PERCUTANEOUS OSTEOTOMIES OF THE FEMUR AND TIBIA USING A COOLED SIDE CUTTING BURR

British Society for Children's Orthopaedic Surgery (BSCOS)



Abstract

Purpose of the Study

A cooled, side cutting burr designed for use in adult foot surgery has been used as a primary bone cutting device in children to facilitate a truly percutaneous method of performing osteotomies. Stabilisation of the femur was using a percutaneous locked nail and for the tibia percutaneous K-wires. The author describes the advantages and disadvantages of this method with results from the first cohort of patients treated.

Method

Patients under going osteotomy of the femur, tibia and fibula using a 2 mm × 20 mm side cutting burr were followed prospectively and assessed for scar size, bone healing time and complications.

Results

Thirty six osteotomies were performed in the femur and tibia in 25 patients. A fibula osteotomy was always performed with a tibial osteotomy. Scar size for the femoral osteotomy was <15 mm and for the nail insertion <25 mm. For the tibia & fibula the scar size was <10 mm. Healing time was by 6 weeks in the tibia and in the femur was within 16 weeks in non-lengthening cases in all but 1 case of non-union (associated with Vitamin D insufficiency). Three burr bits broke during the learning curve including 2 in tibial osteotomies and 1 in the femur. Cortical thickness and slow burr speeds were associated with burr breakage. All wounds healed without infection. The optimum speed for the burr at 50 Nm of torque was established as 200 rpm in children under the age of 13 yrs. The initial recommended speed of 300 rpm increases healing time when performing osteotomies in children.

Conclusion

Truly percutaneous surgery can be performed to osteotomise the femur and tibia using a cooled, side cutting burr with a locked femoral nail and crossed tibial wires with excellent cosmetic results and minimal complications.

Level of Evidence IV