Measurement of Acetabular Cup Anteversion on Lateral Radiographs Using Ischial Axis as an Anatomical Reference

International Society for Technology in Arthroplasty (ISTA)



Abstract

Introduction:

The lateral radiographs are useful in evaluation of the acetabular cup anteversion. However, this method was affected by variations in pelvic position and radiographic technique. In this study, we employed the ischial axis (IA) as an anatomical landmark on the lateral radiographs, and we investigated a relationship between IA and the anterior pelvic plane (APP) using three-dimensional computed tomography (3D-CT). Using these findings, we report a new method for accurate measurement of the acetabular cup anteversion on plain lateral radiographs using IA as an anatomical reference.

Materials and Methods:

At first, preoperative3D-CT images were obtained in 109 patients who underwent total hip arthroplasty. The diagnosis was osteoarthritis in all patients. The angle between the IA (defined by a line connecting the anterior edge of the greater sciatic notch and the lesser sciatic notch) and APP (defined by the bilateral anterosuperior iliac spine and the symphysis) was measured on 3D-CT (Fig. 1). Secondly, postoperative lateral radiographs were obtained at 2 weeks, 4 weeks, 12 weeks, 24 weeks, and 52 weeks after surgery in 15 patients. The angle between a line tangential to the opening of the cup and a line perpendicular to APP was measured (Fig. 2). Three methods of acetebular cup position assessment were compared: 1) the present method, 2) Woo and Morrey method, and 3) software (2D template, Kyocera) method.

Results:

The mean angle between IA and APP was 18.0 ± 3.5°. The mean acetabular cup anteversion measured using present method was 21.3°, Woo and Morrey method was 26.6°, and software method was 21.2°. The mean SDs of present method was 0.64°, Woo and Morrey method was 1.17°, and software method was 0.46°

Conclusions:

APP, considered as vertical in weight bearing, has a relatively consistent relationship between IA. The findings of this study provide a more consistent measurement of acetabular cup by reducing variation due to pelvic position.