header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

CONSTRAINED LINERS: YESTERDAY'S SOLUTION?

Current Concepts in Joint Replacement (CCJR) – Winter 2015 meeting (9–12 December).



Abstract

In the revision situation in general and for recurrent dislocation specifically, it is important to have all options available including tripolar constrained liners in order to optimise the potential for hip stability as well as function. Even with the newer options available, dislocation rates of higher than 5% have been reported in the first two years following revision surgery at institutions where high volumes of revision surgery are performed (Wera et al). Because of the deficient abductors, other soft tissue laxity and the requirement for large diameter cups, revision cases will always have more potential for dislocation. In these situations, in the lower demand patient, tripolar constrained liners provided excellent success in terms of preventing dislocation and maintaining implant construct fixation to bone at intermediate term follow-up. Hence in these situations, tripolar with constraint remains the option we utilise in many cases. We are also confident in using this device in cases with instability or laxity where there is a secure well positioned acetabular shell. We cement a tripolar constrained liner in these situations using the technique described below.

Present indication for tripolar constrained liners: low demand patient, abductor muscle deficiency or soft tissue laxity, large outer diameter cups, multiple operations for instability, instability with well-fixed shells that are adequately positioned

Technique of cementing liner into shell: score acetabular shell if no holes, score liner in spider web configuration, all one or two millimeters of cement mantle

Results: Constrained Tripolar Liner - For Dislocation: 56 Hips; 10 year average f/u; 7% failure of device, 5% femoral loosening, 4% acetabular loosening

Constrained Tripolar Liner - For Difficult Revisions: 101 hips; 10 year average f/u; 6% failure of device, 4% femoral loosening, 4% acetabular loosening

Cementing Liner into Shell: 31 hips; 3.6 year average f/u (2–10 years); 2 of 31 failures

We, like others, are trying to define cases where dual mobility will be as successful or more successful than tripolar constrained liners.