header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Research

PARTICLES FROM VITAMIN-E-DIFFUSED HIGHLY CROSS-LINKED UHMWPE INDUCE LESS OSTEOLYSIS COMPARED WITH VIRGIN UHMWPE IN A MURINE CALVARIAL BONE MODEL

8th Combined Meeting Of Orthopaedic Research Societies (CORS)



Abstract

Summary Statement

Vitamin E-UHMWPE particles have a reduced osteolysis potential in vivo when compared to virgin, highly cross-linked UHMWPE in a murine calvarial bone model.

Introduction

Ultra high-molecular weight polyethylene (UHMWPE) particle-induced osteolysis is one of the major causes of arthroplasty revisions. The lack of particle clearance from the joint inevitably leads to the upregulation of the inflammatory cascade, resulting in bone resorption and implant loosening. Recent in vitro findings (Bladed CL et al. ORS 2011 and J Biomed Mater Res B Appl Biomater, 2012) have suggested that UHMWPE wear particles containing vitamin-E (VE) may have reduced functional biologic activity and decreased potential to cause osteolysis. This is of significant importance since VE-stabilised cross-linked UHMWPEs were recently introduced for clinical use, and there is no in vivo data determining the effects of wear debris from this new generation of implants. In this study we hypothesised that particles from VE-stabilised, radiation cross-linked UHMWPE (VE-UHMWPE) would cause reduced levels of osteolysis in a murine calvarial bone model when compared to virgin gamma irradiated cross-linked UHMWPE.

Methods

Study groups were the following: 1) Radiation cross-linked VE-UHMWPE, approximately 0.8% by weight, diffused after 100 kGy; 2). Radiation cross-linked virgin UHMWPE (virgin UHMWPE); 3). Shams. Particle generation and implantation: UHMWPE was sent to Bioengineering Solutions (Oak Park, IL) for particle generation. After IACUC approval, C57BL/6 mice (n=12 for each group) received equal amount of particulate debris (3mg) overlying the calvarium and were euthanised after 10 days. Micro-CT scans: High resolution micro-CT scans were performed using an X-Tek HMX ST 225 with a set voltage of 70 kV and current of 70 µA. Topographical Grading Scale: Each calvarial bone (interparietal, right and left parietal, right and left frontal) was blindly scored using the following scale: 0=No osteolysis, defined as intact bone; 1=Minimal osteolysis, affecting 1/3 or less of the bone area; 2=Moderate osteolysis, affecting at least 2/3 of the bone area; 3=Severe osteolysis, defined as completely osteolytic bone. Histological Analysis: H&E and TRAP staining was performed on tissue to confirm the micro-CT findings and to quantify osteoclasts. Statistical Analysis: Inter-rater analysis was performed using Cohen's kappa analysis. An inter-rater coefficient >0.65 was considered as high inter-rater agreement. Comparison between groups was made using one-way ANOVA with post hoc Bonferroni correction for multiple comparisons. Correlations are reported as Spearman's rho. A p-value<0.05 was considered statistically significant.

Results

More than 83% of the VE-UHMWPE and more than 85% of the virgin UHMWPE particles measured less than 1 µm in mean particle size. The mean particle size for VE-UHMWPE was 1.12 µm (range 0.28 to 79.08 µm), while virgin UHMWPE particles measured 1.22 µm (range 0.28 to 82.04 µm). There was a statistically significant greater level of osteolysis visualized on the topographical grading scale in calvaria implanted with virgin UHMWPE wear particles. The micro-CT findings were confirmed histologically. A greater amount of inflammatory tissue overlaying the calvaria was observed in the virgin UHMWPE group when compared to both shams and VE-UHMWPE groups. Post hoc analysis revealed significant difference between VE-UHMWPE and virgin UHMWPE for the topographical osteolysis grading score (p = 0.002) but no difference in osteoclast count (p = 0.293).

Discussion/Conclusion

This is the first in vivo study reporting the effects of clinically-relevant UHMWPE particles generated from a VE-UHMWPE implant that is in current clinical use. These results suggest that VE-UHMWPE particles have reduced osteolysis potential in vivo when compared to virgin, highly cross-linked UHMWPE in a murine calvarial bone model. Arthroplasty procedures using VE-UHMWPE might be less susceptible to peri-prosthetic loosening caused by wear debris.