header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

The Radiographic Findings After Total Hip Arthroplasty With Revelation Stem

International Society for Technology in Arthroplasty (ISTA)



Abstract

This study shows the radiographic results of total hip arthroplasty (THA) using Revelation stem ® (Djo surgical USA) for hip joint disease.

We removed cases which we couldn't follow up, the remaining of the patients who had undergone a primary THA July 2007 to December 2009 in our institution using Revelation hip stem® and 58 cases-65 hips (14 men and 44 women) were possible for progress observation on this study, and radiographic evaluation was done at the time of the last observation. The preoperative diagnosis of the hip included osteoarthritis (OA) in 40 patients, idiopatic, necrosis of the hip in 13 patients, Rheumatoid arthritis (RA) in 4 patients, and femoral head fracture in 1 patient.

Three patients had undergone femoral head replacement (FHA) and 55 patients had undergone THA.

At the time of the last evaluation, spot welds were detected in 63 hips (97%) in zone1 and 7. Demarcation lines, which indicates movement of the stem, were detected in zone3,4,5, but not in the proximal of the stem.

Stress shielding were observed in 32 joints, according to Engh's classification, first degree were 27 joints and second degree were 5 joints. Cortical hypertrophy were detected in 13 joints.

The revelation stem features a lateral overhang structure (lateral flare) and the stem has an expanded proximal geometry allowing the device to rest on both medial and lateral cortices at the metaphyseal level. Above all the stem has anterior overhand structure (trapezoidal structure). The medullary cavity occupation rate is made to increase by these structures. Above all, the stability of the stem is increased by a proximal portion by three point support.

The distal stem is tapered and polished, which prevents the stress loading to a distal portion of the device.

Stress shielding more than the third degree is not detected, and spots welds were detected in the proximal of the stem. These facts indicate that the concept of this stem, transmit loading to the femur bone, at the proximal portion of the stem, would be achieved.


*Email: