A NOVEL EXTRAMEDULLARY GUIDE FOR FEMORAL BONE CUTTING BY PRE-OPERATIVE PLANNING USING LONG LEG VIEW IN TKA

The International Society for Technology in Arthroplasty (ISTA), 27th Annual Congress. PART 2.



Abstract

Despite many new methods with preoperative or intra-operative assistance to improve the accuracy of leg alignment, traditional intramedullary (IM) method of bone cutting is still the most commonly used. Traditional TKR using IM guide has more outliers comparing to new computer aided methods, especially in bowing femur which is more prevalent in Asian population. And IM guide could not be used when there is a medullary bony pathology. Avoiding entrance of medullary cavity has been proposed as one of criteria of minimally invasive TKA.

We have designed an extramedullary (EM) guide for the distal femoral bone cutting with millimeter to millimeter increment which is compatible with all posterior referencing instrumentation systems. With mechanical line as the guide line on long leg X-ray film taking with the knee and foot facing anteriorly, the amount of distal femoral bone cutting was measured and recorded on computer screen pre-operatively. During surgery, distal femoral cutting was performed using the EM cutting jig for coronal alignment adjustment tool and anterior femoral cortex and a guide post as sagittal alignment guide.

We retrogratively compared the post-operation long leg X-ray film of two hundreds patients using IM or EM guides, the mechanical alignment of femoral components were measured in coronal and sagittal planes. The results showed no significant difference in distribution and the ratio of outliers. However, if the bowing of femur is more than 8 degree, the outlier is more in the IM group.

We have applied this technique in 8 patients having medullary pathology including plates or nails in femur. All patients got good post-operative limb alignment without the needs of computer assistance device during surgery.

In conclusion, the technique is easy and the instrument is simple. The operative time was not prolonged; the medullary cavity was not entered and compatible with the principle of MIS technique. In case of medullary cavity is obstructed, it is cost-effective by using our EM guide comparing to other methods such as CAOS or PSI.