MATCHED PAIR TKA RETRIEVAL ANALYSIS: OXIDIZED ZIRCONIUM VS. CRCOMO

European Federation of National Associations of Orthopaedics and Traumatology (EFORT) - 12th Congress



Abstract

Introduction

Oxidized zirconium (OxZr) is used as a ceramic surface for femoral components in total knee arthroplasty (TKA). The aim of this study was to investigate its performance by examining retrieved femoral components and their corresponding PE inserts in matched comparison with conventional chrome/cobalt/molybdenum alloy (CrCoMo).

Methods

11 retrieved posterior stabilized (PS) TKA with an OxZr femoral component were included. From a cohort of 56 retrieved TKA with CrCoMo femoral components, pairs were matched according to duration of implantation, patient age, reason for revision, and BMI. The retrieved tibial polyethylene (PE) inserts were analyzed for wear using the Hood classification. Femoral components were optically viewed at 8–32x magnification and screened for scratching, pitting, delamination, and striation. Profilometry was performed to measure surface roughness of the OxZr components using a non-contact white light profiler.

Results

The prostheses were in situ for a mean of 18.5±10.6 (OxZr) and 19.5±14.3 (CrCoMo) months (p=0.41). None of these cases were revised for problems directly related to the use of OxZr. There was one reaction to the implant in the CrCoMo group. The average wear of the tibial PE inserts was significantly lower with OxZr components (41.5±16.8 vs. 60.1±22.0, p=0.01). The average wear score in the visual analysis of the femoral components was significantly lower for the OxZr (1.6±1.3 vs. 9.5±0.6, p=0.005).

Discussion and Conclusion

Femoral components made of OxZr are less sensitive to in vivo wear damage than those of CrCoMo. PE inlays show significantly less in vivo wear damage in combination with an OxZr femoral component.