Microindentation – a tool for measuring cortical bone stiffness?

A systematic review



Download PDF

Abstract

Objectives

Microindentation has the potential to measure the stiffness of an individual patient’s bone. Bone stiffness plays a crucial role in the press-fit stability of orthopaedic implants. Arming surgeons with accurate bone stiffness information may reduce surgical complications including periprosthetic fractures. The question addressed with this systematic review is whether microindentation can accurately measure cortical bone stiffness.

Methods

A systematic review of all English language articles using a keyword search was undertaken using Medline, Embase, PubMed, Scopus and Cochrane databases. Studies that only used nanoindentation, cancellous bone or animal tissue were excluded.

Results

A total of 1094 abstracts were retrieved and 32 papers were included in the analysis, 20 of which used reference point indentation, and 12 of which used traditional depth-sensing indentation. There are several factors that must be considered when using microindentation, such as tip size, depth and method of analysis. Only two studies validated microindentation against traditional mechanical testing techniques. Both studies used reference point indentation (RPI), with one showing that RPI parameters correlate well with mechanical testing, but the other suggested that they do not.

Conclusion

Microindentation has been used in various studies to assess bone stiffness, but only two studies with conflicting results compared microindentation with traditional mechanical testing techniques. Further research, including more studies comparing microindentation with other mechanical testing methods, is needed before microindentation can be used reliably to calculate cortical bone stiffness.

Cite this article: M. Arnold, S. Zhao, S. Ma, F. Giuliani, U. Hansen, J. P. Cobb, R. L. Abel, O. Boughton. Microindentation – a tool for measuring cortical bone stiffness? A systematic review. Bone Joint Res 2017;6:542–549. DOI: 10.1302/2046-3758.69.BJR-2016-0317.R2.

Article focus

  • To provide a comprehensive review on the use of microindentation for measuring cortical bone stiffness.

  • To assess whether microindentation can accurately measure cortical bone stiffness.

Key messages

  • Only two studies were found which directly compared microindentation with traditional mechanical testing methods.

  • These both used reference point indentation.

  • They showed contrasting results, however, and therefore it is currently unclear whether microindentation can accurately measure cortical bone stiffness.

Strengths and limitations

  • The study followed guidelines suggested by Cochrane and PRISMA organisations.

  • It reviews a mechanical testing technique which can further develop clinical practice.

Introduction

Knowledge of a patient’s bone stiffness would enable surgeons to determine the amount of force needed to be applied during impaction of an uncemented implant in joint replacement surgery. With the number of revision operations and periprosthetic fractures increasing each year,1 it is necessary to reduce any factor that may contribute to implant failure or fracture.

Stiffness is the ability of a structure to resist deformation, and is defined as the slope of the linear, reversible portion of the load-deformation plot.2 Stiffer bone undergoes less deformation than more compliant bone when subjected to a given load, and hence more force may be needed in order to seat an implant in a patient who has stiffer bone.3 However, when seated in stiffer bone, the implant will be less likely to loosen because the radial stresses effectively gripping the implant and known as the “elastic grip”,4 will be greater for a given amount of elastic deformation.5 Similarly, in less stiff bone, less force is required to seat the implant, but loosening is more likely as the radial stresses are weaker for the same amount of elastic deformation.5 In the elderly population, the structural stiffness and fracture resistance of bone is reduced due to increasing porosity.6,7 This causes a looser fit for a given amount of elastic deformation, and leads to difficulty in achieving a safe press fit without causing a fracture.8 The elastic interaction between the bone and implant is also critical in ensuring good bony ingrowth.9

It is therefore important to understand the mechanical properties of bone so that implants with the appropriate mechanical properties can be chosen for each patient in order to improve implant longevity.10 Currently, mechanical compression testing is used experimentally to measure bone stiffness,11 but cannot feasibly be performed in vivo. Dual-energy radiograph absorptiometry (DEXA) has been used to assess stiffness,12 although it is not sensitive enough to be used as a clinical tool on its own.13 This is mainly due to cortical bone porosity being fundamental to bone stiffness13 and DEXA does not have the spatial resolution to detect these pores.6

One method that does have the potential to measure bone stiffness is microindentation. Traditional indentation testing involves pressing a hard tip with a known force into a material, and measuring directly or indirectly the contact area. The contact area is usually estimated from the imprint created by the tip on the material, and hardness is defined as the force divided by this area.14 Indentation tips may come in several shapes: spherical; three-sided (Berkovich); or four-sided pyramidal (Vickers) (Fig. 1).14 Macroindentation has been used since the mid-20th century, and recently micro- and nanoindentation methods have been developed.14 Nanoindentation measures the mechanical properties of bone at the level of trabeculae or osteons,15 and microindentation has the potential to measure bone properties at the millimetre-scale level.16

 
          Image on the left: Berkovich tip interacts with cortical bone; right: enlarged view of Berkovich tip; bottom: Spherical tip (~750 µm radius) interacts with cortical bone.

Fig.

Image on the left: Berkovich tip interacts with cortical bone; right: enlarged view of Berkovich tip; bottom: Spherical tip (~750 µm radius) interacts with cortical bone.

Microindentation of bone can broadly be divided into two categories: traditional depth-sensing indentation and the more recently developed reference point indentation (RPI). Traditional indentation, commonly used for macroindentation, involves pressing a sharp, hard tip into a material and measuring the residual hardness impression under a microscope.14 Depth-sensing microindentation was developed so that the hardness and modulus of a material could be calculated by indenting it, but without having to use a microscope to measure the resulting hardness impression.17 It requires a carefully calibrated machine, which must be secured to a surface. From the load displacement data, the elastic modulus and hardness can be calculated with various techniques, including the Oliver-Pharr method.17

Conversely, RPI can be undertaken with a handheld device which uses a secondary probe as a reference point, as it is not fixed to a surface.18 RPI does not calculate stiffness or hardness values, but produces different outputs that are exclusive to the technique, and are described below.

The stiffness of cortical bone is important in orthopaedic surgery, and a reliable method of assessing cortical bone stiffness in patients is needed. Hence, a systematic review of the literature has been conducted to answer the following question: can microindentation accurately measure the stiffness of human cortical bone?

Materials and Methods

A systematic review of published literature relating to microindentation of cortical bone was undertaken using Medline, Embase, Cochrane, PubMed and Scopus databases up to November 2016. A combination of the search terms ‘microindentation’, ‘reference point indentation’, ‘indentation (micro)’, ‘bone’, ‘compact bone’, ‘cortical bone’, ‘elastic modulus’, ‘Young’s modulus’, ‘elasticity’, ‘rigidity’ and ‘stiffness’ was used. Exclusion criteria were as follows: studies only using nanoindentation or cancellous bone; studies on animal tissue only; foreign language papers; and papers where the full text could not be accessed. The first two authors selected articles for review, and any disagreements as to whether a paper should be included were resolved via discussion as recommended by the Cochrane Collaboration’s guidelines.19

Results

Search results

Figure 2 shows the systematic review flowchart, according to the Preferred Reporting Items for Systematic Review and Meta-Analyses (PRISMA) guidelines.20 A total of 1076 abstracts were retrieved, as well as 18 from other sources, amounting to 1094 abstracts to be reviewed. There were 132 abstracts remaining, which after screening and removal of duplicates, left 90 eligible full-text papers. Following review of these 90 papers, 32 papers were deemed to fit the inclusion criteria for testing cortical bone using microindentation. A total of 20 of the studies used RPI (BioDent = 13 and OsteoProbe = 7, both produced by Active Life Scientific, Santa Barbara, California) and 12 involved conventional depth-sensing microindentation.

Fig. 2 
            A flow diagram showing the search strategy, inclusion and exclusion criteria.

Fig. 2

A flow diagram showing the search strategy, inclusion and exclusion criteria.

Qualitative assessment

Only two papers directly compared microindentation measurements with traditional bulk tissue compression testing and both these studies used RPI.21,22 Papers were also retrieved which compared nanoindentation measures of bone mechanical properties with bulk tissue testing measures.23 As this review is focused on microindentation rather than nanoindentation, these latter papers will not be discussed further. Table I16,21,22,24-52 displays details of each included study. When extracting results, values were only given for parameters relevant to stiffness of bone in order to comply with PRISMA guidelines.20

Table I.

Papers included in qualitative synthesis

Author Bone used Indentation machine Indenter tip (diameter) Settings Outcomes (mean)
Abraham et al24 Tibia (cortical) BioDent (Active Life Scientific, Santa Barbara, California) 90° cono-spherical (375 mm) and 2.5 nm tip 20 cycles with a peak force of 10 N IDI
Beutel and Kennedy25 Tibia (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip 20 cycles with peak force 8 N TID, IDI, US, US 1st, LS and ED
Boivin et al26 Ilium (whole bone) MicroMet 5104 (Buehler, Lake Bluff, Illinois) Vickers Does not specify Microhardness
Coutts et al27 Femoral neck (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip 10 cycles with peak force 10 N TID, IDI, CID
Dall'Ara et al16 Vertebra (whole bone) Nano Hardness Tester (NHT; CSM Instruments, SA) Berkovich Depth = 2.5 nm. Displacement = 120 mN/min Elastic modulus and hardness
Dall'Ara et al28 Vertebra (whole bone) Nano-hardness tester Berkovich Depth = 2.5 nm. Displacement = 60 nm/min Stiffness (14.6 MPa in axial direction, 12.3 MPa in circumferential direction and 8.3 MPa in radial direction)
Diez-Perez et al29 Tibia (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip 20 cycles with a peak force of 11N IDI, TID, CID
Duarte Sosa et al30 Femur (cortical) OsteoProbe (Active Life Scientific Santa Barbara, California) 90° conical tip (375 nm) and tip radius < 10 nm OsteoProbe protocol BMSi
Farr et al31 Tibia (cortical) OsteoProbe 90° conical tip (375 nm) and tip radius < 10 nm OsteoProbe protocol BMSi
Granke et al21 Femur (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip Variable depth. 20 cycles with peak force of 10 N per cycle IDI and TID and elastic modulus (variable)
Granke et al32 Femoral midshaft (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip 20 cycles with peak force of 10 N per cycle Fracture toughness, TID, IDI, ED and LS
Grant et al33 Vertebra (endplate) Dynact Model I-PP3-B5 (Dynact Inc., Orchard Park, New York) Hemi-spherical (3 mm) Depth = 3 mm. Displacement = 0.2 mm/s Elastic modulus (multiple)
Grant et al34 Vertebra (endplate) Dynact Model I-PP3-B5 Hemi-spherical (3 mm) Depth = 3 mm. Displacement = 0.2 mm/s Elastic modulus (multiple) and strength
Güerri-Fernandez et al35 Tibia (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip  20 cycles with peak force 11N TID, IDI, CID
Hansma et al36 Tibia (cortical) OsteoProbe 90° conical tip (375 nm) and tip radius < 10 nm OsteoProbe protocol IDI
Jasiuk37 Does not specify BioDent 90° cono-spherical (375 mm) and 2.5 nm tip Does not specify Elastic modulus (100 to 300 MPa)
Jenkins et al38 Femoral neck (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip 10 cycles with peak force of 10 N IDI, TID, CID
Johnston et al39 Tibia (cortical) Instron 8874 (Instron Corporation, Norwood, Massachusetts) Flat-ended tip (3.5 mm) Depth = max 0.5 mm. Displacement = 2 mm/min, max load 250 N Elastic modulus (452 MPa)
Katsamenis et al40 Femur (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip Variable depth. 10 cycles with maximum peak force of 2 N IDI, TDI, creep indentation cycle and mean ED
Kerrigan et al41 Patella (cortical) BN23 (Industrial Devices Corporation, Petaluma, California (motor)); and Model 31 (Honeywell Sensotec, Columbus, Ohio (load cell)) Spherical (2 mm/6.5 mm) Depth = 0.1 mm/0.65 mm Elastic modulus (325 MPa (2 mm diameter tip)/206 MPa 6.5 mm diameter tip))
Krege et al22 Femur (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip Variable depth.10 cycles with peak force 10 N TID, IDI
Malgo et al42 Tibia (cortical) OsteoProbe 90° conical tip (375 nm) and tip radius < 10 nm OsteoProbe protocol BMSi
Mellibovsky et al43 Tibia (cortical) OsteoProbe 90° conical tip (375 nm) and tip radius < 10 nm OsteoProbe protocol BMSi
Milovanovic et al44 Femoral neck (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip 10 cycles with peak force 2 N TID, IDI, CID, ED
Milovanovic et al45 Femur (cortical) BioDent 90° cono-spherical (375 mm) and 2.5 nm tip 10 cycles with peak force 10 N ID, TID, ED
Mimar et al46 Glenoid (cortical) Shimadzu Autograph (Shimadzu Corporation, Kyoto, Japan) Cylindrical (2.5 mm) Depth = 3 mm. 2 mm/min displacement Strength (26 to 27 MPa)/elastic modulus (119 to 234 MPa)
Mirzaali et al47 Femur (cortical) Ultra Nano Hardness Tester (UNHT; CSM Instruments, SA) Berkovich Depth = 1 μm. 100 mN/min displacement and 400 mN/min unloading rate Elastic modulus (multiple) and hardness (multiple)
Noshchenko et al48 Vertebra (cortical) Instron 1321 servo-hydraulic test machine (Instron) Hemi-spherical (3 mm) Depth = 1.8 mm. 0.2 mm/s displacement Elastic modulus (74.8 N/ mm), end plate density/thickness and trabecular separation
Oxland et al49 Vertebra (endplate) Dynact Model I-PP3-B5 Hemispherical (3 mm) Depth = 3 mm. Displacement = 0.2 mm/s Elastic modulus (multiple)
Rudäng et al50 Tibia (cortical) OsteoProbe 90° conical tip (375 nm) and tip radius < 10 nm OsteoProbe protocol BMSI
Tan et al51 Vertebra (cortical) Instron 8874 Kidney/ elliptical/ cloverleaf Depth = 20% of vertebral height. Displacement = 0.2 mm/s Elastic modulus (396 to 805 N/mm)
Thurner et al52 Vertebra and tibia (whole bone) OsteoProbe 90° conical tip (375 nm) and tip radius < 10 nm OsteoProbe protocol Elastic modulus (112 and 49 MPa after NaF)
  1. BMSi, bone material strength index; CID, creep indentation distance; ED, energy dissipation; IDI, indentation distance increase; LS, loading slope; TID, total indentation distance; US, mean unloading slope; US 1st, first cycle unloading slope; BMD, bone mineral density; MPa, Megapascal; NaF, Sodium fluoride

Discussion

Microindentation, which has two main categories: depth-sensing microindentation; and RPI, has been used to evaluate cortical bone stiffness. However, in this systematic review we found only two studies that have validated this technique by comparing microindentation of human bone with other mechanical testing methods,21,22 both of which used RPI but had contrasting results.

Currently, there are two instruments using RPI, the BioDent and the OsteoProbe. The BioDent produces parameters such as indentation distance increase (IDI: the difference in depth between the first and last indentation), total indentation distance (TID), and creep indentation distance (CID: progressive indentation distance during the stable force phase of the first indentation cycle at the maximum 10 N force).18 In contrast, the OsteoProbe has an output, referred to as bone material strength index (BMSi).18

The BioDent uses one of three 700 mm diameter reference probes which differ in their tip morphology: a tri-bevelled surface (BP1); a bevelled surface with a blunted end (BP2); or a flat concentric surface (BP3). Based on the manufacturer’s recommendations, BP1 probes are ideal for samples with intact soft tissue because the probe can be used to scrape the soft tissue away from the test site, which is important for in vivo studies. BP2 probes are used for ex vivo work on large bones, and BP3 probes are for small animal work. Within each of the three reference probes is a similar test probe (375 mm in diameter, 90° cono-spherical, 2.5 mm tip radius).

Interpretation of RPI values has proven to be problematical. Although no direct comparative studies have been undertaken, it is generally accepted that a higher BMSi value from the OsteoProbe indicates better bone mechanical properties.18 However, there is disagreement regarding the BioDent, with only two studies directly comparing the BioDent with traditional mechanical testing. Granke et al21 suggested that bone toughness was inversely related to RPI measurements (IDI and TID), but Krege et al22 contradicted this when they demonstrated that chemically treated bone, that is by demineralisation, drying, or ashing, was shown to have weaker bone toughness when tested mechanically, but had a decreased IDI and TID. This could be due to the inability of the BioDent to generate enough force to damage the tissue. Furthermore, chemical treatment of the bone may not have affected the bone uniformly, and a greater effect may have been found on the surface where the microindentation was carried out. This may have affected results obtained from microindentation more than those from the mechanical testing, and highlights the complex task of comparing RPI measurements with traditional mechanical testing. Both of these studies21,22 tested bone in a hydrated state.

Several reports have focused on interpreting RPI parameters. Two, which used the OsteoProbe, have assessed the relationship between BMSi and occurrence of fracture in patients. However, these studies produced differing results, with one finding no significant correlation50 while the other reporting that patients with a fragility fracture had significantly reduced BMSi compared with patients with no fracture.53

Another important issue with the OsteoProbe is the possible harm it may cause to bone. It has been shown that microcracks form during testing,54 although it is not yet known what effect this may have on a patient’s bone when testing in vivo. Studies have shown no difference in TID between cycles, and that a large part of IDI is achieved in the first cycle which is not affected by increasing the number of cycles. Beutal and Kennedy,25 however, suggested that 6 N to 8 N of ten to 25 cycles was sufficient for testing, and does not harm bone. Therefore in fragile bone, future studies will need to focus on this point.

Although this review concentrated on human bone, it is worth reviewing results from animal studies. Using RPI, one study on dog and rat bone compared microindentation with three-point bending and axial compression and concluded that IDI was inversely correlated to toughness, as calculated by mechanical testing.55 A canine study investigating raloxifene, a drug which is used in the treatment of osteoporosis, showed that increasing concentrations of raloxifene significantly decreased IDI compared with untreated dogs.56 As previous studies have demonstrated that raloxifene increases bone toughness,57 IDI may therefore be inversely correlated with toughness.

The level of bone hydration has been shown to affect its mechanical properties. The elastic modulus of vertebral cancellous bone, calculated by microindentation, was reduced when the bone was tested hydrated rather than dry.55 This is in agreement with nanoindentation studies, which have shown that bone has a lower elastic modulus in a hydrated compared with a dry state.56 Hence, studies comparing indentation with other forms of mechanical testing need to perform both tests in the same state of hydration in order to ensure a fair comparison.

The direction of indentation, which requires taking into consideration the heterogeneity of bone, is an important variable during microindentation. One study showed that the mechanical properties of vertebral bones are affected by both the indentation direction and region, where it was found that the elastic modulus was higher axially then circumferentially25 confirming previous studies undertaken using nanoindentation methods.57

RPI has been shown to be sensitive to tissue organisation. For example, stiffness is greater in the longitudinal axis than in the transverse direction.21 Coutts et al27 measured the level of heterogeneity in RPI to see whether analysing a small part of the bone is representative of the whole. They found that sites on the same bone yield different values, and therefore in order to assess bone mechanical properties at a region of interest accurately, testing must be undertaken at that specific region.

The depth of the indenter is another factor that needs to be considered. Indentation measurements are sensitive to surface roughness at a lamellar level and testing at a higher depth avoids lamellar heterogeneity. If one orientation of lamellae is favoured, there is a risk of overestimating the mechanical behaviour of bone, thus by indenting several lamellae simultaneously, these risks are reduced. Mechanical properties of lamellae vary significantly across the osteon and thus measuring multiple lamellae provides a mean measurement of the overall bone tissue.58

The size of indenter tip is also important. One study compared large and small spherical indenter tips (2 mm and 6.5 mm in diameter), reaching differing depths (0.1 mm and 0.65 mm), and resulting in different elastic moduli (206 MPa and 325 MPa, respectively).41 Increased stiffness is thought to be due to the smaller indenter making contact with the denser bone near to the surface, while the larger indenter made contact with lower density areas deeper in the bone structure.41 The larger indenter tip engages with more of the porous bone, which reduces the apparent modulus.59

In addition, when attempting to measure the elastic modulus of a material, plastic or permanent deformation should be avoided. Microindentation using a sharp tip will cause some plastic deformation of the bone especially at the tip, which may result in calculating the elastic modulus of damaged bone, rather than unaffected bone.60 Oyen59 suggested that this can be overcome using a spherical indenter, where inconsequential plastic deformation occurs if the indentation strain is of a lower value than the yield strain of the bone.

Although indentation was originally designed to measure the hardness of a material, it is also possible to calculate the elastic modulus from the data collected. Most studies use the technique developed by Oliver and Pharr17 which calculates the elastic modulus from the unloading curve of the load-displacement graph. One disadvantage of using the Oliver-Pharr method is that the technique assumes the unloading response is purely elastic. Due to the time-dependent behaviour of bone, unloading is considered viscoelastic, and some studies have tried to correct for this by introducing a creep hold at peak load. This could still affect results, however, and a model that takes into account the viscoelastic properties of bone needs to be considered.59

One of the main obstacles facing microindentation as a testing method is recognising what part of the complex hierarchical structure of bone is being tested. Due to the porous nature of cortical bone at the millimetre scale, the apparent elastic modulus at the millimetre scale will be less than the modulus at a smaller material scale, where pores will not affect its stiffness. By combining indentation with high-resolution imaging, displaying the porosity and structure of the bone, a more accurate measurement of the whole bone structure may be calculated. This has been demonstrated by Hengsberger et al,23 who combined nanoindentation with synchrotron CT, demonstrating a good prediction of stiffness as measured by traditional mechanical testing.

When trying to assess the millimetre-scale elastic properties of bone using indentation, based on this review of the literature, we would suggest using a large spherical indenter tip, as the spherical tip causes less plastic deformation and the larger tip includes some of the porosity of cortical bone at this length-scale.

In conclusion, microindentation has been used to measure bone mechanical properties such as stiffness and toughness, producing several contrasting results. Indenter tip size, indentation depth and method of analysis are among the factors that affect indentation results. Once these variables have been fully evaluated and put in place, including using large spherical indenter tips, indentation should be assessed against other mechanical testing methods so that the reliability and reproducibility of the technique can be determined. Microindentation may be able to provide clinicians with vital bone quality information, potentially allowing a surgeon to customise the operative technique and implant to suit the mechanical properties of an individual patient’s bone.


O. Boughton; email:
Author Contribution

M. Arnold: Study retrieval, Review write up.

S. Zhao: Study retrieval, Review edit.

S. Ma: Image creation, Review edit.

F. Giuliani: Review edit.

U. Hansen: Review edit.

J. P. Cobb: Review edit.

R. L. Abel: Review edit.

O. Boughtonl: Idea of review, Review write up, Review edit.


We gratefully acknowledge R. Jones at Charing Cross Hospital Library (Imperial College London) for her help with the literature search, and the UK National Institute for Health Research (NIHR) and The Michael Uren Foundation for funding this research.

  • Funding Statement

    None declared

  • Conflicts of Interest Statement

    None declared

  • References

    1 Vanhegan IS , Malik AK , Jayakumar P , Ul Islam S , Haddad FS . A financial analysis of revision hip arthroplasty: the economic burden in relation to the national tariff. J Bone Joint Surg [Br]2012;94-B:619-623.CrossrefPubMed Google Scholar

    2 Miles AW , Gheduzzi S . Basic biomechanics and biomaterials. Surg2009;27:90-95. Google Scholar

    3 Squire M , Griffin WL , Mason JB , Peindl RD , Odum S . Acetabular component deformation with press-fit fixation. J Arthroplasty2006;21(Suppl 2):72-77.CrossrefPubMed Google Scholar

    4 Macdonald W , Carlsson LV , Charnley GJ , Jacobsson CM . Press-fit acetabular cup fixation: principles and testing. Proc Inst Mech Eng H1999;213:33-39.CrossrefPubMed Google Scholar

    5 Winter W , Karl M . Basic considerations for determining the amount of press fit in acetabular cup endoprostheses as a function of the elastic bone behavior. Biomed Tech (Berl)2014;59:413-420.CrossrefPubMed Google Scholar

    6 Zebaze RMD , Ghasem-Zadeh A , Bohte A , et al. Intracortical remodelling and porosity in the distal radius and post-mortem femurs of women: a cross-sectional study. Lancet2010;375:1729-1736.CrossrefPubMed Google Scholar

    7 Burr DB . Cortical bone: a target for fracture prevention?Lancet2010;375:1672-1673.CrossrefPubMed Google Scholar

    8 Sharkey PF , Hozack WJ , Callaghan JJ , et al. Acetabular fracture associated with cementless acetabular component insertion: a report of 13 cases. J Arthroplasty1999;14:426-431.CrossrefPubMed Google Scholar

    9 Macdonald W , Carlsson LV , Charnley GJ , Jacobsson CM , Johansson CB . Inaccuracy of acetabular reaming under surgical conditions. J Arthroplasty1999;14:730-737.CrossrefPubMed Google Scholar

    10 Huiskes R , Weinans H , van Rietbergen B . The relationship between stress shielding and bone resorption around total hip stems and the effects of flexible materials. Clin Orthop Relat Res1992;274:124-134.PubMed Google Scholar

    11 Keaveny TM , Pinilla TP , Crawford RP , Kopperdahl DL , Lou A . Systematic and random errors in compression testing of trabecular bone. J Orthop Res1997;15:101-110.CrossrefPubMed Google Scholar

    12 Liu XS , Cohen A , Shane E , et al. Bone density, geometry, microstructure, and stiffness: relationships between peripheral and central skeletal sites assessed by DXA, HR-pQCT, and cQCT in premenopausal women. J Bone Miner Res2010;25:2229-2238.CrossrefPubMed Google Scholar

    13 Granke M , Grimal Q , Saïed A , et al. Change in porosity is the major determinant of the variation of cortical bone elasticity at the millimeter scale in aged women. Bone2011;49:1020-1026.CrossrefPubMed Google Scholar

    14 Zysset PK . Indentation of bone tissue: a short review. Osteoporos Int2009;20:1049-1055.CrossrefPubMed Google Scholar

    15 Lewis G , Nyman JS . The use of nanoindentation for characterizing the properties of mineralized hard tissues: state-of-the art review. J Biomed Mater Res B Appl Biomater2008;87:286-301.CrossrefPubMed Google Scholar

    16 Dall’Ara E , Schmidt R , Zysset P . Microindentation can discriminate between damaged and intact human bone tissue. Bone2012;50:925-929.CrossrefPubMed Google Scholar

    17 Oliver WC , Pharr GM . An improved technique for determining hardness and elastic modulus using load and displacement sensing indentation experiments. J Mater Res1992;7:1564-1583. Google Scholar

    18 Allen MR , McNerny EM , Organ JM , Wallace JM . True gold or pyrite: a review of reference point indentation for assessing bone mechanical properties in vivo. J Bone Miner Res2015;30:1539-1550.CrossrefPubMed Google Scholar

    19 Higgins J , Deeks J , . Cochrane Handbook of Systematic Reviews of Interventions. HigginsJPT, GreenS, ed. Chichester, UK: John Wiley & Sons; 2008. Google Scholar

    20 Moher D , Liberati A , Tetzlaff J , Altman DG ; PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. Ann Intern Med2009;151:264-269, W64.CrossrefPubMed Google Scholar

    21 Granke M , Coulmier A , Uppuganti S , et al. Insights into reference point indentation involving human cortical bone: sensitivity to tissue anisotropy and mechanical behavior. J Mech Behav Biomed Mater2014;37:174-185.CrossrefPubMed Google Scholar

    22 Krege JB , Aref MW , McNerny E , et al. Reference point indentation is insufficient for detecting alterations in traditional mechanical properties of bone under common experimental conditions. Bone2016;87:97-101.CrossrefPubMed Google Scholar

    23 Hengsberger S , Enstroem J , Peyrin F , Zysset P . How is the indentation modulus of bone tissue related to its macroscopic elastic response? A validation study. J Biomech2003;36:1503-1509.CrossrefPubMed Google Scholar

    24 Abraham AC , Agarwalla A , Yadavalli A , et al. Multiscale predictors of femoral neck in situ strength in aging women: contributions of BMD, cortical porosity, reference point indentation, and nonenzymatic glycation. J Bone Miner Res2015;30:2207-2214.CrossrefPubMed Google Scholar

    25 Beutel BG , Kennedy OD . Characterization of damage mechanisms associated with reference point indentation in human bone. Bone2015;75:1-7.CrossrefPubMed Google Scholar

    26 Boivin G , Bala Y , Doublier A , et al. The role of mineralization and organic matrix in the microhardness of bone tissue from controls and osteoporotic patients. Bone2008;43:532-538.CrossrefPubMed Google Scholar

    27 Coutts LV , Jenkins T , Li T , et al. Variability in reference point microindentation and recommendations for testing cortical bone: location, thickness and orientation heterogeneity. J Mech Behav Biomed Mater2015;46:292-304.CrossrefPubMed Google Scholar

    28 Dall’Ara E , Karl C , Mazza G , et al. Tissue properties of the human vertebral body sub-structures evaluated by means of microindentation. J Mech Behav Biomed Mater2013;25:23-32.CrossrefPubMed Google Scholar

    29 Diez-Perez A , Güerri R , Nogues X , et al. Microindentation for in vivo measurement of bone tissue mechanical properties in humans. J Bone Miner Res2010;25:1877-1885.CrossrefPubMed Google Scholar

    30 Duarte Sosa D , Vilaplana L , Güerri R , et al. Are the high hip fracture rates among Norwegian women explained by impaired bone material properties? J Bone Miner Res 2015;30:1784-1789.CrossrefPubMed Google Scholar

    31 Farr JN , Drake MT , Amin S , et al. In vivo assessment of bone quality in postmenopausal women with type 2 diabetes. J Bone Miner Res2014;29:787-795.CrossrefPubMed Google Scholar

    32 Granke M , Makowski AJ , Uppuganti S , Does MD , Nyman JS . Identifying Novel Clinical Surrogates to Assess Human Bone Fracture Toughness. J Bone Miner Res2015;30:1290-1300.CrossrefPubMed Google Scholar

    33 Grant JP , Oxland TR , Dvorak MF . Mapping the structural properties of the lumbosacral vertebral endplates. Spine (Phila Pa 1976)2001;26:889-896.CrossrefPubMed Google Scholar

    34 Grant JP , Oxland TR , Dvorak MF , Fisher CG . The effects of bone density and disc degeneration on the structural property distributions in the lower lumbar vertebral endplates. J Orthop Res2002;20:1115-1120.CrossrefPubMed Google Scholar

    35 Güerri-Fernández RC , Nogués X , Quesada Gómez JM , et al. Microindentation for in vivo measurement of bone tissue material properties in atypical femoral fracture patients and controls. J Bone Miner Res2013;28:162-168.CrossrefPubMed Google Scholar

    36 Hansma P , Turner P , Drake B , et al. The bone diagnostic instrument II: indentation distance increase. Rev Sci Instrum2008;79:064303.CrossrefPubMed Google Scholar

    37 Jasiuk I . Multi-scale analysis of bone. In: Proceedings of the Conference on Summer Computer Simulation. Society for Computer Simulation International2015:1-5http://dl.acm.org/citation.cfm?id=2874953 (date last accessed 31 August 2017).CrossrefPubMed Google Scholar

    38 Jenkins T , Coutts LV , D’Angelo S , et al. Site-dependent reference point microindentation complements clinical measures for improved fracture risk assessment at the human femoral neck. J Bone Miner Res2016;31:196-203.CrossrefPubMed Google Scholar

    39 Johnston JD , Kontulainen SA , Masri BA , Wilson DR . Predicting subchondral bone stiffness using a depth-specific CT topographic mapping technique in normal and osteoarthritic proximal tibiae. Clin Biomech (Bristol, Avon)2011;26:1012-1018.CrossrefPubMed Google Scholar

    40 Katsamenis OL , Jenkins T , Thurner PJ . Toughness and damage susceptibility in human cortical bone is proportional to mechanical inhomogeneity at the osteonal-level. Bone2015;76:158-168.CrossrefPubMed Google Scholar

    41 Kerrigan JR , Sanchez-Molina D , Neggers J , et al. Indentation response of human patella with elastic modulus correlation to localized fractal dimension and bone mineral density. J Mech Behav Biomed Mater2014;33:99-108.CrossrefPubMed Google Scholar

    42 Malgo F , Hamdy NA , Papapoulos SE , Appelman-Dijkstra NM . Bone material strength as measured by microindentation in vivo is decreased in patients with fragility fractures independently of bone mineral density. J Clin Endocrinol Metab2015;100:2039-2045.CrossrefPubMed Google Scholar

    43 Mellibovsky L , Prieto-Alhambra D , Mellibovsky F , et al. Bone tissue properties measurement by reference point indentation in glucocorticoid-induced osteoporosis. J Bone Miner Res2015;30:1651-1656.CrossrefPubMed Google Scholar

    44 Milovanovic P , Rakocevic Z , Djonic D , et al. Nano-structural, compositional and micro-architectural signs of cortical bone fragility at the superolateral femoral neck in elderly hip fracture patients vs. healthy aged controls. Exp Gerontol2014;55:19-28.CrossrefPubMed Google Scholar

    45 Milovanovic P , Zimmermann EA , Riedel C , et al. Multi-level characterization of human femoral cortices and their underlying osteocyte network reveal trends in quality of young, aged, osteoporotic and antiresorptive-treated bone. Biomaterials2015;45:46-55.CrossrefPubMed Google Scholar

    46 Mimar R , Limb D , Hall RM . Evaluation of the mechanical and architectural properties of glenoid bone. J Shoulder Elbow Surg2008;17:336-341.CrossrefPubMed Google Scholar

    47 Mirzaali MJ , Schwiedrzik JJ , Thaiwichai S , et al. Mechanical properties of cortical bone and their relationships with age, gender, composition and microindentation properties in the elderly. Bone2016;93:196-211.CrossrefPubMed Google Scholar

    48 Noshchenko A , Plaseied A , Patel VV , et al. Correlation of vertebral strength topography with 3-dimensional computed tomographic structure. Spine(Phila Pa 1976)2013;38:339-349.CrossrefPubMed Google Scholar

    49 Oxland TR , Grant JP , Dvorak MF , Fisher CG . Effects of endplate removal on the structural properties of the lower lumbar vertebral bodies. Spine(Phila Pa 1976)2003;28:771-777.PubMed Google Scholar

    50 Rudäng R , Zoulakis M , Sundh D , et al. Bone material strength is associated with areal BMD but not with prevalent fractures in older women. Osteoporos Int2016;27:1585-1592. Google Scholar

    51 Tan JS , Bailey CS , Dvorak MF , Fisher CG , Oxland TR . Interbody device shape and size are important to strengthen the vertebra-implant interface. Spine(Phila Pa 1976)2005;30:638-644.CrossrefPubMed Google Scholar

    52 Thurner PJ , Erickson B , Turner P , et al. The effect of NaF in vitro on the mechanical and material properties of trabecular and cortical bone. Adv Mater2009;21:451-457. Google Scholar

    53 Malgo F , Hamdy NA , Papapoulos SE , Appelman-Dijkstra NM . Bone material strength as measured by microindentation in vivo is decreased in patients with fragility fractures independently of bone mineral density. J Clin Endocrinol Metab2015;100:2039-2045.CrossrefPubMed Google Scholar

    54 Beutel BG , Kennedy OD . Characterization of damage mechanisms associated with reference point indentation in human bone. Bone2015;75:1-7.CrossrefPubMed Google Scholar

    55 Wolfram U , Wilke H-J , Zysset PK . Rehydration of vertebral trabecular bone: influences on its anisotropy, its stiffness and the indentation work with a view to age, gender and vertebral level. Bone2010;46:348-354.CrossrefPubMed Google Scholar

    56 Burstein AH , Currey JD , Frankel VH , Reilly DT . The ultimate properties of bone tissue: the effects of yielding. J Biomech1972;5:35-44.CrossrefPubMed Google Scholar

    57 Roy ME , Rho JY , Tsui TY , Evans ND , Pharr GM . Mechanical and morphological variation of the human lumbar vertebral cortical and trabecular bone. J Biomed Mater Res1999;44:191-197.CrossrefPubMed Google Scholar

    58 Gupta HS , Stachewicz U , Wagermaier W , et al. Mechanical modulation at the lamellar level in osteonal bone. J Mater Res2006;21:1913-1921. Google Scholar

    59 Oyen ML . Analytical techniques for indentation of viscoelastic materials. Philos Mag2006;86:5625-5641. Google Scholar

    60 Kruzic JJ , Kim DK , Koester KJ , Ritchie RO . Indentation techniques for evaluating the fracture toughness of biomaterials and hard tissues. J Mech Behav Biomed Mater2009;2:384-395.CrossrefPubMed Google Scholar