General Orthopaedics

Measurement of Serum Cobalt and Chromium Ion Levels in Metal on Metal Total Hip Arthroplasty With Different Femoral Head Size

International Society for Technology in Arthroplasty (ISTA)



Abstract

Metal on metal total hip arthroplasty provides the potential improvement in articular wear and the use of large-diameter femoral heads following the prospect for reduction in the risk of dislocation. The purpose of this study was to compare the clinical and radiographic outcomes as well as serum metal ion level between the two different component designs with small and large femoral heads in metal on metal total hip arthroplasty.

We studied 39 patients with large head (Magnum®, Biomet; cup size minus 6 mm) and 37 patients with small head (M2a taper®, Biomet; 28 or 32 mm head) of metal on metal total hip arthroplasty between December 2009 to October 2011 with follow-up of 2.1 years (1.0–3.3 years) after surgery. Harris Hip Score, UCLA activity score, EQ-5D, radiographic assessment, and serum cobalt and chromium ion levels were evaluated.

Harris Hip Score, UCLA activity score, and EQ-5D were improved after surgery in small and large head groups, however, no significant differences were observed between both groups. Cup inclination was below 50 degree in all prosthesis. No loosening and no osteolysis were observed. Cobalt and chromium ion was not detected before surgery; however, metal ion levels of both groups were increased after surgery in time dependent manner. There was no significant difference between two groups at one year after surgery. One patient in each group showed the increase of cobalt ion level over 7 ppb (15.4 ppb, 12.9 ppb) without any clinical symptom including pain. Cup inclination was 29 degree in both patients and cup anteversion was 38 and 41 degree, respectively.

There was no significant difference of ion levels between both groups. No dislocation was observed in large head group while one dislocation occurred in small head group. No patients required the revision surgery.

This metal on metal component, especially with large femoral heads, showed the good clinical results at the maximum follow up of 3.3 year after surgery. However, the ion level of two patients increased over 7 ppb and longer follow-up will be needed.


*Email: