THE NEW FEMORAL NECK SYSTEM FOR FEMORAL NECK FRACTURE FIXATION: BIOMECHANICAL PERFORMANCE VERSUS DIVERSE STANDARD IMPLANTS FOR OSTEOSYNTHESIS

European Orthopaedic Research Society (EORS) 24th Annual Meeting, 14–16 September 2016. Part 2.



Abstract

Three Cannulated Screws (3CS), Dynamic Hip Screw (DHS) with antirotation screw (DHS–Screw) or with a Blade (DHS–Blade) are the gold standards for fixation of unstable femoral neck fractures. Compared to 3CS, both DHS systems require larger skin incision with more extensive soft tissue dissection while providing the benefit of superior stability. The newly designed Femoral Neck System (FNS) for dynamic fixation combines the advantages of angular stability with a less invasive surgical technique. The aim of this study is to evaluate the biomechanical performance of FNS in comparison to established methods for fixation of the femoral neck in a human cadaveric model.

Twenty pairs of fresh–frozen human cadaveric femora were instrumented with either DHS–Screw, DHS–Blade, 3CS or FNS. A reduced unstable femoral neck fracture 70° Pauwels III, AO/OTA31–B2.3 was simulated with 30° distal and 15° posterior wedges. Cyclic axial loading was applied in 16° adduction, starting at 500N and with progressive peak force increase of 0.1N/cycle until construct failure. Relative interfragmentary movements were evaluated with motion tracking.

Highest axial stiffness was observed for FNS (748.9 ± 66.8 N/mm), followed by DHS–Screw (688.8 ± 44.2 N/mm), DHS–Blade (629.1 ± 31.4 N/mm) and 3CS (584.1 ± 47.2 N/mm) with no statistical significances between the implant constructs. Cycles until 15 mm leg shortening were comparable for DHS–Screw (20542 ± 2488), DHS–Blade (19161 ± 1264) and FNS (17372 ± 947), and significantly higher than 3CS (7293 ± 850), p<0.001. Similarly, cycles until 15 mm femoral neck shortening were comparable between DHS–Screw (20846 ± 2446), DHS–Blade (18974 ± 1344) and FNS (18171 ± 818), and significantly higher than 3CS (8039 ± 838), p<0.001.

From a biomechanical point of view, the Femoral Neck System is a valid alternative to treat unstable femoral neck fractures, representing the advantages of a minimal invasive angle–stable implant for dynamic fixation with comparable stability to the two DHS systems with blade or screw, and superior to Three Cannulated Screws.