header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Research

QUANTITATIVE MEASUREMENT OF MECHANICAL ALIGNMENT AND CORONAL LAXITY DURING EARLY KNEE FLEXION

West of Scotland Research Society (WOSORS) - Glasgow Meeting of Orthopaedic Research (GLAMOR)



Abstract

Non-invasive assessment of lower limb mechanical alignment and assessment of knee laxity using navigation technology is now possible during knee flexion owing to recent software developments. We report a comparison of this new technology with a validated commercially available invasive navigation system.

We tested cadaveric lower limbs (n=12) with a commercial invasive navigation system against the non-invasive system. Mechanical femorotibial angle (MFTA) was measured with no stress, then with 15 Nm of varus and valgus moment. MFTA was recorded at 10° intervals from full knee extension to 90° flexion. The investigator was blinded to all MFTA measurements. Repeatability coefficient was calculated to reflect each system's level of precision, and agreement between the systems; 3° was chosen as the upper limit of precision and agreement when measuring MFTA in the clinical setting based on current literature.

Precision of the invasive system was superior and acceptable in all conditions of stress throughout flexion (repeatability coefficient <2°). Precision of the non-invasive system was acceptable from extension until 60° flexion (repeatability coefficient <3°), beyond which precision was unacceptable. Agreement between invasive and non-invasive systems was within 1.7° from extension to 50° flexion when measuring MFTA with no varus / valgus applied. When applying varus / valgus stress agreement between the systems was acceptable from full extension to 30° knee flexion (repeatability coefficient <3°). Beyond this the systems did not demonstrate sufficient agreement.

These results indicate that the non-invasive system can provide reliable quantitative data on MFTA and laxity in the range relevant to knee examination.


Correspondence should be sent to: Mr D. Russell; email: