header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Research

CHARACTERISATION OF NOISE IN CERAMIC-ON-CERAMIC TOTAL HIP ARTHROPLASTY

European Orthopaedic Research Society (EORS) 24th Annual Meeting, 14–16 September 2016. Part 2.



Abstract

Ceramic-on-ceramic (CoC) total hip arthroplasty (THA) can produce articular noise during the normal activities, generating discomfort to the patient. THA noise has to be investigated also as a potential predictor and a clinical sign of prosthetic failure.

An observational study has been carried out to characterize the noise in CoC cementless THA, and to analyze the related factors. A total of 46 patients with noisy hip have been enrolled in 38 months, within the follow-up protocol normally applied for the early diagnosis of ceramic liner fracture [1]. Noise recording was based on a high-quality audible recorder (mod. LS 3, Olympus, Japan) and a portable ultrasonic transducer (mod USB AE 1ch, PAC, USA). The sensors for noise recording were applied to the hip of the patient during a sequence of repeatable motorial activities (forward and backward walking, squat, sit in a chair, flexion and extension of the leg). Sessions were also video-recorded to associate the noise emission to the specific movements.

Each noise event was initially identified by the operator and therefore classified by comparison to the spectral characteristics (duration, intensity and frequency) of the main noise types. Number and spectral characteristics of noise events were obtained and correlated to the factors describing the clinical status of the patient, the surgical approach, the prosthetic device implanted. The study investigated also the noise as a sign of implant failure, by comparison with the total number of implants failed in the cohort during the study.

We observed three types of noise with the main spectral characteristics in agreement to the literature: clicking, squeaking and popping. Among the identified types of noise, squeaking showed the longest duration and the highest amplitude. The 63% of hip presented the emission of just one type of noise, while the remaining a mix of types. The movement with the highest presence of noise was walking, followed by squat. Correlation was found between the noise type and the dimension of the ceramic head (p<0.001), with the sizes of 32 mm more affected by squeaking that the smaller one. Squeaking appeared before during the follow-up than the other types of noise. The 35% (16/46) of the noisy hips were revised during the study. Among the revised hips, the 81% (13/16) were affected by impingement and/or severe damage of the prosthetic components. The antiversion of the cup (p=0.008), the presence of debris in the synovial fluid (p=0.021) and the average frequency of squeaking (p=0.006) were significant predictors for the revision, but it has to be mentioned that the squeaking data was obtained on a small subset of revised patients. Ultrasonic analysis did not show significant correlations.

The study presented and validated an experimental procedure to analyze noisy hips in clinical trials. Noise is confirmed to be a significant parameter in the follow-up evaluation of ceramic THA.