header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

Ct-Based Patient-Specific Instrumentation Is Effective in Patients With Pre-Existing Hardware About the Knee

International Society for Technology in Arthroplasty (ISTA)



Abstract

Hardware in or about the knee joint presents a number of challenges to the surgeon in performance of Total Knee Arthroplasty (TKA). Conventional instrumentation usually requires a modification of technique or removal of the metallic implants. Computer-Assisted TKA (CAOS) is another option, but adds complexity and time to the procedure. MRI-based Patient-Specific Instrumentation (PSI) cannot be used as metal causes unwanted artifact and renders the images for planning, useless. However, CT scans are not affected by metal and thus CT-based PSI can be used in TKA patients with pre-existing hardware.

The present IRB approved study evaluates 12 consecutive knees (10 patients) with pre-existing hardware using CT-based PSI (MyKnee®, Medacta International, SA, Castel San Pietro, Switzerland).

In this technique, CT scan of the lower extremity is obtained, and from these images, the knee is reconstructed 3-dimensionally. Surgical and implant-size planning are performed according to surgeon preference, with the goal to create a neutral mechanical axis. Once planned and approved, the blocks are made [Fig 1].

During surgery, the PSI cutting block is registered on the femur first and secured with smooth pins. The distal femoral resection is performed directly through the block. An appropriate sized 4-in-1 block is placed and the remaining femoral resections are performed. The tibial resection block is registered and resection performed. Final bone preparation, patella resurfacing, and trialing is performed as is standard to all surgical techniques.

Of the 12 TKAs, there were 5 left and 7 right knees performed in 6 females and 6 males. The average BMI was 33.19 and average age was 53 (range 44–63). All diagnoses were either osteoarthritis or post-traumatic osteoarthritis. Follow-up averaged 59 weeks (range 18.6–113.7).

Nine patients had pre-operative varus deformities with HKA deformities average of 171.9° (range 154°–178.5°). One patient had pre-operative valgus deformity of 184.5°. Two patients were neutral (180°). Post-operative alignment for all patients (n = 11) was 179° (range 177°–180°). All patients were within 3° neutral, post operatively. Four patients measured 180°, 4 measured at 179°, 2 measured at 178°, and only one at 177°.

Hardware consisted of 5 patients with femur or tibia staples, 3 with plate(s) and screws [Fig. 2], 3 patients with ACL interference screws, and one titanium rod. No hardware was removed unless necessary for implantation. Only 3 patients required some hardware removal.

The pre-operative Range of Motion (ROM) averaged 2.9° to 98.3° (Extension range 0–15° and flexion range 30–115°). Post-operative ROM was 2.9° to 101.3°. (Extension range 0–5° and flexion range 65–125°). Knee Society Score (KSS) improved from 42.3 to 82.3, and KSS Function Score improved from 52.1 to 77.5. No intraoperative complications were recorded. Average tourniquet time was 42.1 minutes (range 28–102).

Regardless of the deformity, the patient's post-operative mechanical axes HKA averaged 179° (range 177–180). Clinical scores were typical for TKA patients with improvement in both KSS and ROM.

In conclusion, early results using PSI in patients with pre-existing hardware in or about the joint, is safe, efficient, and accurate in performance of TKA.


*Email: