header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

ASSESSMENT OF PRECISION AND ACCURACY OF COMPUTER NAVIGATION IN TOTAL HIP ARTHROPLASTY

Computer Assisted Orthopaedic Surgery (CAOS) 13th Annual Meeting of CAOS International



Abstract

Introduction

Restoration of normal hip biomechanics is vital for success of total hip arthroplasty (THA). This requires accurate placement of implants and restoration of limb length and offset. The purpose of this study was to assess the precision and accuracy of computer navigation system in predicting cup placement and restoring limb length and offset.

Material and Methods

An analysis of 259 consecutive patients who had THA performed with imageless computer navigation system was carried out. All surgeries were done by single surgeon (KD) using similar technique. Acetabular cup abduction and anteversion, medialisation or lateralisation of offset and limb length change were compared between navigation measurements and follow-up radiographs. Precision, accuracy, sensitivity and specificity were calculated to assess navigation for cup orientation and student t-test used for evaluation of offset and limb length change. A p value of <0.05 was considered significant for evaluation.

Results

Mean cup abduction and anteversion was 40.35° (SD, 5.81) and 18.46° (SD, 6.79) in postop radiographs compared to 41° (SD, 5.03) and 14.76° (SD, 6.11) for navigation measurements. Intraoperative navigation measurements had high precision and specificity for determining cup abduction and anteversion (precision >95%, specificity >90%). Accuracy for determining cup abduction was 96.13% compared to 72.2% for cup anteversion. Change in limb length and offset was mean 6.46mm (SD, 5.68) and −1.07mm (SD, 5.75) on radiograph evaluation and 5.41mm (SD, 5.11) and −2.24mm (SD, 5.87) from navigation measurements respectively, the difference being not significant in both (p value > 0.2). Radiograph and navigation had a mean difference of 1.01mm (SD, 2.83) for offset measurements and a mean difference of 1.05mm (SD, 4.37) for postop limb length assessment.

Discussion

To the best of our knowledge this is the largest single surgeon study of navigated THA. We found that computer navigation assessment of acetabular cup abduction and anteversion and limb length and offset restoration has high probability of predicting correct placement of implants. To conclude, navigation can serve as an excellent tool for appropriate placement of implants and restoring limb length and offset in THA.


Email: