CEMENTLESS TOTAL HIP ARTHROPLASTY WITH THE TAPERED RECTANGULAR TITANIUM ZWEYMÜLLER STEM. A MINIMUM 20-YEAR FOLLOW-UP

European Federation of National Associations of Orthopaedics and Traumatology (EFORT) - 12th Congress



Abstract

We present our results of cementless total hip arthroplasty with a tapered, rectangular stem made of titanium-aluminum-niobium alloy. This implant is used since 1979 with only minor modifications. The design of the femoral component achieves primary stability through precision rasping and press-fit implantation.

Between October 1986 and November 1987, two hundred consecutive patients (208 hips) underwent total hip arthroplasty with this tapered, rectangular stem. In all cases the acetabular component was a threaded cup made of titanium.

At a minimum follow-up of twenty years eighty-seven patients were still alive. Sixty-seven patients (69 hips) were available for clinical and radiographic follow-up. The probability of survival of the stem was 0.96 (95% confidence interval, 0.91 to 0.98) and that of the cup was 0.72 (0.62 to 0.80). The probability of survival of both the stem and the acetabular component with revision for any reason as the end point was 0.71 (0.61 to 0.78).

Two stems have been revised due to aseptic loosening. We found various degrees of osteolysis around the acetabular and femoral component (61,7%). At the time of the 20-year follow-up no stem was deemed at risk for loosening.

The key findings of our twenty-year follow-up are the very low rate of revisions of the femoral component and the low rate of distal femoral osteolysis associated with this stem. Our data show that femoral fixation of the stem continues to be secure at a follow-up of twenty years.