header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

CONTEMPORARY POLYMERS: ALL OUR BURDENS NOW OVER?

Current Concepts in Joint Replacement (CCJR) – Winter 2015 meeting (9–12 December).



Abstract

As a generic technology, intentionally crosslinked polyethylene has improved the durability of total hip replacement. Regardless of the manufacturing method, the wear rates have been reduced on the order of 90% compared to historical materials, with a substantial reduction in the occurrence of osteolysis. Most of the data is with 28 and 32 mm bearings. Larger diameter bearings have been shown to reduce the occurrence of dislocation. However, there is clinical evidence that volumetric wear is increased with larger diameter crosslinked polyethylene bearings, and this may increase the occurrence of osteolysis. Further, modular liner fracture is more likely with larger diameter bearings (thinner liners), which is generally associated with increased cup abduction angle and/or increased anteversion. Contemporary polymers are better than their predecessors, but there is always opportunity for improvement.