header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

INTRA-OPERATIVE STABILITY ANALYSES WITH USE OF NAVIGATION SYSTEM AND TENSOR DEVICE BEFORE/ AFTER PCL RESECTION IN TKA USING CRUCIATE-SUBSTITUTING (CS) POLYETHYLENE INSERT

The International Society for Technology in Arthroplasty (ISTA), 29th Annual Congress, October 2016. PART 4.



Abstract

Introduction

Recently, tibial insert design of cruciate-substituting (CS) polyethylene insert is employed and widely used. However, in vivo kinematics of using CS polyethylene insert is still unclear. In this study, it is hypothesized that CS polyethylene insert leads to stability of femoro-tibial joint as well as posterior-stabilized (PS) polyethylene insert, even if PCL is sacrificed after TKA. The purpose of this study is an investigation of in vivo kinematics of femoro-tibial joint with use of CS polyethylene insert before and after PCL resction using computer assisted navigation system and tensor device intra-operatively in TKA.

Materials and Methods

Sixty-one consecutive patients who had knees of osteoarthritis with varus deformity were investigated in this study. All TKAs (Triathlon, Stryker) were performed using computer assisted navigation system. During surgery, using a tensor device, after bony cut of femur and tibia, joint gaps were assessed in 0 and 90 degrees in flexion. Then, CS polyethylene tibial trial insert were inserted after trial implantation of femoral and tibial components, before and after resection of PCL, respectively. The kinematic parameters of the soft-tissue balance, and amount of coronal and sagittal relative movement between femur and tibia were obtained by interpreting kinematics, which display tables throughout the range of motion (ROM) in the navigation system. In each ROM (30, 45, 60, 90, max degrees), the data were analyzed with a ANOVA test, and mean values were compared by the multiple comparison test (Turkey HSD test) (p< 0.05).

Results

Joint gap assessment revealed significant enlargement in both of extension and 90 degrees in flexion after PCL resection compared with before resection.

In kinematic analyses in navigation system, regarding to amount of sagittal movement of tibia, there were significances between before and after PCL resection in 60 and 90 degrees in flexion, 1.2mm difference in 60 degrees, and 2.3mm difference in 90 degrees in flexion. There were no significance between before and after PCL resection in the other degrees in flexion. Regarding to the other analyses, varus/ valgus and rotation, there were no differences between before and after resection of PCL.

In addition, concerning ROM, maximum extension angle is significantly lower, and maximum flexion angle is significantly higher after than before PCL resection.

Discussion

These results demonstrated that CS polyethylene insert might have a stability of femoro-tibial joint nearly after PCL resection as well as before PCL resection. The main design feature of Triathlon CS insert is single radius and rotary arc, in addition, the posterior lip is same as that of Triathlon CR, which can be the factor to avoid paradoxical anterior movement and to permit internal and external rotation between femoral and tibial component. Due to the design features and benefits, there is a high possibility that use of CS insert without PCL can lead same stability as PCL remained, and improvement of ROM. Based on these backgrounds, it is suggested that CS insert may have an additional choice of PCL resection in case of tight gap of flexion in TKA.


*Email: