Bone Quality Assessment With Intraoperative Trabecular Torque Measurement During Total Hip Arthroplasty

International Society for Technology in Arthroplasty (ISTA)



Abstract

In cases of poor bone quality intraoperative torque measurement might be an alternative to preoperative dual energy x-ray absorptiometry (DXA) to assess bone quality in Total Hip Arthroplasty (THA). 14 paired fresh frozen human femurs were included for trabecular peak torque measurement. We evaluated an existing intraoperative torque measurement method to assess bone quality and bone strength. We modified the approach to use this method in total hip arthroplasty (THA), which has not been published before. Since there are several approaches used in THA to exposure the hip joint, we decided to prefer the measurement in the femoral head which allows every surgeon to perform this measurement. Here a 6.5 × 23 mm blade was inserted into the proximal femur without harming the lateral cortical bone (figure 1). Further tests of the proximal femur evaluated the results of this new method: DXA, micro-computed tomography (μCT) and biomechanical load tests. Basic statistical analyses and multiple regressions were done. In the femoral head mean trabecular peak torque was 4.38 ± 1.86 Nm. These values showed a strong correlation with the values of the DXA, the μCT and the biomechanical load test. In comparison to the bone mineral density captured by DXA, the results of the intraoperative torque measurement showed a superior correlation with high sensitive bone quality evaluating methods (mechanical load tests and micro-computed tomography). Hence, the use of this intraoperative torque measurement seems to be more accurate in evaluating bone strength and bone quality than DXA during THA. The torque measurement provides sensitive information about the bone strength, which may affect the choice of implant in cases of poor bone stock and osteoporosis. In clinical use the surgeon may alter the prosthesis if the device indicates poor bone quality. Furthermore, we assume that the disadvantages associated with DXA scans like radiation exposure or errors caused by potential extraosteal sclerosis and interindividual soft-tissue artifacts could be excluded.


*Email: