrosc . 2020 . Crossref PubMed Google Scholar

46. Cacciola G , Mancino F , De Meo F , et al. Mid-Term survivorship and clinical outcome of the medial stabilized systems in primary total knee arthroplasty:A systematic review . J Orthop . 2021 ; 24 : 157 164 . Google Scholar

47. Scott G , Imam MA , Eifert A , et al. Can a total knee arthroplasty be both rotationally unconstrained and anteroposteriorly stabilised? A pulsed fluoroscopic investigation . Bone Joint Res . 2016 ; 5 ( 3 ): 80 86 . Crossref PubMed Google Scholar

48. Key S , Scott G , Stammers JG , et al. Does lateral lift-off occur in static and dynamic activity in a medially spherical total knee arthroplasty? A pulsed-fluoroscopic investigation . Bone Joint Res . 2019 ; 8 ( 5 ): 207 215 . Crossref PubMed Google Scholar

49. Øhrn FD , Ø G , Låstad Lygre SH , et al. Decreased survival of medial pivot designs compared with cruciate-retaining designs in TKA without patellar resurfacing . Clin Orthop Relat Res . 2020 ; 478 ( 6 ): 1207 1218 . Crossref PubMed Google Scholar

50. Cacciola G , Martino D , De Meo F . Does the medial pivot knee improve the clinical and radiographic outcome of total knee arthroplasty? A single centre study on two hundred and ninety seven patients . Int Orthop . 2020 ; 44 ( 2 ): 291 299 . Crossref PubMed Google Scholar

51. Molt M , Toksvig-Larsen S . Similar early migration when comparing CR and PS in Triathlon TKA: A prospective randomised RSA trial . Knee . 2014 ; 21 ( 5 ): 949 954 . Google Scholar

52. No authors listed . Australian Orthopaedic Association National Joint Replacement Registry Hip, Knee & Shoulder arthroplasty: 2020 Annual Report, Adelaide . Australian Orthopaedic Association . 2020 . https://aoanjrr.sahmri.com/documents/10180/689619/Hip%2C+Knee+%26+Shoulder+Arthroplasty+New/6a07a3b8-8767-06cf-9069-d165dc9baca7 ( date last accessed 12 August 2021 ). Google Scholar

53. No authors listed . 17th Annual Report 2020 . National Joint Registry . 2020 . https://reports.njrcentre.org.uk/Portals/0/PDFdownloads/NJR%2017th%20Annual%20Report%202020.pdf ( date last accessed 12 August 2021 ). Google Scholar

54. Malchau H . Introducing new technology: a stepwise algorithm . Spine (Phila Pa 1976) . 2000 ; 25 ( 3 ): 285 . Crossref PubMed Google Scholar

55. Nelissen RG , Pijls BG , Karrholm J , et al. RSA and registries: the quest for phased introduction of new implants . J Bone Joint Surg Am . 2011 ; 93 ( Suppl 3 ): 62 65 . Crossref PubMed Google Scholar

56. Flatoy B , Rohrl SM , Rydinge J , et al. Triple taper stem design shows promising fixation and bone remodelling characteristics: radiostereometric analysis in a randomised controlled trial . Bone Joint J . 2015 ; 97-b ( 6 ): 755 761 . Crossref PubMed Google Scholar

57. Ohrn FD , Van Leeuwen J , Tsukanaka M , et al. A 2-year RSA study of the vanguard CR total knee system: A randomized controlled trial comparing patient-specific positioning guides with conventional technique . Acta Orthop . 2018 . Crossref PubMed Google Scholar

58. Kibsgard TJ , Roise O , Stuge B , et al. Precision and accuracy measurement of radiostereometric analysis applied to movement of the sacroiliac joint . Clin Orthop Relat Res . 2012 ; 470 ( 11 ): 3187 3194 . Crossref PubMed Google Scholar

59. Sikorski JM . Alignment in total knee replacement . J Bone Joint Surg Br . 2008 ; 90 ( 9 ): 1121 1127 . Google Scholar

Author contributions

F-D. Øhrn: Methodology, Formal analysis, Writing – original draft.

Ø. B. Lian: Supervision, Writing – review & editing.

M. Tsukanaka: Methodology, Writing – review & editing.

S. M. Röhrl: Conceptualization, Methodology, Supervision, Writing – review & editing.

Funding statement

Although none of the authors has received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this article, benefits have been or will be received but will be directed solely to a research fund, foundation, educational institution, or other non-profit organization with which one or more of the authors are associated. F-D. Øhrn's PhD scholarship from Møre and Romsdal Health Trust covered the open access funding.

ICMJE COI statement

The authors declare an insitutional grant from Mecdata International, paid to the Orthopaedic Department of Ullevaal Hospital, related to this study. F-D. Øhrn declares a PhD scholarship from Møre and Romsdal Health Trust.

Acknowledgements

We would like to thank biostatistician Are H. Pripp, orthopaedic surgeon Bernhard Flatøy, and the staff at CIRRO, including research coordinator Marte Traae Magnusson and radiographers Silje Clausen, Alexis Hinojosa and Mona Risdal for their valuable participation in the study.

Ethical review statement

The study was approved by the Western Norway Regional Committee for Medical and Health Research Ethics (REC West, approval number 2014/1075). All patients gave written consent before inclusion. The study was registered at clinicaltrials.org with unique protocol identification 424444-1.

Twitter

Follow F. Øhrn @fdoehrn

© 2021 Author(s) et al. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (CC BY-NC-ND 4.0) licence, which permits the copying and redistribution of the work only, and provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc-nd/4.0/