header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

Bone Mineral Density Has a Direct Effect on the Fixation Strength of Cementless Femoral Knee Component

International Society for Technology in Arthroplasty (ISTA)



Abstract

To achieve desirable outcomes in cementless total knee replacement (TKR), sufficient primary stability is essential. The primary stability inhibits excessive motions at the bone-implant interface, hence providing the necessary condition for osseointegration [1]. Primary stability for cementless TKR is provided by press-fit forces between the bone and implant. The press-fit forces depend on several factors including interference fit, friction between bone and implant surface, and the bone material properties. It is expected that bone mineral density (BMD) will affect the stability of cementless TKR [2]. However, the effect of BMD on the primary stability of cementless femoral knee component has not been investigated in vitro.

Phantom calibrated CT-scans of 9 distal femora were obtained after the surgical cuts were made by an experienced surgeon. Since the press-fit forces of the femoral component mainly occur in the Anteroposterior (AP) direction, the BMD was measured in the anterior and posterior faces for a depth of 5 mm; this depth was based on stress distributions from a Finite Element Analysis of the same implant design. In addition, four strain gauges were connected to different locations on the implant's outer surface and implant strain measured throughout as an indication of underlying bone strain. A cementless Sigma CR femoral component (DePuy Synthes Joint Reconstruction, Leeds, UK) was then implanted using an MTS machine. In order to simulate a ‘normal’ bone condition, the implanted bone was preconditioned for one hour at a cyclic load of 250–1500 N, and a rate of 1 Hz. Finally, the implants were pushed-off from the bone in a high-flex position. Forces and displacements were recorded both during insertion and push-off tests.

Strong correlations were found for insertion and push-off forces with BMD, R2 = 0.88 and R2 = 0.88, respectively (p < 0.001), so although implantation may be harder in patients with higher BMD, initial stability is also improved. A correlation was also found between final strain and push-off forces (R2 = 0.89, p < 0.01) and BMD also showed a strong reverse correlation with total bone relaxation (R2 = 0.76, p = 0.023). These results indicate that higher BMD induces higher bone strain, which can lead to improved fixation strength.

There is no consensus on the best fixation method for the TKR but some surgeons prefer a cementless design for young and active patients. The results of our study showed that the primary stability of a cementless femoral knee component is directly correlated with the bone mineral density. Therefore, patient selection based on bone quality may increase the likelihood of good osseointegration and adequate long-term fixation for cementless femoral knee components.


*Email: