PREDICTION OF WEAR IN MOBILE AND FIXED BEARING, CROSS-LINKED AND NON-CROSS-LINKED, UNICOMPARTMENTAL AND TRICOMPARTMENTAL KNEE ARTHROPLASTY

The International Society for Technology in Arthroplasty (ISTA), 27th Annual Congress. PART 1.



Abstract

Background

Wear and fatigue damage to polyethylene components remain major factors leading to complications after total knee and unicompartmental arthroplasty. A number of wear simulations have been reported using mechanical test equipment as well as computer models. Computational models of knee wear have generally not replicated experimental wear under diverse conditions. This is partly because of the complexity of quantifying the effect of cross-shear at the articular interface and partly because the results of pin-on-disk experiments cannot be extrapolated to total knee arthroplasty wear. Our premise is that diverse experimental knee wear simulation studies are needed to generate validated computational models. We combined five experimental wear simulation studies to develop and validate a finite-element model that accurately predicted polyethylene wear in high and low crosslinked polyethylene, mobile and fixed bearing, and unicompartmental (UKA) and tricompartmental knee arthroplasty (TKA).

Methods

Low crosslinked polyethylene (PE). A finite element analysis (FEA) of two different experimental wear simulations involving TKA components of low crosslinked polyethylene inserts, with two different loading patterns and knee kinematics conducted in an AMTI knee wear simulator: a low intensity and a high intensity. Wear coefficients incorporating contact pressure, sliding distance, and cross-shear were generated by inverse FEA using the experimentally measured volume of wear loss as the target outcome measure. The FE models and wear coefficients were validated by predicting wear in a mobile bearing UKA design.

Highly crosslinked polyethylene (XLPE). Two FEA models were constructed involving TKA and UKA XLPE inserts with different loading patterns and knee kinematics conducted in an AMTI knee wear simulator. Wear coefficients were generated by inverse FEA.

Results

Predicted wear rates were within 5% of experimental wear rates during validation tests.

Unicompartmental mobile bearing back-side wear accounted for 46% of the total wear in the mobile bearing. Wear during the swing phase was 38% to 44% of total wear.

Discussion & Conclusions

Crosslinking polyethylene primarily decreased (by nearly 10-fold) the wear generated by cross-shear. This result can be explained by the reduced propensity of crosslinked polyethylene molecules to orient in the dominant direction of sliding. A highly crosslinked fixed-bearing polyethylene insert can provide high wear performance without the increased risk for mobile bearing dislocation. Finite element analysis can be a robust and efficient method for predicting experimental wear. The value of this model is in rapidly conducting screening studies for design development, assessing the effect of varying patient activity, and assessing newer biomaterials. This FEA model was experimentally validated but requires clinical validation.