Intra-Operative Photographic Assessment of Acetabular Cup Inclination During Total Hip Arthroplasty Compared with Post-Operative Radiographic Evaluation

International Society for Technology in Arthroplasty (ISTA)



Abstract

The purpose of this study was to examine the utility of the acetabular component introducer as a tool to intra-operatively predict implant inclination in total hip arthroplasty. This study investigated (1) the correlation between intra-operative photographic assessment of cup inclination using the acetabular introducer and that measured on post-operative radiograph; and (2) the accuracy of intra-operative prediction of abduction angle.

For this study, we prospectively recruited 56 patients scheduled to receive primary hip arthroplasty from one of two senior surgeons. During the procedure, the lead surgeon provided a prediction of the abduction angle based on the alignment of the impactor attached to the cup in its final seated position. A standardized anteroposterior (AP) photograph was then taken of the acetabular impactor in situ. Abduction angles were measured by two observers on the photographs and post-operative AP pelvis radiographs. Linear regression was used to determine the correlation between the angle of the guide measured on the photographs and the actual position of the implant measured on the radiograph. Descriptive statistics were further used to analyze the accuracy of the intra-operative prediction as compared with the abduction angle measured on the photographs.

Measurements of cup position made from post-operative radiographs were significantly correlated with the measurements as assessed by intra-operative photographs (r = 0.34, p = 0.00). Our findings demonstrate that radiological abduction angles tend to be greater than that assessed by intra-operative photographs by a mean of 5.6 degrees (SD = 6.6 degrees; 95% CI = 7.3 to 3.9 degrees). Conversely, surgeon prediction of cup inclination based on the acetabular introducer differed from the radiographic measurements by a mean of 6.8 degrees (SD = 8.7 degrees). There was good agreement between the two observers in both photographic and radiographic measurement (k = 0.95, k = 0.96, respectively).

In conclusion, we found that the intra-operative photographic assessment of acetabular cup inclination by acetabular impactor alignment tends to underestimate the abduction angle by a mean of approximately 5 degrees. In addition, intra-operative surgeon estimation of acetabular inclination did not appear accurate in this study demonstrating that cup position should rely on additional visual cues beyond that captured in the anteroposterior view of the cup introducer.