header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Research

NOISE CHARACTERISTICS IN TWO TYPES OF TOTAL HIP ARTHOPLASTIES: A COMPARATIVE COHORT STUDY

West of Scotland Research Society (WOSORS) - Glasgow Meeting of Orthopaedic Research (GLAMOR)



Abstract

The link between squeaking and ceramic on ceramic (CoC) bearings has been widely reported in orthopaedic literature and is described as a hard bearing phenomenon. We aim to look at the incidence of noise in CoC bearings compared to Metal on Polyethylene (MoP) bearing, which have yet to be linked to squeaking.

We developed a noise characterizing hip questionnaire and sent that along with the Oxford Hip Score (OHS) to 1000 patients; 3:2 ratio of CoC to MoP. 282 CoC patients and 227 MoP patients returned the questions: 509 patients in total. Our patient database provided details on femoral head size and the acetabular inclination angle, for each respondent

47 (17%) of the CoC hip patients reported noise compared to 19 (8%) of the MoP hip patients (P=0.054). 9 CoC patients and 4 MoP patients reported squeaking, while clicking was the most frequent answer in both groups. 27% patients with noise reported avoiding recreational activities because of it. Patient's with noisy hips scored on average, 5 points worse in the OHS (CoC: P = 0.04 and MoP: P = 0.007) and were on average 5 years younger (CoC: P<0.001 and MoP: P=0.007). No correlation was found between noisy hips and femoral head size or inclination angle.

The squeaking hip phenomenon is not exclusive to hard bearing THA. Noise from patient's hips may have social implications and this should be highlighted when consenting a patient for either of these hip procedures. In both implants, we showed there to be a correlation between noise production and a lower OHS. However, longer follow up studies are needed to link noise to a poorly functioning implant


Correspondence should be sent to: Mr B. Jones; email: