DIFFERENCES IN FIXATION BETWEEN CEMENTLESS STEM WITH HYDROXYAPATITE COATING ON ROUGH SURFACE AND SAME SHAPE CEMENTLESS STEM WITHOUT HYDROXYAPATITE COATING ON ROUGH SURFACE

The International Society for Technology in Arthroplasty (ISTA), 29th Annual Congress, October 2016. PART 2.



Abstract

Introduction

To evaluate the effect of hydroxyapatite coating, two same shape cementless stems were compared in this randomized control trial study.

Methods

Between May 2003 and February 2010, 88 patients had a primary cementless total hip arthroplasty with two different types of cementless stems. Forty-three patients had Proarc stems (P group) (Kyocera Medical, Osaka, Japan), and Forty-five patients had Proarc HA stems (HA group) (Kyocera Medical, Osaka, Japan) which was coated with thin (20 micrometer) hydroxyapatite on Proarc rough porous coating. Gender distribution, average age at surgery, average weight and average follow-up period were same in the two groups. The average follow-up period was 8.5 years (range, 5 to 13 years). The average age at the time of surgery was 63 years. Porous acetabular shells and highly crosslinked polyethylene liners made by Kyocera Medical corporation were implanted into all hips. Stems were implanted with a modified Hardinge surgical approach without trochanteric osteotomy. Harris Hip Score was used for clinical evaluation. Post-op radiographs of these patients were evaluated. Fisher's exact probability test was used for statistical analysis. P values of less than 0.05 were considered to be significant.

Results

The mean preoperative score and postoperative score of P group were 39 points and 86 points, respectively. The mean preoperative score and postoperative score of HA group were 46 points and 87 points, respectively. All stems were evaluated as bone-ingrown fixation in both groups. The rate of varus inclination was not different between two groups. The rate of severe stress shielding was not different

Discussion

However hydroxyapatite coating is useful for early fixation, the adverse effect, such as delamination for long-term in vivo situation is questionable. There was no significant difference between P-group and HA group in the present study. Longer follow-up is required to evaluate hydroxyapatite coating.