General Orthopaedics

PILOT IN VIVO EVALUATION OF A RESORBABLE ANTIBIOTIC-ELUTING BONE VOID FILLER TO PREVENT PERIPROSTHETIC JOINT INFECTION

European Bone And Joint Infection Society (EBJIS) 34th Annual Meeting: PART 1



Abstract

A resorbable, antibiotic-eluting bone void filler (AEBVF) was developed to address device-related infections. The AEBVF provides two functions: osteoconductive matrix for bone restoration, and local antibiotic delivery to treat device-related infections. In vitro evaluations of this AEBVF demonstrated antimicrobial activity to 7 weeks against Staphylococcus aureus (S. aureus).1 Subsequent rabbit studies demonstrated bactericidal capacity2 of the AEBVF against 105 CFU S. aureus and osteoconductivity.1 We hypothesized that the AEBVF would restore bone volume while eliminating 105 CFU S. aureus in a pilot sheep femoral condyle defect model.

Four groups (n=2/group) were utilized to assess osteoconductivity (Group A-commercial ProOsteon & B-AEBVF) and antimicrobial activity (Group C-ProOsteon with 105 CFU S. aureus & D-AEBVF with 105 CFU S. aureus). AEBVF devices comprised degradable polymers (PCL, PEG, PLGA), ProOsteon (Biomet, USA), CaCl2, and tobramycin.3 Devices (1.5cc ProOsteon or 6 AEBVF croutons) were implanted into rectangular defects in the medial face of each sheep femoral condyle. Defects were evaluated using backscatter electron microscopy, mineral apposition rate (MAR) analysis, and light microscopy with Sanderson's Rapid Bone Stain (SRBS).

All animals in Groups A, B, and D survived to the 12-week endpoint. In contrast, Group C animals were euthanized 11 days post-op. MAR and SRBS demonstrated comparable bone remodeling and defect restoration after 12 weeks in Groups A, B, and D. Notably, implant volumes of Groups A and D were greatly diminished (0.16±0.1%; 0.35%) after 12 weeks, compared to Group A (13.23±3.2%) and Time “0” (16.8%).

These data show the AEBVF device's ability to: eliminate 105 CFU S. aureus, promote bone remodeling comparable to known bone void filler, and degrade at rates that do not interfere with bone remodeling.