header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

FEMORAL COMPONENT ROTATION IN TOTAL KNEE ARTHROPLASTY: ANATOMICAL LANDMARK METHODS VERSUS A FORCE SENSOR DEVICE

The International Society for Technology in Arthroplasty (ISTA), 28th Annual Congress. PART 2.



Abstract

Introduction

11%–19% of patients are unsatisfied with outcomes from Total Knee Arthroplasty (TKA). This may be due to problems of alignment or soft-tissue balancing. In TKA, often a neutral mechanical axis is established followed by soft tissue releases to balance and match the flexion/extension gaps with the distal femoral and proximal tibial resections at right angles to the mechanical axis. Potential issues with establishment of soft tissue balance are due to associated structures such as bone tissue of the knee, the static (or passive) stabilizers of the joint (medial and lateral collateral ligaments, capsule, and anterior and posterior cruciate ligaments), and the dynamic (or active) stabilizers around the knee. An optimized balance among these systems is crucial to the successful outcome of a TKA. Additionally, the importance of correct femoral rotation has been well documented due to its effect on patella alignment and flexion instability, range of motion, and polyethylene wear.

There are several methods used in TKA procedures to establish femoral component rotation. The more prominent ones are a conventional method of referencing to the posterior condylar axis with a standard external rotation of 3° (PCR), anterior-posterior line or “Whiteside's line” (AP axis), transepicondylar axis (TEA) (Figure 1), and the gap balancing technique, however, it is not yet clear, which method is superior for femoral rotational component alignment.

In the current study, we sought to investigate an alternative method based on soft-tissue, dynamic knee balancing (DKB) while using an alternative analysis approach. DKB dictates femoral component rotation on the basis of ligament balance and force measures. DKB has become more prominent in TKA surgeries. While retaining ligament balance in TKA, it is possible that this technique also leads to higher precision of rotational alignment to the anatomical axis. The primary objective of this study was to compare efficiency of DKB versus other methods for rotational implant alignment based on post-surgery computed tomography (CT).

Methods

31 patients underwent computer-navigated total knee arthroplasty for osteoarthritis with femoral rotation established via a flexion gap balance device (Synvasive eLibra). Alternative, hypothetical alignments were assessed based on anatomical landmarks during the surgery. Postoperative computed tomography (CT) scans were analyzed to investigate post-surgery rotational alignment. Repeated measures ANOVA and Cochran's Q test were utilized to test differences between the DKB method and the other techniques.

Results

Significant differences were observed between the DKB method and TEA method (p=0.02), between DKB and AP method (p=0.04), and DKB and PCR method (p=0.02): The DKB method showed the lowest rotational deviation from CT-determined true anatomical TEA (aTEA)(Figure 2). The DKB method established femoral rotation within ±3 more often than the other techniques (Figure 3), further analysis revealed a significant proportional difference between DKB and PCR method (p=0.01), between DKB and TEA (p=0.02) and DKB and AP (p=0.04).

Conclusions

DKB showed promising results in our study regarding femoral rotation accuracy in comparison to other methods. DKB may be a valuable tool due to its ability to establish soft-tissue balance in addition to high accuracy of femoral rotation.


*Email: