The Accuracy of CT-Based Fluoroscopy-Matching Navigation System in Revision THA

International Society for Technology in Arthroplasty (ISTA)



Abstract

Introduction:

Since2007, we have used CT-based fluoroscopy-matching navigation system (Vector Vision Hip Ver.3.5.2, BrainLAB, Germany) in revision total hip arthroplasty. This system completes the registration procedure semi-automatically by matching the contours of fluoroscopic images and touching 3 adequate points to the contours of 3D bone model created in the computer. Registration procedure using fluoroscopic figures has finished before making surgical incision. It needs no elongation time during the operation. The objective of this study was to evaluate the accuracy of CT-based fluoroscopy-matching navigation system in revision THA.

Material and method:

We analysed the acetabular cup in consecutive 33 hips with both intra-operative and post-operative alignment data (based on navigation system and CT evaluation) We further compared these measurements with results from primary THA. Data for primary THA were therefore obtained from 40 consecutive patients who underwent primary THA between August 2007 and May 2013 using the same navigation system by postero-lateral approach. We aimed the cup angle for Revision THA as following, the inclination: 40 degrees, the anteversion: 20 degrees Anteversion on the navigation system must be adjusted by the pelvic tilt.

Results:

There was one dislocation in 33 Revision THAs. There was no other obvious complication (nerve palsy, VTE and Infection). The all cup alignments were within 7 degrees from the preoperative orientation. In the Revision THA group the differences between the intra- and post-operative measurement of cup inclination were 2.3 ± 1.9 degrees. The differences of cup anteversion were 2.7 ± 2.5 degrees. In the primary THA group, the differences between the intra- and post-operative measurement of cup inclination were 1.9 ± 2.1 degrees. The differences of cup anteversion were 2.1 ± 2.5 degrees. There was no significant difference with two groups.

Discussion:

CT-based navigation THA is very useful for severe deformity of hip osteoarthritis. We had used CT-based navigation system (landmark matching) since 2003. It needs some technical skills to improve the accuracy of landmark matching. The registration with CT-based fluoroscopy-matching navigation system is much easier and more simple than with landmark matching navigation system. CT images of revision patients included metal artifacts caused by implants. However this system is not so affected by metal artifacts. And we found this system provided high accuracy even in revision THA.


*Email: